Scope
GB/T 6730.28-2021 is the English-translated version of 铁矿石 氟含量的测定 离子选择电极法.
GB/T 6730.28-2021 specifies an ion-selective electrode method for determining the fluorine content of iron ores, over a range of 0.005 % to 1.00 % by mass. It applies to natural iron ores, iron concentrates and lump ores, including sintered products.
The sample is fused with sodium hydroxide, dissolved in water and hydrochloric acid and dry filtered, and the fluoride is measured potentiometrically in a sodium citrate buffer at pH 5.0.
The standard lists the reagents, fluorine standard solutions and equipment, the blank and verification tests, the preparation of calibration solutions and the calibration curve, the calculation and acceptance of results against reference materials, and the test report.
It replaces GB/T 6730.28-2006, was issued on 31 December 2021 and has applied since 1 July 2022.
Found what you need?
Document preview — GB/T 6730.28-2021
National Standard of the People's Republic of China
- ICS
- 73.060.10
- Classification
- D 31
- Replacing
- GB/T 6730.28-2006
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- Introduction5
- 1 Scope6
- 2 Normative references6
- 3 Terms and definitions7
- 4 Principle7
- 5 Reagents and materials7
- 6 Instruments and equipment8
- 7 Sampling and sample preparation9
- 8 Analysis steps9
- 9 Result calculation and representation12
- 10 Test report14
- Annex A (informative) Constituent documents of GB/T 673016
- Annex B (normative) Procedure for acceptance of sample analysis results22
- Annex C (informative) Raw data of precision test23
1 Scope
This document specifies a method for the determination of fluorine content using ion-selective electrode method.
This document applies to the determination of fluorine content in natural iron ores, iron concentrates, and lump ores, including sintered products. The determination range (mass fraction): 0.005 % ~ 1.00 %.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 6379.1 Accuracy (trueness and precision) of measurement methods and results - Part 1: General principles and definitions
GB/T 6379.2 Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
GB/T 6682 Water for analytical laboratory use - Specification and test methods
GB/T 6730.1 Iron ores - Preparation of predried test samples for chemical analysis
GB/T 8170 Rules of rounding off for numerical values & expression and judgement of limiting values
GB/T 10322.1 Iron ores - Sampling and sample preparation procedures
GB/T 12806 Laboratory glassware - One-mark volumetric flasks
GB/T 12807 Laboratory glassware - Graduated pipettes
GB/T 12808 Laboratory glassware - One-mark pipettes
3 Terms and definitions
There are no terms or definitions to be defined in this document.
4 Principle
The sample is melted with sodium hydroxide and dissolved in water and hydrochloric acid. After dry filtration, in the presence of sodium citrate buffer solution, adjust the pH value of the test solution to 5.0 ± 0.1, and directly conduct potentiometric determination with a fluoride ion-selective electrode.
5 Reagents and materials
Unless otherwise stated, only approved analytical reagents and secondary water complying with GB/T 6682 or water equivalent to its purity shall be used in the analysis.
5.1 Sodium hydroxide, granular, dry.
5.2 Sodium hydroxide solution, 300 g/L.
DISSOLVE 75 g of sodium hydroxide in 250 mL of water.
5.3 Sodium hydroxide solution, 20 g/L.
DISSOLVE 5 g of sodium hydroxide in 250 mL of water.
5.4 Hydrochloric acid, rho ~ 1.19 g/mL.
5.5 Hydrochloric acid, 1 + 2.
5.6 Hydrochloric acid, 1 + 9.
5.7 Sodium citrate buffer solution, c(Na3C6H5O7 · 2H2O) = 1 mol/L.
In a 1 L beaker, WEIGH 294.1 g of sodium citrate and DISSOLVE it in 800 mL of water, USE hydrochloric acid (see 5.5) to adjust the pH to 5.0 ± 0.1. TRANSFER to a 1 L volumetric flask, DILUTE with water to the mark, and MIX well.
5.8 Fluorine standard solution A, stored in plastic bottles.
DRY an appropriate amount of sodium fluoride (purity greater than 99.7 %) at 105 °C.
WEIGH 1.108 g of dried sodium fluoride, DISSOLVE it in water, TRANSFER it to a 1 L volumetric flask, DILUTE it with water to the mark, and MIX well. 1 mL of this solution contains 500 µg of fluorine.
5.9 Fluorine standard solution B, stored in plastic bottles.
USE a pipette to transfer 100.00 mL of fluorine standard solution A (see 5.8), PUT it into a 500 mL volumetric flask, DILUTE it with water to the mark, and MIX well. 1 mL of this solution contains 100 µg of fluorine.
5.10 Fluorine standard solution C, stored in plastic bottles.
USE a pipette to transfer 50.00 mL of fluorine standard solution B (see 5.9), PUT it into a 500 mL volumetric flask, DILUTE it with water to the mark, and MIX well. This solution shall be prepared fresh before use. 1 mL of this solution contains 10 µg of fluorine.
5.11 Fluorine standard solution D, stored in plastic bottles.
USE a pipette to transfer 50.00 mL of fluorine standard solution C (see 5.10), PUT it into a 500 mL volumetric flask, DILUTE it with water to the mark, and MIX well. This solution shall be prepared fresh before use. 1 mL of this solution contains 1 µg of fluorine.
