GB/T 15981-2021Evaluating method for the efficacy of sterilization for disinfection equipment (English PDF)

消毒器械灭菌效果评价方法

Preview PDF

This is a limited preview

Buy now to download the full PDF (24 pages)

Need quotation for your management?

Get a formal quotation for GB/T 15981-2021 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

December 31, 2021

Implementation date

July 1, 2022

Scope

GB/T 15981-2021 is the English-translated version of 消毒器械灭菌效果评价方法.

GB/T 15981-2021 fixes how the sterilising efficacy of a disinfection equipment is verified.

It sets the test equipment, the test procedures, the rules by which the result is evaluated and the precautions to be observed, for pressure steam sterilizers, dry heat sterilizers, ethylene oxide sterilizers, low temperature steam formaldehyde sterilizers and hydrogen peroxide gas plasma sterilizers.

It applies to the identification testing of the sterilising efficacy of those machines. Hospitals, sterilizer makers and the laboratories that qualify a machine use it.

It replaces GB 15981-1995, was issued on 31 December 2021 and has applied since 1 July 2022.

Found what you need?

Document preview — GB/T 15981-2021

National Standard of the People's Republic of China

ICS
11.080
Classification
C 59
Replacing
GB 15981-1995

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Terms and definitions5
  • 4 Identification test for efficacy of sterilization for disinfection equipment6
  • Annex A (informative) Reagents and medium recipes18

1 Scope

This document specifies the test equipment, test procedures, evaluation regulations and precautions for the identification test of the efficacy of sterilization for pressure steam sterilizer, dry heat sterilizer (cabinet), ethylene oxide sterilizer, low temperature steam formaldehyde sterilizer, hydrogen peroxide gas plasma sterilizer.

This document is applicable to the evaluation of the efficacy of sterilization for pressure steam sterilizers, dry heat sterilizers (cabinet), ethylene oxide sterilizers, low temperature steam formaldehyde sterilizers, and hydrogen peroxide gas plasma sterilizers.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GBZ 2.1, Occupational exposure limits for hazardous agents in the workplace - Part

1 Chemical hazardous agents

WS/T 649, Hygienic requirement for medical purpose-low temperature steam and formaldehyde sterilizers

WS/T 683, Microbiological requirements for disinfection test

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

3.1 D value

Under the set exposure conditions, the time required to kill 90% of the total number of specific test microorganisms.

3.2 Carrier

Support for test microorganisms.

3.4 Process challenge device; PCD

A specially designed device to simulate the sterilization of articles, resistant to a specific sterilization process, and used to evaluate the effectiveness of the sterilization process.

4 Identification test for efficacy of sterilization for disinfection equipment 4.1 Identification test for efficacy of sterilization for pressure steam sterilizer

4 Identification test for efficacy of sterilization for disinfection equipment

4.2 Identification test for efficacy of sterilization for dry heat sterilizer (cabinet)

4.3 Identification test for efficacy of sterilization for ethylene oxide sterilizer

4.4 Identification test for efficacy of sterilization for low temperature steam formaldehyde sterilizer

4.5 Identification test for efficacy of sterilization for hydrogen peroxide gas plasma sterilizer

4.1.2 Test steps

The test steps are as follows.

4.1.3 Evaluation regulations

4.1.3.1 To determine that the efficacy of sterilization of the sterilizer is qualified, it must meet the requirements at the same time. The color of the positive control medium turns yellow for each test. The negative control medium color does not change. The recovered bacteria volume of the control bacteria tablets is >=1 x 10^5 CFU/tablet. The color of the test group remained unchanged.

4.1.3.2 When self-contained biological indicators are used for evaluation, the results are determined according to the instruction manual.

4.1.4 Precautions
4.2.2 Test steps

The test steps are as follows.

4.2.4 Precautions

4.2.4.1 Strictly abide by aseptic technique regulations.

4.2.4.2 The test shall be carried out under full load conditions.

4.3.2 Test steps

The test steps are as follows.

4.3.3 Evaluation regulations

4.3.3.1 To determine the sterilization effect of the sterilizer is qualified, it must meet the following requirements. The amount of recovered bacteria in the control group shall be greater than or equal to 1 x 10^6 CFU/tablet. There is bacterial growth in the positive control group. There is aseptic growth in the negative control group. All the test groups are aseptically grown.

4.3.4 Precautions

4.3.4.1 Ethylene oxide is flammable, explosive and toxic. To ensure the safety of the test, the operators and test personnel shall be familiar with the performance of ethylene oxide and the operating procedures of the equipment.

4.3.4.2 The working environment shall be well ventilated. The maximum allowable concentration of ethylene oxide in the air at the work site is 2mg/m3.If a person inhales too much ethylene oxide gas, it will cause poisoning symptoms. Severe cases can cause pulmonary edema. If symptoms of poisoning appear, leave the scene immediately and rest in a well-ventilated place. Mild cases breathe fresh air until symptoms clear up.

Severe cases shall be sent to hospital for treatment in time.

4.4.2 Test steps

The test steps are as follows.

4.4.3 Evaluation regulations

4.4.3.1 To determine that the sterilization effect of the sterilizer is qualified, it must meet the requirements at the same time. The amount of testing recovered bacteria of the bacterial count control group in each test is 1 x 10^6 CFU/tablet~5 x 10^6 CFU/tablet. There is bacterial growth in the positive control group. There is aseptic growth in the negative control group. There is aseptic growth in the test group.

4.5.1 Test equipment

4.5.1.1 Test contamination carrier. Bacillus stearothermophilus (ATCC 7953 or SSI K31) spores, with qualified resistance identification. The contamination carrier is a stainless-steel needle with a diameter of <=0.4mm.

4.5.1.2 Reagents. Phosphate buffered saline (PBS, see A.1); tryptone soy agar medium (TSA, see A.3); bromocresol purple peptone medium (see A.4).

4.5.1.4 Fully loaded items. The items specified in the instruction manual.

4.5.1.3 Simulated medical device lumen
4.5.2 Test steps

The test steps are as follows.

4.5.4 Precautions

4.5.4.1 Items must be placed in compliance with the regulations. Sterilized items shall be clean and dry.

4.5.4.2 If the hydrogen peroxide gas plasma sterilizer has multiple sterilization programs, each sterilization program shall be verified separately.

4.5.4.3 The identification of the efficacy of sterilization of the hydrogen peroxide gas sterilizer is carried out with reference to the test method in 4.5.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of GB/T 15981

EditionTitleRevisionStatus
GB/T 15981-2021Evaluating method for the efficacy of sterilization for disinfection equipmentcurrent editionCurrent
GB 15981-1995Evaluating method for the efficacy of sterilization for disinfection equipmentprevious editionSuperseded

This page sells the current edition, GB/T 15981-2021. Earlier editions are listed for reference only.

How to Download GB/T 15981-2021

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
24 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 15981-2021

$210.00