YY/T 1723-2020High-throughput gene sequencer (English PDF)
高通量基因测序仪
This is a limited preview
Buy now to download the full PDF (18 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1723-2020 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
June 30, 2020
Implementation date
December 1, 2021
Scope
YY/T 1723-2020 is the English-translated version of 高通量基因测序仪.
YY/T 1723-2020 is the Chinese pharmaceutical industry standard for the high-throughput gene sequencer.
It sets the requirements, the test methods, the labels and instructions for use and the rules for packaging, transport and storage, covering the output of the run and the quality of the reads, the accuracy of the base calling, the coverage uniformity, the carry-over between runs, the stability of the instrument and its software functions.
It applies to massively parallel sequencers whatever their principle: semiconductor sequencing, reversible terminator sequencing, combinatorial probe anchor polymerization or ligation, and pyrosequencing. Issued on 30 June 2020, it has applied since 1 December 2021.
Found what you need?
Document preview — YY/T 1723-2020
National Standard of the People's Republic of China
- Classification
- C 30
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Requirements
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the requirements for high-throughput gene sequencers, test methods, labels and instructions for use, packaging, transportation and storage.
This standard applies to semiconductor sequencing methods, reversible end termination sequencing methods, combined probe-anchored polymerization sequencing methods, and combined probe-anchored ligation Parallel sequencing high-throughput gene sequencer with main technical principles such as sequencing method and pyrosequencing.
This standard does not apply to gene sequencers with Sanger sequencing as the main technical principle.
This standard does not apply to single-molecule sequencing with main technical principles such as total internal reflection microscope, biological nanopore, and solid-state nanopore technology. High-throughput gene sequencer.
2 Normative references
The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this document. For undated reference documents, the latest version (including all amendments) is applicable to this document.
GB/T 191 Graphical signs for packaging, storage and transportation
GB 4793.1 Safety requirements for electrical equipment for measurement, control and laboratory use Part 1: General requirements
GB 4793.9 Safety requirements for electrical equipment for measurement, control and laboratory use Part 9: Laboratory use for analysis and other purposes Special requirements for mobile and semi-automatic equipment
GB/T 14710 Environmental requirements and test methods for medical electrical appliances
GB/T 18266.1 Electromagnetic compatibility requirements for electrical equipment for measurement, control and laboratory use-Part 1: General requirements
GB/T 18268.26 Electromagnetic compatibility requirements for electrical equipment for measurement, control and laboratory use Part 26: Special requirements External diagnostic (IVD) medical equipment
GB/T 29791.3 Information provided by manufacturers of in vitro diagnostic medical devices (labeling) Part 3: In vitro diagnostic devices for professional use
YY 0648 Safety requirements for electrical equipment for measurement, control and laboratory use Part 2-101.In vitro diagnostic (IVD) medical equipment Dedicated requirements
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 Sequencing
The determination of the different base types of nucleic acid molecules, that is, the determination of adenine (A), guanine (G), cytosine (C) and thymus
The composition or sequence of bases such as adenine (T) or uracil (U).
Note: Rewrite GB/T 30989-2014, definition 3.18.
3.2 Data output per run
The number of gene fragments for which sequence information can be obtained in a single sequencing or the deoxyribonucleic acid and ribonucleic acid (expressed in bases) that can be measured Quantity. [GB/T 30989-2014, definition 3.21]
3.3 Output per run
The number of gene fragments or bases that are qualified in a single run
Note: Rewrite GB/T 30989-2014, definition 3.23.
3.4 Read length
The length of a sequence fragment of acceptable quality that can be read in a single run, usually expressed in bases.
Note: Rewrite GB/T 30989-2014 and define 3.24.
3.5 Base calling
The process of converting fluorescent signals or other signals generated by the sequencing reaction into sequence information in the sequencing process.
[GB/T 30989-2014, definition 3.26]
3.6 Base quality value
A measure of the probability of the base being correctly identified. Usually expressed directly in ASCII code.
Note: Rewrite GB/T 30989-2014 and define 3.29.
3.7 High-throughput sequencing
The gene sequencer, which is obviously different from Sanger sequencing, is mainly characterized in that it is not necessary to pre-clear the sequence of the primers of the target fragments.
Columns, based on fragmented DNA, relying on independent reaction system for clonal amplification, capable of performing hundreds of thousands to millions of nucleic acid sequences at a time (DNA) The technical characteristics of parallel molecular sequence determination and generally shorter read lengths are signs.
Note: At present, high-throughput gene sequencers are roughly distinguished according to their throughput and suitability for clinical applications such as human genome, whole exome, and target genes.
Such as sequencing throughput >=20GB/run high-throughput gene sequencer, sequencing throughput <20GB/run and >=2GB/run high-throughput gene sequencer and measurement Sequence throughput< 2Gb/run and >=100Mb/run high-throughput gene sequencer.
4 Requirements
4.1 Appearance requirements
The appearance of the instrument meets the following requirements.
--- The surface of the equipment should not have obvious bumps, scratches, paint stripping, flow marks, and metal parts should not be rusted and mechanically damaged;
---The switches and keys should be flexible and reliable, and the fasteners should be tight and not loose;
---Signs and symbols should be clear and correct.
4.2 Sequencing read length and throughput
Under the prescribed read length mode of sequencing, the sequencing throughput and the effective sequencing throughput should meet the requirements specified by the manufacturer.
Note: The sequencing read length mode is specified according to the performance requirements of the reagents used, such as double-ended 150 bp, single-ended.200 bp and so on.
4.3 The percentage of base recognition quality
Single sequencing, count the percentage of base recognition quality above the specified threshold, and the average percentage of base recognition quality should not be lower than the manufacturer Prescribed requirements.
