Scope

GB/T 21186-2007 is the English-translated version of 傅立叶变换红外光谱仪.

GB/T 21186-2007 is the Chinese standard "Fourier transform infrared spectrometer". Its scope clause reads: This Standard specifies the requirements, test methods, inspection rules, marking, packaging, transportation, storage, quality assurance, etc. of Fourier transform infrared spectrometer.

This Standard applies to the Fourier transform infrared spectrometer of the measuring band in the mid-infrared region (hereinafter known as "instrument" for short).

Its clauses include requirements; test methods; inspection rules; marking, packaging, transportation, and storage; quality assurance. It was issued by the General Administration of Quality Supervision, Inspection and Quarantine on SEPTEMBER 12, 2007 and took effect on MAY 01, 2008.

Found what you need?

Document preview — GB/T 21186-2007

National Standard of the People's Republic of China

ICS
71.040.01
Classification
N 53

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Requirements4
  • 4 Test methods7
  • 5 Inspection rules10
  • 6 Marking, packaging, transportation, and storage13
  • 7 Quality assurance14

1 Scope

This Standard specifies the requirements, test methods, inspection rules, marking, packaging, transportation, storage, quality assurance, etc. of Fourier transform infrared spectrometer.

This Standard applies to the Fourier transform infrared spectrometer of the measuring band in the mid-infrared region (hereinafter known as "instrument" for short).

2 Normative references

The following documents contain provisions which, through reference in this Standard, constitute provisions of this Standard. For the dated references, their subsequent amendments (excluding corrections) or revisions do not apply to this Standard. However, the parties who enter into agreement based on this Standard are encouraged to investigate whether the latest editions of these documents are applicable. For undated reference documents, the latest editions apply to this Standard.

GB/T 191-2000 Packaging - Pictorial marking for handling of goods (eqv ISO 780:1997)

GB/T 2829-2002 Sampling procedures and tables for periodic inspection by attributes (Apply to inspection of process stability)

GB/T 15464-1995 General-purpose specification for the packaging of instrumentation products

JB/T 9329-1999 Basic environmental conditions and testing methods for instruments transportation and storage in the transportation

3 Requirements

3.3 100%tau line tilt range The tilt range of the multi-band 100%tau line of the instrument is shown in Table 1.

3.1 Normal working conditions of instrument

3.2 Background spectral energy distribution

The energy value at 4000 cm-1 shall be no less than 20% of the energy value at the highest point.

3.5 Transmittance repeatability

The transmittance repeatability of the instrument shall be no more than 0.5%tau.

3.6 Resolution

The highest resolution of the instrument shall be selected from 4 cm-1, 2 cm-1, 1 cm-1, 0.5 cm-1, 0.25 cm-1, and 0.125 cm-1.

4 Test methods

4.3 100%tau line tilt range Same as 4.2 setting. The air background spectrum and the air sample spectrum are collected. The 100%tau line is measured. The transmittances of wave-number segments of 800 cm-1~500 cm-1, 2200 cm-1~1900 cm-1, 3200 cm-1~2800 cm-1, and 4400 cm-1~4000 cm-1 are measured respectively.

4.4 100%tau line noise Same as 4.2 setting. The air background spectrum and the air sample spectrum are collected. Obtain the 100%tau line. The RMS values of transmittances of wave-number segments of 1000 cm-1~900 cm-1, 2200 cm-1~2100 cm-1 (or 2100 cm-1~2000 cm-1), and 4100 cm-1~4000 cm-1 are calculated.

4.1 Test conditions

4.2 Background spectral energy distribution

SET the resolution to 4 cm-1.The scanning speed is placed at the optimal position. The number of scans is 32.

4.6 Resolution

4.6.1 SET the highest resolution of the instrument. The scanning speed is placed at the optimal position. The number of scans is 32.

4.7 Wave-number accuracy

4.7.1 For instruments with a resolution better than 0.5 cm-1 (including 0.5 cm-1), SET the highest resolution of the instrument. The scanning speed is placed at the optimal position. The number of scans is 32.The apodization function is BOXCOR. The instrument diaphragm is at the minimum.

4.8 Wave-number repeatability

4.8.1 For instruments with a resolution better than 0.5 cm-1 (including 0.5 cm-1), SET the highest resolution of the instrument. The scanning speed is placed at the optimal position. The number of scans is 32.The apodization function is BOXCOR. The instrument diaphragm is at the minimum. COLLECT the background spectrum. PUT in the gas cell filled with carbon monoxide gas (see Table 3 for the pressure), to collect the sample's transmittance spectrum. The accuracy of carbon monoxide gas 2193.36cm-1 is measured. The measurement is repeated 6 times in succession. Take the difference between the maximum and the minimum.

4.9 Safety tests

No. Absorption peak position

4.9.1 Insulation resistance

The power plug of the instrument is not connected to the grid. The power switch is placed in the on position. USE an insulation resistance meter to apply a 500 V DC test voltage between the phase-to-center connection of power plug and and the ground wire. After 5 s of stabilization, the insulation resistance is measured.

4.10 Instrument appearance

Visual and hand touch inspection.

4.11 Instrument completeness

Visual inspection.

4.12 Transportation, storage in the transportation

It shall be carried out according to the methods of 4.1~4.5 in JB/T 9329-1999.

5 Inspection rules

5.1 Inspection classification

The inspection of the instrument is classified into exit-factory inspection and type inspection.

5.2 Exit-factory inspection

5.3 Type inspection

5.3.1 The instrument shall, in any of the following cases, in accordance with the requirements of 3.2~3.12, be subjected to type inspection.

5.3.2 Samples for type inspection shall be randomly selected from instruments which have passed the exit-factory inspection.

5.3.5 If the type inspection is qualified, the lot which has passed the exit-factory inspection, as a qualified product, can leave the factory or be put in storage. If the storage is more than 12 months before leaving the factory, the exit-factory inspection shall be carried out again.

6 Marking, packaging, transportation, and storage

6.1 Marking

6.1.1 Instrument marking
6.1.2 Packaging marking

6.2 Packaging

6.2.1 The packaging of the instrument shall comply with the provisions of moisture-proof and shock-proof packaging in GB/T 15464-1995.

6.2.2 Instrument accompanying documents […]

6.3 Transportation

The instrument, when the package is complete, can be transported by general means of transport.

6.4 Storage

When the instrument is in the packaged state, it shall be stored in the room where the ambient temperature is 0°C~40°C, the relative humidity shall not be more than 85%, and the air shall not contain corrosive gases.

7 Quality assurance

Under the condition that the user complies with the rules of custody and use, when the instrument, within 12 months from the date of delivery, fails to work normally due to poor manufacturing quality, the manufacturer shall repair or replace the parts for the user for free (excluding the replacement of quick-wear and vulnerable parts). […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download GB/T 21186-2007

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 21186-2007

$125.00