YY/T 1540-2017Guide for the check of medical class II biological safety cabinets (English PDF)
医用II级生物安全柜核查指南
This is a limited preview
Buy now to download the full PDF (17 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1540-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
May 2, 2017
Implementation date
April 1, 2018
Scope
YY/T 1540-2017 is the English-translated version of 医用II级生物安全柜核查指南.
YY/T 1540-2017 is the Chinese pharmaceutical industry standard that guides the checking of class II biological safety cabinets in healthcare institutions: the acceptance test when a cabinet is installed, the routine inspection and the annual verification.
It sets the requirements and the test methods for each, covering the downflow and inflow velocities, the airflow smoke pattern, the integrity of the HEPA filters, the containment, the alarms and the electrical safety, so that the user can confirm that a cabinet still protects the operator, the product and the room. Issued on 2 May 2017, it has applied since 1 April 2018.
Found what you need?
Document preview — YY/T 1540-2017
National Standard of the People's Republic of China
- ICS
- 11.140
- Classification
- C 48
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Preface
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Requirements
- 4.1 Acceptance Verification
- 4.2 inspection inspection
- 4.3 Annual Verification
- 4.4 Verification records
- 4.5 Verification conclusion
- 5 test methods
- 5.1 Appearance
- 5.2 sampling port
- 5.3 Alarm and interlocking system
- 5.4 wind speed display
- 5.5 high efficiency filter integrity
- 5.6 Noise
- 5.7 illumination
- 5.8 down the air flow rate
- 5.9 Inflow air flow rate
- 5.10 Airflow Mode
- 5.11 UV lamp (if any)
- 5.12 Personnel, Products and Cross-contamination Protection
- Appendix A (Informative) Class II biological safety cabinet verification of the original record
- A.1 Acceptance Verification Record Form
- A.2 Inspection Record Form
- A.3 Annual Verification Records Form
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
Introduction
The purpose of this standard is to verify the performance of Class II biological safety cabinets in medical institutions, to establish standardized methods, requirements and test methods Provided a guide.
To ensure that Class II biological safety cabinets in normal use, the need for regular verification. This standard specifies the acceptance inspection, inspection Verification and annual verification of the three types of verification. Acceptance inspection refers to the installation of Class II biological safety cabinet after the mobile location and replacement of high efficiency Filter and internal components after maintenance performance verification; inspection inspection is a simple quarterly check on the operation of equipment; annual inspection is to inspect Verification added, provides for a more detailed verification.
Inspectors of medical institutions, such as medical practitioners, undertake inspection inspections. Acceptance verification, annual verification by a skilled verifier, Such as biomedical engineers, medical physicists, medical equipment maintenance personnel, professional inspectors, inspection agencies, or manufacturers bear.
1 Scope
This standard specifies the medical institutions II level biological safety cabinet (hereinafter referred to as "safety cabinet") acceptance inspection, inspection inspection and annual verification Requirements and test methods.
The methods described in this standard are intended to assist users of safety cabinets in verifying the performance of such equipment and are intended primarily for medical practitioners, students Medical engineers, medical physicians, medical equipment maintenance personnel, professional inspectors, inspection agencies or manufacturers.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
YY 0569-2011 Class II biological safety cabinet
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 II biological safety cabinets class IIbiologicalsafetycabinet; BSC
Negative pressure filtration exhaust hood, to prevent the operator and the environment exposed to bioaerosols generated during the experiment.
Class II cabinets have a front window operator, the operator can operate through the front window in the safety cabinet, the front window to inhale mouth Of the negative pressure air flow to protect the operator's safety; the descending air flow filtered by the high efficiency filter is used to protect the experimental items in the safety cabinet; the air flow Filter by high-efficiency filter and discharge safety cabinet to protect the environment. Safety cabinet according to the proportion of exhaust gas flow accounts for the total system flow and internal design structure Divided into A1, A2, B1, B2 total of four types. A1 and A2 type of 70% gas circulation, 30% gas efflux; B1 type of 30% gas circulation Ring, 70% gas efflux; B2 type 100% gas efflux. As B1 and B2 type safety cabinet gas exhaust need to connect a dedicated exhaust line, Therefore, the test method of its discharge filter and A1, A2 efflux filter test different.
3.2 Inspector
After professional training, with the appropriate qualifications, access to safety cabinets for performance testing of personnel, including medical practitioners, biomedical Engineers, medical physicists, medical equipment maintenance personnel, professional inspectors, inspection agencies, or manufacturers.
3.3 Nominal flow rate
The safety cabinet operating point specified by the manufacturer is the descending and inflow airflow rates set by the safety cabinet during normal operation.
3.4 Check check
Systematic actions taken to ensure the safety and effectiveness of safety cabinets in medical facilities include acceptance checks, inspections and annual inspections.
3.5 Acceptance Verification
After the safety cabinets have been delivered to the customer for the first test and after the position has been changed, replace the high-efficiency filter and internal parts after repair test.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
How to Download YY/T 1540-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY 0569-2011 — Class II biological safety cabinets
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0463-2026 — Dentistry - Refractory investment and die materials
YY/T 1540-2017
$205.00