GB/T 14637-2021Determination of copper,iron and zinc in industrial circulating cooling water and scale - Atomic absorption spectrometric method (English PDF)

工业循环冷却水及水垢中铜、铁、锌的测定 原子吸收光谱法

Open the GB/T 14637-2021 preview as PDF

Preview — first pages of GB/T 14637-2021 (full document: 15 pages)

This is a limited preview

Buy now to download the full PDF (15 pages)

Issued by

State Administration for Market Regulation; Standardization Administration of China

Level / Type

National · Recommended

Issue date

August 20, 2021

Implementation date

March 1, 2022

Scope

GB/T 14637-2021 is the English-translated version of 工业循环冷却水及水垢中铜、铁、锌的测定 原子吸收光谱法.

GB/T 14637-2021 is the Chinese national standard on determination of copper,iron and zinc in industrial circulating cooling water and scale - atomic absorption spectrometric method, in the field of chemical technology. The /T suffix marks it as a recommended standard: it is not compulsory by itself, but it becomes binding as soon as a contract, a tender or a customer specification calls it up - which in practice is how most foreign buyers meet it. It was issued on 20 August 2021 by the State Administration for Market Regulation; Standardization Administration of China, and has been in force since 1 March 2022. Classification: ICS 71.040.40, CCS G76. This page is published from the official record of the standard held by the Chinese standards administration: the identification, the dates, the classification and the issuing body are taken from there. The clause text, the tables and the numeric limits are in the document itself, which is delivered complete in English translation.

Document preview — GB/T 14637-2021

National Standard of the People's Republic of China

ICS
71.040.40
Classification
G76

Issued by: State Administration for Market Regulation; Standardization Administration of China

Contents

  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Principle
  • 5 Reagents or materials
  • 5.10 Zinc standard solution II: 5mg/L. Accurately pipette 5.00mL of zinc standard
  • 6 Instruments and equipment...
  • 7 Selection of instrument conditions...
  • 8 Specimen preparation...
  • 9 Determination steps
  • 9.1 Determination of copper content
  • 9.2 Determination of iron content
  • 9.3 Determination of zinc content
  • 10 Result calculation

1 Scope

GB/T 14637-2021 is the Chinese national standard on determination of copper,iron and zinc in industrial circulating cooling water and scale - atomic absorption spectrometric method, in the field of chemical technology. The /T suffix marks it as a recommended standard: it is not compulsory by itself, but it becomes binding as soon as a contract, a tender or a customer specification calls it up - which in practice is how most foreign buyers meet it. It was issued on 20 August 2021 by the State Administration for Market Regulation; Standardization Administration of China, and has been in force since 1 March 2022. Classification: ICS 71.040.40, CCS G76. This page is published from the official record of the standard held by the Chinese standards administration: the identification, the dates, the classification and the issuing body are taken from there. The clause text, the tables and the numeric limits are in the document itself, which is delivered complete in English translation.

This document specifies the determination method of copper, iron and zinc content in industrial circulating cooling water and scale in boiler water system or circulating water system - atomic absorption spectrometry. This document is applicable to the determination of copper content of 0.1mg/L~20mg/L, iron content of 0.1mg/L~20mg/L and zinc content of 0.1mg/L~20mg/L in industrial circulating cooling water; the determination of copper content >= 0.005%, iron content >= 0.01%, zinc content >= 0.005% in scale. This document is also applicable to the determination of copper, iron and zinc in boiler water and the determination of copper, iron and zinc in scale in other industrial water, raw water and water systems, as well as the determination of zinc content in phosphorus-zinc pre-film solution for industrial circulating cooling water.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 4470, Analytical spectroscopic methods. Flame emission, atomic absorption and atomic fluorescence. Vocabulary

GB/T 6682-2008, Water for analytical laboratory use - Specification and test methods DL/T 1151.2, Analytical methods of scale and corrosion products in power plants. Part 2: Sampling and dissolving method for scale and corrosion products

HG/T 3530, Industrial circulating cooling water. Sampling and production for sludge and corrosion products

3 Terms and definitions

For the purposes of this document, the terms and definitions defined in GB/T 4470 apply.

