YY/T 1823-2022Cardiovascular implants-Standard test method for nickel ion release of nickel-titanium alloy (English PDF)
心血管植入物 镍钛合金镍离子释放试验方法
This is a limited preview
Buy now to download the full PDF (25 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1823-2022 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
May 18, 2022
Implementation date
June 1, 2023
Scope
YY/T 1823-2022 is the English-translated version of 心血管植入物 镍钛合金镍离子释放试验方法.
YY/T 1823-2022 is the Chinese pharmaceutical industry standard that gives the test method for the release of nickel ions in vitro from nickel-titanium alloy cardiovascular implants, and shows by an example how the maximum allowable limit is calculated.
The implant or its specimen is immersed in a defined medium for a defined time at body temperature and the nickel released into that medium is measured, so that the exposure of the patient can be estimated.
It applies to nickel-titanium cardiovascular implants such as vascular stents, cardiac occluders, vena cava filters and heart valves. Issued on 18 May 2022, it has applied since 1 June 2023.
Found what you need?
Document preview — YY/T 1823-2022
National Standard of the People's Republic of China
- Classification
- C 35
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Preface
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions
- 4 Test method
- 5 Result calculation
- 6 Report
- Appendix A (Informative) In Vitro Determination of Nickel Ion AL6 Reference
Foreword
This document is in accordance with the provisions of GB/T 1.1-2020 "Guidelines for Standardization Work Part 1: Structure and Drafting Rules of Standardization Documents" drafted.
Introduction
Nickel ion is a heavy metal ion. Excessive intake of nickel can cause central circulation and respiratory disorders, and cause water in the heart muscle, brain, lung and kidney.
Swelling, bleeding or degeneration may increase the incidence of leukemia and cancer. Studies have shown that nickel has immunotoxicity, especially for those allergic to nickel. crowd. In the adult population, nickel allergy is estimated to occur in 10% of women (17%) higher than men (3%) [22].
Therefore, after Nitinol cardiovascular implants are implanted in the human body, the release rate and release amount of Ni ions from Nitinol alloys are used to evaluate their clinical use.
Important content for security. However, due to the long time, high cost, and low accuracy of the test results in vivo (the body's interference with the test results)
In the preclinical evaluation stage, it is advisable to establish an in vitro nickel ion release model to form an in vitro test for nickel ion release evaluation. method.
At present, there is no similar test standard for in vitro release of nickel ions in China. The formulation of this document will fill this gap and establish Unified test method for nickel ion release in vitro from nitinol cardiovascular implants. If the pitting test results do not meet the expected defined acceptance criteria If the surface finishing process of the device does not have a good history of clinical use [8], it is recommended to use this document to evaluate the nickel ion content through the in vitro test.
1 Scope
This document specifies a test method for the in vitro release of nickel ions from nickel-titanium alloy cardiovascular implants. An example of the calculation of the maximum allowable limit.
This document applies to nickel-titanium alloy cardiovascular implants, including vascular stents, cardiac occluders, vena cava filters, heart valves, etc.
2 Normative references
The contents of the following documents constitute essential provisions of this document through normative references in the text. Among them, dated citations documents, only the version corresponding to that date applies to this document; for undated references, the latest edition (including all amendments) applies to this document.
GB/T 6682 Analysis Laboratory Water Specifications and Test Methods
GB/T 16886.1-2011 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing in Risk Management Process
GB/T 16886.17-2005 Biological Evaluation of Medical Devices Part 17: Establishment of Allowable Limits for Leachables Chinese Pharmacopoeia (four volumes of the 2020 edition)
3 Terms and Definitions
The terms and definitions defined in GB/T 16886.17-2005 and the following apply to this document.
3.1 Detectionlimit
Under the optimal test conditions, measure no less than 7 blank sample solutions to continuously measure 3 times the standard of the response value of the blank sample solution The concentration of the element to be measured corresponding to the deviation (3SD).
3.2 Quantitativelimit
Under the best test conditions, measure no less than 7 blank sample solutions to continuously measure 10 times the standard deviation of the response value of the blank solution.
The concentration of the element to be tested corresponding to the difference (10SD).
3.3 Simulated-use extraction
The test material or sample is extracted with a suitable medium under conditions simulating the use of the product, the purpose of which is to evaluate the Potential hazards to the patient or user during bed use.
Note. The simulated use of leaching should consider all possible conditions such as the tissue contacted by the device, contact temperature and contact time, and assume the most severe conditions of use.
3.4 Sample test area
The surface area of the Nitinol portion of the device that comes into contact with the human body.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 6682Water for analytical laboratory use - Specification and test methods
- GB/T 16886.1-2011Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- GB/T 16886.17-2005Biological evaluation of medical devices - Part 17: Toxicological risk assessment of medical device constituents
Similar standards
How to Download YY/T 1823-2022
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 16886.17-2005 — Biological evaluation of medical devices -- Part 17: Establishment of allowable limits for leachable substances
GB/T 6682-2008 — Water for analytical laboratory use - Specification and test methods
YY/T 0297-2026 — General requirements for clinical investigation of medical devices