GB/T 40219-2021General specification for Raman spectrometers (English PDF)

拉曼光谱仪通用规范

Preview PDF

This is a limited preview

Buy now to download the full PDF (20 pages)

Need quotation for your management?

Get a formal quotation for GB/T 40219-2021 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

May 21, 2021

Implementation date

December 1, 2021

Scope

GB/T 40219-2021 is the English-translated version of 拉曼光谱仪通用规范.

GB/T 40219-2021 is the general specification for Raman spectrometers that use a laser as the excitation light source; instruments with other light sources may refer to it.

It defines Raman scattering, Raman shift and Raman spectrum, classifies the instruments and sets out requirements for spectral resolution, shift accuracy, environmental conditions, electromagnetic compatibility and transport.

The test methods explain how to measure spectral resolution with element lamps and shift accuracy with monocrystalline silicon, sulfur, polystyrene and naphthalene reference samples, along with immunity and transport tests.

It also gives the rules for factory and type inspection, and for marking, packaging, transport and storage. Annexes list example lamp spectra and reference Raman shift values.

Instrument makers and test laboratories use it. It was issued on 21 May 2021 and has applied since 1 December 2021.

Found what you need?

Document preview — GB/T 40219-2021

National Standard of the People's Republic of China

ICS
71.040.10
Classification
N 53

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions5
  • 4 Categories6
  • 5 Requirements7
  • 6 Test method9
  • 7 Inspection rules13
  • 8 Marking, packaging, transportation, storage15
  • Appendix A (Informative) Example of low-voltage discharge lamp spectrum17
  • Appendix B (Informative) Reference shift value of Raman spectrum18
  • References20

1 Scope

This standard specifies the terms and definitions, classification, requirements, test methods, inspection rules, markings, packaging, transportation, storage of Raman spectrometers.

This standard applies to Raman spectrometers, which use laser as the excitation light source (hereinafter referred to as the instrument). The instruments, which use other light sources, may make reference to this standard.

2 Normative references

The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) is applicable to this standard.

GB/T 191-2008 Packaging - Pictorial marking for handling of goods

GB/T 2829-2002 Sampling procedures and tables for periodic inspection by attributes (Apply to inspection of process stability)

GB 4793.1-2007 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements

GB 7247.1-2012 Safety of laser products - Part 1: Equipment classification and requirements

GB/T 11606-2007 Methods of environmental test for analytical instruments

GB/T 13384-2008 General specifications for packing of mechanical and electrical product

GB/T 13966-2013 Terminology for analytical instruments

GB/T 17626.2-2018 Electromagnetic compatibility - Testing and measurement techniques - Electrostatic discharge immunity test

GB/T 17626.3-2016 Electromagnetic compatibility - Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test

GB/T 17626.4-2018 Electromagnetic compatibility - Testing and measurement techniques - Electrical fast transient/burst immunity test

GB/T 17626.5-2019 Electromagnetic compatibility - Testing and measurement techniques - Surge immunity test

GB/T 18268.1-2010 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements

GB/T 34065-2017 Safety requirements for analyzers

3 Terms and definitions

The terms and definitions as defined in GB/T 13966-2013, as well as the following terms and definitions, apply to this document.

3.1 Raman scattering

A phenomenon, where monochromatic incident light interacts with the medium to produce scattered light, which has a frequency different from the original incident light; meanwhile the relative frequency shift is symmetrically distributed on both sides of the incident light frequency.

3.2 Raman shift

The difference, BETWEEN the absolute wavenumber of incident light AND the absolute wavenumber of scattered light.

Note: The unit is cm-1.

3.3 Raman spectrum

The distribution of Raman scattering intensity versus Raman shift.

3.4 Raman spectrometer

An instrument, which is used to collect Raman spectrum of specimen, […]

5 Requirements

5.7 Transportation, transportation, storage

When the instrument is transported and packaged, it shall select the test conditions for transportation and storage, according to the requirements of the environmental condition grouping in Table 1 of GB/T 11606-2007. After the test according to 6.12, the package shall not have significant deformation or damage; the instrument shall not have mechanical damage, such as deformation or loosening, peeling off of coating, etc.; the spectral resolution and shift accuracy shall meet the requirements of 5.3.