4 Principle

The specimen is atomized and sprayed into the air-acetylene flame. Copper, iron, and zinc are pyrolyzed into ground state atoms. Take copper resonance line 324.7nm, iron resonance line 248.3nm, and zinc resonance line 213.9nm as analysis lines to measure the absorbance of copper, iron, and zinc atoms.

5 Reagents or materials

WARNING -- Acetylene gas used in this document is flammable and explosive when mixed with air. Pay attention to safety. Prevent leakage. Operate strictly according to the specifications. Strong acids are corrosive, so be careful when using them. When splashed on body, rinse with plenty of water. Avoid inhalation or contact with skin. Reagents used in this document, unless otherwise specified, are only analytically pure reagents. See GB 6819 for the requirements of acetylene gas used in the test.

5.1 Water: GB/T 6682-2008, grade two.

5.2 Hydrochloric acid.

5.3 Nitric acid.

5.4 Perchloric acid.

5.5 Nitric acid solution: 1+1.

5.6 Nitric acid solution: 1+99.

5.7 Silver nitrate solution: 10g/L.

5.8 Copper, iron, zinc standard stock solution: 1000mg/L. Commercially available or prepare according to Annex A.

5.9 Copper, iron, zinc standard solution I: 50mg/L. Accurately pipette 5.00mL of standard stock solutions of copper, iron and zinc respectively. Put each into a 100mL volumetric flask. Use water to dilute to the scale. Shake well. The solution is valid for 1 month.

5.10 Zinc standard solution II: 5mg/L. Accurately pipette 5.00mL of zinc standard

a) Weigh about 0.2g~0.3g of the prepared scale sample, accurate to 0.1mg. Place it in a 250mL glass beaker.

b) Add a small amount of water to the beaker to fully wet the scale sample. Slowly add 15mL of hydrochloric acid and 5mL of nitric acid. Cover with a watch glass. Shake well. Slowly heat and boil on a hot plate or an adjustable electric stove for 20min. If there is still a brown or brown residue, add 10mL of hydrochloric acid. Boil until the solution is clear.

c) Remove the beaker. Add 10mL of perchloric acid when the beaker is slightly cooled. Reheat until it starts to emit thick white smoke. Move the watch glass slightly. Continue to heat slowly for 15min~20min. Never evaporate the solution to dryness.

d) Remove the beaker. Add 10mL of nitric acid solution (5.6) while the breaker is hot. Stir well to dissolve the salt on the wall of the beaker.

e) If the scale sample is completely dissolved, transfer the solution to a 100mL volumetric flask. Use a straw to absorb the nitric acid solution (5.6) to wash the inner wall of the beaker (not less than 3 times). Collect the washing solution together in a 100mL volumetric flask. Add nitric acid solution (5.6) to dilute to the scale. Shake well. It shall be the scale sample test solution. If the dissolved solution contains white suspended matter, use medium-speed quantitative filter paper to filter. Use nitric acid solution (5.6) (about 10mL each time) to wash the beaker wall and the sediment and filter paper attached to the wall several times (not less than 5 times). The filtrate and washings are collected together in a volumetric flask. Add nitric acid solution (5.6) to dilute to the scale. Shake well. It shall be the scale sample test solution.

8.2.3 Dissolve scale samples by microwave digestion. Do as follows:

a) Weigh 0.2g~0.3g of the prepared scale sample, accurate to 0.1mg. Place in the digestion inner tank of microwave digestion instrument. Moisten with a few drops of water. Slowly add 6mL of hydrochloric acid and 2mL of nitric acid. Shake the digestion inner tank. Let the gas run out (if a large amount of gas is generated, pre- digest the inner tank on the electric heating plate at 100°C for 20min, and then add the same proportion of digestion solution to about 10mL). Cover the tank tightly and put it in the outer tank. Put the digestion tank into the microwave digester. Digest the scale sample according to the instrument instructions.

b) After the digestion is complete, transfer the solution in the tank to a 250mL glass beaker. Wash the digestion tank and lid with a small amount of nitric acid solution (5.6) and pour it into a beaker. Add 10mL of perchloric acid. Place beaker on hot plate or adjustable hotplate. Slowly heat until white smoke is emitted for 15min~20min. Never evaporate the solution to dryness. Remove the beaker from hot plate or adjustable hotplate. Add 10mL of nitric acid solution (5.6) while the breaker is hot. Stir well to dissolve the salt on the wall of the beaker.