Note: Sequencers with different principles have different expressions of base recognition quality, such as Q20, Q30, etc.
4.4 Performance Index
4.4.1.1 Sequencing coverage and average sequencing depth
The manufacturer shall specify the sequencing coverage and average sequencing depth of the national reference or standard products.
Note: Refer to Appendix A for the description of national reference products involved in this standard.
4.4.1.2 Sequencing accuracy rate
Under the sequencing coverage and average sequencing depth specified by the manufacturer, the following requirements are met.
a) For detection of human genome (DNA) reference or standard products, the comparison rate should meet the manufacturer's requirements, and should be consistent with single base variation (SNP), insertion Incoming deletion variant (Indel) reference data set comparison, the accuracy and sensitivity of SNP and Indel should meet the manufacturer's requirements;
b) For the detection of human genome DNA reference or standard products, the comparison rate should meet the manufacturer's requirements, and the human genome DNA reference sequence The 1GB reference sequence alignment is specified in the column, and the accuracy of sequencing consensus sequence should not be less than 99.0%;
c) Detect bacterial and viral DNA reference products and compare them with corresponding reference sequences. The accuracy of sequencing consensus sequences should not be less than 99.0%.
Note 1. For gene sequencers that are limited to the detection of human genome DNA, only the reference or standard product of human genome DNA can be tested, that is, a) and b) Detection.
Note 2. For gene sequencers that are limited to the detection of pathogenic microorganisms, only bacterial and viral DNA reference detections can be performed, that is, the detection of c).
Note 3. Gene sequencers for single-stranded sequence analysis can increase the detection indicators of haplotype genome DNA reference products.
4.4.1.3 Repeatability
Performing 3 repetitions of sequencing can meet the requirements of 4.4.1.1 and 4.4.1.2.
4.4.2.1 Sequencing coverage and average sequencing depth
The manufacturer shall specify the sequencing coverage and average sequencing depth of the national reference or standard products.
4.4.2.2 Sequencing accuracy rate
Under the sequencing coverage and average sequencing depth specified by the manufacturer, the following requirements are met.
a) Detect the all-exon region specified in the human genome DNA reference product or standard product, and the comparison rate should meet the manufacturer's requirements, and For comparison of the designated SNP and Indel reference data sets for all exon regions, the accuracy and sensitivity of SNP and Indel should be in line with the manufacturer The requirements;
b) Detect all exon regions specified in the DNA reference product or standard product of the human genome. The comparison rate should meet the manufacturer's requirements.
Alignment of the entire exon region specified in the DNA reference sequence of the human genome, and the accuracy of the sequencing consensus sequence should not be less than 99.0%;
c) Detect bacterial and viral DNA reference products and compare them with corresponding reference sequences. The accuracy of sequencing consensus sequences should not be less than 99.0%.
Note 1. For gene sequencers that are limited to the detection of human genome DNA, only the reference or standard product of human genome DNA can be tested, that is, a) and b) Detection.
Note 2. For gene sequencers that are limited to the detection of pathogenic microorganisms, only bacterial and viral DNA reference detections can be performed, that is, the detection of c).
Note 3. Gene sequencers for single-stranded sequence analysis can increase the detection indicators of haplotype genome DNA reference products.
4.4.2.3 Repeatability
Performing three repetitions of sequencing should all meet the requirements of 4.4.2.1 and 4.4.2.2.
4.4.3.1 Sequencing coverage and average sequencing depth
The manufacturer shall specify the sequencing coverage and average sequencing depth of the national reference or standard products.
4.4.3.2 Sequencing accuracy rate
Under the sequencing coverage and average sequencing depth specified by the manufacturer, the following requirements are met.
a) To detect the 17Kb region specified in the human genome DNA reference product or standard product, the comparison rate should meet the manufacturer's requirements, and Determine the 10 designated SNP sites in the 17Kb area, and the accuracy of SNP detection should not be less than 90%;
b) Detect the sequence alignment of the designated 17Kb region in the human genome DNA reference product or standard product, and the accuracy of the sequencing consensus sequence should Not less than 99.0%;
c) Detect bacterial and viral DNA reference products and compare them with corresponding reference sequences. The accuracy of sequencing consensus sequences should not be less than 99.0%.
Note 1. For gene sequencers that are limited to the detection of human genome DNA, only the reference or standard product of human genome DNA can be tested, that is, a) and b) Detection.
Note 2. For gene sequencers that are limited to the detection of pathogenic microorganisms, only bacterial and viral DNA reference detections can be performed, that is, the detection of c).
Note 3. Gene sequencers for single-stranded sequence analysis can increase the detection indicators of haplotype genome DNA reference products.
4.4.3.3 Repeatability
Performing three repetitions of sequencing should all meet the requirements of 4.4.3.1 and 4.4.3.2.
4.5 Software functions
The software function of a high-throughput gene sequencer should include at least two parts. the instrument operation control function and the signal real-time acquisition function.
a) The instrument operation control software should be able to guide the completion of instrument operation and control, and meet the manufacturer's requirements; […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB 4793.1Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
- GB/T 14710The environmental requirement and test methods for medical electrical equipment
- GB/T 18266.1Ranking standard for gymnasium and playground--Part 1:Star-rating standard for bowling playground
- GB/T 18268.26Electrical equipment for measurement, control and laboratory use -- EMC requirements -- Part 26: Particular requirements -- In vitro diagnostic (IVD) medical equipment
- GB/T 29791.3In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 3: In vitro diagnostic instruments for professional use
GB 4793.9
Similar standards
How to Download YY/T 1723-2020
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 4793.1-2007 — Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
GB/T 18266.1-2000 — Ranking standard for gymnasium and playground--Part 1:Star-rating standard for bowling playground
Secure payment via Stripe
Payments accepted
YY/T 1723-2020
$175.00