Valid

GB/T 25752-2010Differential pressure air leak tester (English PDF)

差压式气密检漏仪

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for GB/T 25752-2010 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

December 23, 2010

Implementation date

October 1, 2011

Scope

GB/T 25752-2010 is the English-translated version of 差压式气密检漏仪.

GB/T 25752-2010 is the Chinese standard "Differential pressure air leak tester". Its scope clause reads: This standard specifies the terms and definitions differential pressure airtight leak detectors, classification, structure and basic parameters, test conditions, test methods, inspection Rules, marking, packaging, transportation and storage.

This standard applies to the differential pressure sensor for detecting a pressure difference between the measured object and the reference material for the pressure-sensing element to determine whether the presence of analyte vent Leak size and leak rates of all types of differential pressure type airtight leak detector (hereinafter referred to as the leak detector).

Its clauses include terms and definitions; classification, structure and basic parameters.

Found what you need?

Document preview — GB/T 25752-2010

National Standard of the People's Republic of China

ICS
23.160
Classification
J 78

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • 1 Scope
  • 2 Normative references
  • 3 Terms and Definitions
  • 4 classification, structure and basic parameters

1 Scope

This standard specifies the terms and definitions differential pressure airtight leak detectors, classification, structure and basic parameters, test conditions, test methods, inspection

Rules, marking, packaging, transportation and storage.

This standard applies to the differential pressure sensor for detecting a pressure difference between the measured object and the reference material for the pressure-sensing element to determine whether the presence of analyte vent

Leak size and leak rates of all types of differential pressure type airtight leak detector (hereinafter referred to as the leak detector).

2 Normative references

The following documents contain provisions which, through reference in this standard and become the standard terms. For dated references, subsequent

Amendments (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement are based on research

Whether the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.

GB/T 191 Packaging - Pictorial signs (GB/T 191-2008, ISO 780.1997, MOD)

GB/T 15479-1995 industrial automation instruments of insulation resistance, dielectric strength technical requirements and test methods

3 Terms and Definitions

The following terms and definitions apply to this standard.

3.1

In the differential pressure sensor detects the pressure difference between the measured object and the reference material for the pressure-sensing element to determine whether there is leakage measured object size and leak rate

An instrument or device.

3.2 Reference material master

And the shape, material and volume of the same analyte and the leakage rate and the measured object can be neglected compared to the reference object.

3.3 Volume corrector

Through precise, small amount of change in volume, so that the leak detector generates a pressure difference of different devices.

4 classification, structure and basic parameters

4.1 Classification

Leak test pressure divided by.

a) Vacuum leak test pressure of less than 1 × 105Pa (gauge pressure range of -100kPa ~ 0kPa);

b) micro-pressure leak test pressure in the range of 1 × 105Pa ~ 1.5 × 105Pa (gauge pressure range of 0kPa ~ 50kPa);

c) a low pressure leak test pressure in the range of 1.5 × 105Pa ~ 3 × 105Pa (gauge pressure range of 50kPa ~ 200kPa);

d) the pressure leak test pressure in the range of 3 × 105Pa ~ 8 × 105Pa (gauge pressure range of 200kPa ~ 700kPa);

e) high-pressure leak test pressure is higher than 8 × 105Pa (gauge pressure greater than 700kPa).

4.2 Structure

Detection of analytes with general inflation port of the leak detector (the structure is shown in Figure 1). Unit detected no inflation port sealing components

Leak detector (the structure is shown in Figure 2).

HG/T 6093-2022 | GB 19815-2021 | JB/T 8947-2008 | GB/T 25753.4-2015 | GB/T 25755-2010 | GB/T 25754-2010 | […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Similar standards

How to Download GB/T 25752-2010

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included