c) If the scale sample is completely dissolved, transfer the solution to a 100mL volumetric flask. Use a straw to absorb the nitric acid solution (5.6) to wash the inner wall of the beaker (not less than 3 times). Collect the washing solution together in a 100mL volumetric flask. Add nitric acid solution (5.6) to dilute to the scale. Shake well. It shall be the scale sample test solution. If the dissolved solution contains white suspended matter, filter it with medium-speed quantitative filter paper. Use nitric acid solution (5.6) (about 10mL each time) to wash the beaker wall and the sediment and filter paper attached to the wall several times (not less than 5 times). The filtrate and washings are collected together in a volumetric flask. Add nitric acid solution (5.6) to dilute to the scale. Shake well. It shall be the scale sample test solution.

8.2.4 Except that the scale sample is not added, follow the same operation as in

8.2.3 to prepare the scale sample test solution to prepare a dissolved sample blank solution.

9.1 Determination of copper content

9.1.1 Drawing of copper calibration curve Accurately pipette 0.00mL (blank), 0.50mL, 1.00mL, 1.50mL, 2.00mL of copper standard solution I. Respectively place in 50mL volumetric flasks. Use nitric acid solution (5.6) to dilute to the scale. Shake well. The copper content of this series of calibration solutions is

0.00 mg/L,

0.50 mg/L, 1.00mg/L, 1.50mg/L, 2.00mg/L. Under the best working conditions of the instrument, at a wavelength of 324.7nm, zero with reagent blank. Measure its absorbance. Take the measured absorbance as the ordinate, and the corresponding copper content (mg/L) as the abscissa, to draw a calibration curve or calculate a regression equation. The linear correlation coefficient of the calibration curve shall be greater than 0.999, otherwise it shall be redrawn.

9.1.2 Determination of water samples Accurately pipette an appropriate amount of specimen solution (8.1). Place in a 50mL volumetric flask. Use nitric acid solution (5.6) to dilute to the scale. Shake well. According to the same instrument conditions in the preparation of the calibration curve, adjust to zero with reagent blank. Measure its absorbance. Find out the corresponding copper content from the calibration curve. If the copper content is lower than the lower limit of detection or exceeds the range of the calibration curve, the dilution factor can be adjusted and measured again.

9.1.3 Determination of scale samples Accurately pipette appropriate amount of scale sample test solution (

8.2.2 or 8.2.3) and dissolved sample blank solution (8.2.4). Place in a 50mL volumetric flask. Use nitric acid solution (5.6) to dilute to the scale. Shake well. According to the same instrument conditions in the preparation of the calibration curve, zero with reagent blank. Measure the absorbance of the dissolved sample blank solution and the scaled sample solution after diluting to constant volume, respectively. Find out the corresponding copper content from the calibration curve. If the copper content is lower than the lower limit of detection or exceeds the range of the calibration curve, the dilution factor can be adjusted and measured again.

9.2 Determination of iron content

9.2.1 Preparation of iron calibration curve Accurately pipette 0.00mL (blank), 1.00mL, 2.00mL, 4.00mL, 6.00mL of iron standard solution I. Respectively place in 50mL volumetric flasks. Use nitric acid solution (5.6) to dilute to the scale. Shake well. The iron content of this standard series is

0.00 mg/L,

6.00 mg/L, respectively. Under the best working conditions of the instrument, at a wavelength of 248.3nm, zero with reagent blank. Measure its absorbance. Take the measured absorbance as the ordinate, and the corresponding iron content (mg/L) as the abscissa, to draw a calibration curve or calculate a regression equation. The linear correlation coefficient of the calibration curve shall be greater than 0.999, otherwise it shall be redrawn.

9.2.2 Determination of water samples Accurately pipette an appropriate amount of specimen solution (8.1). Put in a 50mL volumetric flask. Use nitric acid solution (5.6) to dilute to the scale. Shake well. According to the same instrument conditions in the preparation of the calibration curve, the reagent blank is used for zero adjustment. Measure its absorbance. Find out the corresponding iron content from the calibration curve. If the iron content is lower than the lower limit of detection or exceeds the range of the calibration curve, the dilution factor can be adjusted and re-measured.