6 Test method

6.1 Preparation before test

Set the instrument parameters, according to the requirements of the instrument manual, such as laser wavelength and power, objective lens, slit width, grating line, other related conditions. The laser power is the nominal power OR the recommended value in the instrument manual. When testing the instrument, it shall indicate the environmental temperature, the wavenumber measurement range declared by the instrument, the wavelength and power of the excitation light, the size of the slit, the number of grating lines and focal length of the spectroscopic system, the sampling integration time, as well as other instrument setting conditions. For confocal Raman microscopy spectrometers, it shall also indicate the obvious micro objective lens, pinhole size, as well as other instrument parameters, which are closely related to the confocal performance.

Unless otherwise specified, the following test methods from 6.2 to 6.8 are required to be carried out, under the same conditions as above.

6.2 Test conditions

After the instrument has been prepared for the test, the performance test shall be started under the following test conditions:

a) Unless otherwise specified, the test shall be carried out under the conditions specified in 5.1; the temperature fluctuation shall not exceed ±2 °C;

b) The instrument shall be preheated, in accordance with the requirements of the instrument specification, before the test;

c) Test supplies: Element lamps: neon lamps, argon lamps, mercury lamps, element lamps of other wave band, etc.;

d) Test sample: monocrystalline silicon (111); elemental sulfur (analytical pure), naphthalene (analytical pure), polystyrene flakes.

6.3 Functional inspection

In accordance with the provisions of 5.2, carry out visual observation and hand-feeling inspection, item by item.

6.4 Spectral resolution

Turn off the laser. Within the measurable spectrum range of the instrument, select the appropriate characteristic spectral lines of the element lamp, near the lower limit, upper limit, middle, respectively (see Appendix A for example). After the intensity of the light, which is output by the element lamp, is attenuated, introduce the measurement optical path of the spectrometer. Select three characteristic spectral lines of strong signals, at approximately equal interval, for measurement. Measure and record the full width at half maximum of each spectral peak of the element lamp. Repeat the measurement 6 times. Take the average value as the resolution of each band, under the corresponding excitation wavelength. Take the maximum value as the spectral resolution of the instrument.

6.5 Shift accuracy

Turn on the laser. According to the test conditions in 6.2, place the standard sample at the focus of the instrument signal acquisition probe. Measure the Raman spectrum of the reference sample, according to the wavenumber measurement range, which is declared by the instrument. In the range of 50 cm-1 ~ 200 cm-1, measure the Raman peak of elemental sulfur, at 153.8 cm-1.

In the range of 200 cm-1 ~ 2000 cm-1, measure the Raman peak of polystyrene, at 1001.4 cm-1. In the range of 2000 cm-1 ~ 4000 cm-1, measure the Raman peak of naphthalene, at 3056.4 cm-1. Record the shift value of the selected characteristic peak. Repeat the measurement 3 times. Calculate the difference, BETWEEN the measured shift value of the corresponding characteristic peak AND the standard shift value (see Appendix B), according to formula (1). Take the difference, which has the largest absolute value, as the shift accuracy of instrument.

6.11.3.1 Electrostatic discharge immunity

Carry out the test, in accordance with the contact discharge test procedure, which is specified in GB/T 17626.2-2018.

6.11.3.2 Radio frequency electromagnetic field radiation immunity

Carry out the test, in accordance with the procedure, which is specified in GB/T 17626.3-2016.

6.11.3.3 Electrical fast transient pulse group immunity

Carry out the test, in accordance with the procedure, which is specified in GB/T 17626.4-2018.

6.11.3.4 Surge (shock) immunity

Carry out the test, in accordance with the procedure, which is specified in GB/T 17626.5-2019.

6.12 Transportation, transportation, storage test

When the instrument is in the packaged state, it is tested according to the methods in Chapter 8, Chapter 15, Chapter 16, Chapter 17, Chapter 18 in GB/T 11606-2007.

7 Inspection rules

7.1 Classification of inspections

Instrument inspection is divided into exit-factory inspection and type inspection.

7.2 Exit-factory inspection

7.2.1 See Table 2 for the item categories of the exit-factory inspection.

7.2.2 Each instrument shall be qualified in the inspection. If the exit-factory inspection fails OR if there is nonconformances, it shall be reworked and re-inspected. Only after the re-inspection of all items are qualified, it can be exit-factory.

7.3 Type inspection

7.3.1 Timing of inspection

Type inspection is carried out in one of the following situations:

a) When the design is type-finalized OR the production is type-finalized; […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 20 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download GB/T 40219-2021

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
20 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 40219-2021

$225.00