YY/T 1564-2017Cardiovascular implants-Pulmonary valve conduit -Pulsatile flow test methods (English PDF)
心血管植入物肺动脉带瓣管道体外脉动流性能测试方法
This is a limited preview
Buy now to download the full PDF (19 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1564-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 28, 2017
Implementation date
April 1, 2018
Scope
YY/T 1564-2017 is the English-translated version of 心血管植入物肺动脉带瓣管道体外脉动流性能测试方法.
YY/T 1564-2017 is the Chinese pharmaceutical industry standard that gives the in vitro pulsatile flow test for heart valve substitutes.
It sets out the pulse duplicator and the test fluid, the flow and pressure conditions that stand for the circulation, the measurements of the pressure difference across the valve, the regurgitant volume and the effective orifice area, and the way the results are reported, so that valves can be compared. Issued on 28 March 2017, it has applied since 1 April 2018.
Found what you need?
Document preview — YY/T 1564-2017
National Standard of the People's Republic of China
- ICS
- 11.040.30
- Classification
- C 30
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Test methods
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the valve with pulsating flow in vitro performance testing methods.
This standard applies to flasks in vitro pulsatile flow test.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB 12279 Cardiovascular implants Artificial heart valves (ISO 5840:1996, IDT)
ISO 5840-3-2013 Cardiovascular implants Artificial heart valves - Part 3. Catheter-implanted prosthetic heart valves (Cardi-ovascularimplants-Cardiacvalve prostheses-Part 3. Heartvalvesubstitutesimplantedbytran-scathetertechniques)
3 Terms and definitions
Terms and definitions defined in GB 12279 and ISO 5840-3:2013 apply to this document.
4 Test methods
4.1 Test system
4.1.1 in vitro pulsatile flow test system should be able to produce similar physiological conditions pulsation, the physiological conditions should be consistent with Table 1, Table 2 Claim.
4.1.2 In vitro pulsatile flow performance test The performance of the system can be confirmed by reference to the valve.
4.1.3 In vitro pulsatile flow performance test system should be able to test the pressure over time, the volume flow rate, velocity field and turbulent shear stress field.
4.1.4 In vitro pulsatile flow performance test system repeatability should be evaluated.
4.1.5 in vitro pulsatile flow test system geometry and mechanical properties should be able to represent the expected site of implantation.
4.1.6 Testers should be able to observe and photograph blood inflow and outflow during all phases of circulation.
4.1.7 The pressure measurement system shall have a natural frequency of at least 50 times the test cycle rate and a measurement accuracy of ± 0.26 kPa (± 2 mmHg) Range.
4.1.8 Flow measurement system measurement accuracy should be within ± 2mL range.
4.1.9 The minimum measurement accuracy of all other measuring equipment shall be within ± 5% of the full scale reading.
4.2 Test conditions
According to the population of pulmonary artery valves, select the appropriate test conditions in Table 1 and Table 2.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 19 pages — is available in the English PDF.
Referenced standards
Normative references
GB 12279 · ISO 5840-3-2013
Similar standards
How to Download YY/T 1564-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
YY/T 0492-2026 — Implantable leads
Secure payment via Stripe
Payments accepted
YY/T 1564-2017
$85.00