9.2.3 Determination of scale samples Respectively and accurately pipette appropriate amount of scale sample test solution (

8.2.2 or 8.2.3) and dissolved sample blank solution (8.2.4). Place in 50mL volumetric flasks (choose the appropriate dilution factor according to the scale source and scale type. Generally, the scale sample mainly composed of calcium and magnesium scale is diluted by about 10 times to 100 times. The scale sample mainly composed of iron oxide scale is diluted by about 100 times to 500 times. Dissolved sample blank solution and scale test solution use the same dilution factor). Use nitric acid solution (5.6) to dilute to the scale. Shake well. According to the same instrument conditions in the preparation of the calibration curve, the reagent blank is used for zero adjustment. Measure the absorbance of the dissolved sample blank solution and the scaled sample solution after diluting to constant volume, respectively. Find out the corresponding iron content from the calibration curve. If the iron content is lower than the lower limit of detection or exceeds the range of the calibration curve, the dilution factor can be adjusted and re- measured.

9.3 Determination of zinc content

9.3.1 Drawing of zinc calibration curve Accurately pipette 0.00mL (blank), 2.00mL, 4.00mL, 6.00mL, 8.00mL of zinc standard solution II. Place in 50mL volumetric flasks, respectively. Use nitric acid solution (5.6) to dilute to the scale. The zinc content of this standard series is 0.00mg/L, 0.20mg/L, 0.40mg/L, 0.60mg/L, 0.80mg/L. Under the optimal conditions of the instrument, at a wavelength of 213.9nm, zero with reagent blank. Measure its absorbance. Take the measured absorbance as the ordinate, and the corresponding zinc content (mg/L) as the abscissa, to draw a calibration curve or calculate a regression equation. The linear correlation coefficient of the calibration curve shall be greater than 0.999, otherwise it shall be redrawn.

9.3.2 Determination of water samples Accurately pipette an appropriate amount of sample solution (8.1). Put it into a 50mL volumetric flask. Use nitric acid solution (5.6) to dilute to the scale. Shake well. According to the same instrument conditions in the preparation of the calibration curve, adjust the reagent blank to zero. Measure its absorbance. Find out the corresponding zinc content from the calibration curve. If the zinc content is lower than the lower limit of detection or exceeds the range of the calibration curve, the dilution factor can be adjusted and re-measured.

9.3.3 Determination of scale samples Respectively and accurately pipette appropriate amount of scale sample solution (

8.2.2 or 8.2.3) and dissolved sample blank solution (8.2.4). Place in 50mL volumetric flasks. Use nitric acid solution (5.6) to dilute to the scale. Shake well. According to the same instrument conditions in the preparation of the calibration curve, the reagent blank is used for zero adjustment. Measure the absorbance of the dissolved sample blank solution and the scale sample solution after diluting to constant volume, respectively. Find out the corresponding zinc content from the calibration curve. If the zinc content is lower than the lower limit of detection or exceeds the range of the calibration curve, the dilution factor can be adjusted and re-measured.

10 Result calculation

Note 1. 2. **Search (Optional).** Enter the keyword '**GB/T 14637-2021**' in the search bar, if it is not already shown. 3.

Note 2. 9.

Note 3. 10. **Verification (USD

Notes and Explanations

Note 1:**[GB/T 14637-2021](https://

Note 3:** If you are unable to redirect back to our Return Page, do not worry. **Additionally and independently**, our automated system will still deliver your PDF and Invoice via email **(the second mechanism to ensure the delivery)**, typically within 3 minutes. Once your payment is complete, you have two instant ways to access your PDF and invoice: you may stay on the Return Page for download in 3 minutes typically, or simply check your email inbox where a copy is automatically delivered in 3 minutes typically.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of GB/T 14637

EditionTitleRevisionStatus
GB/T 14637-2021Determination of copper,iron and zinc in industrial circulating cooling water and scale - Atomic absorption spectrometric methodcurrent editionCurrent
GB/T 14637-2007Industrial circulating cooling water and scale-Determination of copper zinc-Atomic absorption spectrometric methodprevious editionSuperseded

This page sells the current edition, GB/T 14637-2021. Earlier editions are listed for reference only.

How to Buy GB/T 14637-2021

  1. 1Add to cart. Click the "Buy GB/T 14637-2021" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 14637-2021

$155.00

$130.00for partners