6 Instruments and equipment
Unless otherwise stated during analysis, use ordinary experimental instruments. One-mark volumetric flasks, graduated pipettes, and one-mark pipettes shall comply with the requirements of GB/T 12806, GB/T 12807, and GB/T 12808, respectively.
6.1 Balance: the scale value is 0.1 mg.
6.2 Ion meter (or pH meter): the scale value is 0.1 mV.
6.3 Magnetic stirrer: with a 2.5 cm × 0.4cm stirring rod covered with polyethylene.
6.4 Fluoride ion-selective electrode.
6.5 Calomel reference electrode.
6.6 pH glass electrode.
6.7 Silver crucible or nickel crucible: the generally used capacity is 25 mL ~ 30 mL, but 100 mL can also be used.
6.8 Muffle furnace: temperature control of 650 °C ~ 700 °C.
6.9 Plastic beaker: 100 mL and 200 mL.
8 Analysis steps
8.3 Blank test and verification test
8.3.1 Blank test
A blank test is performed along with the sample analysis, and all reagents shall be taken from the same reagent bottle.
8.3.2 Verification test
During each analysis, an iron ore reference material of the same type and similar properties to the analyzed sample is used and analyzed under the same analysis conditions at the same time. Pre-drying of iron ore reference materials shall be carried out in accordance with 7.2.
When analyzing several similar types of ores at the same time, it can use only one reference material for analysis at the same time.
8.4 Determination
WARNING: It is recommended to wear safety glasses and gloves when melting sodium hydroxide, and be careful when dissolving the melt.
8.4.1 Decomposition of sample
PUT the sample (see 8.2) into a crucible, ADD 3 g of sodium hydroxide (see 5.1) to cover the sample, PUT the crucible into a muffle furnace at 600 °C ~ 650 °C to melt for 10 minutes, TAKE OUT the crucible and ROTATE it for a few seconds, then PUT it back into the furnace to melt for 5 minutes.
TAKE OUT the crucible, COOL it and PLACE it in a 200 mL plastic beaker, ADD 60 mL of hot water to extract the melt, WASH the crucible with 10 mL of hydrochloric acid (see 5.5) and WIPE the crucible wall. After the test solution is cooled to room temperature, TRANSFER it to a 100 mL plastic volumetric flask, DILUTE it with water to the mark, and MIX well. DRY FILTER with slow quantitative filter paper, DISCARD the initial filtrate, and COLLECT the filtrate in a plastic beaker. COLLECT about 40 mL of the filtrate for splitting.
Before using a silver crucible or nickel crucible, it shall be cleaned by melting a small amount of sodium hydroxide (see 5.1).
When using a 100 mL silver crucible or nickel crucible, extract the melt directly in the crucible.
8.4.2 Preparation of calibration solutions
In a set of 100 mL plastic volumetric flasks, ADD fluorine standard solution (see 5.8 ~ 5.11) according to the amount given in Table 2, ADD 5 mL of sodium hydroxide […]
8.4.5 Calibration curve
On semi-logarithmic paper, take the potential value as the ordinate, the fluoride ion concentration as the abscissa, plot a calibration curve segmentally with 0.30 µg/mL as the node, and read the fluorine content in the test solution on the calibration curve. Or take the potential value as the ordinate, the negative logarithm of the fluoride ion concentration as the abscissa, plot a calibration curve segmentally with 0.30 µg/mL as the node. Calibration curve regression equations can be established using a computer.
9 Result calculation and representation
9.1 Calculation of fluorine content
Calculate the fluorine content omegaF in the sample according to formula (1) or formula (2), expressed in mass fraction (%). Formula (1) is used when plotting a calibration curve on semi-logarithmic paper, and formula (2) is used when establishing a calibration curve with the negative logarithm of the fluoride ion concentration and using the potential value as the ordinate. where: omegaF - the fluorine content in the sample, %; […]
9.2.5 Calculation of final result
The final result of the sample is the arithmetic mean of the acceptable analytical values, or the value determined in accordance with the provisions in Annex B, and is rounded off in accordance with the provisions of GB/T 8170. When the fluorine content is less than 0.01 %, the final result is rounded off to the fourth decimal place; when the fluorine content is greater than or equal to 0.01 %, the final result is rounded off to the third decimal place.
10 Test report
The test report shall include the following information:
a) laboratory name and address; […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 6379.1Accuracy (trueness and precision) of measurement methods and results -- Part 1: General principles and definitions
- GB/T 6379.2Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
- GB/T 6682Water for analytical laboratory use - Specification and test methods
- GB/T 6730.1Iron ores -- Preparation of pre-dried test samples for chemical analysis
- GB/T 8170Rules of rounding off for numerical values & expression and judgement of limiting values
- GB/T 10322.1Iron ores - Sampling and sample preparation procedures
Similar standards
Editions of GB/T 6730.28
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 6730.28-2021 | Iron ores—Determination of fluorine content—Ion-selective electrode method | current edition | Current |
| GB/T 6730.28-2006 | Iron ores—Determination of fluorine content—Ion-selective electrode method | previous edition | Superseded |
This page sells the current edition, GB/T 6730.28-2021. Earlier editions are listed for reference only.
How to Download GB/T 6730.28-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 10322.1-2023 — Iron ores - Sampling and sample preparation procedures
GB/T 12806-2011 — Laboratory glassware -- One-mark volumetric flasks
GB/T 12807-2021 — Laboratory glassware - Graduated pipettes
Secure payment via Stripe
Payments accepted
GB/T 6730.28-2021
$210.00