Valid

YY/T 1160-2009Carcinoembryonic Antigen Quantitative Detection Reagent(Kit) (Chemiluminescent Immunoassay) (English PDF)

癌胚抗原(CEA)定量测定试剂(盒)(化学发光免疫分析法)

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1160-2009 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

December 30, 2009

Implementation date

June 1, 2011

Scope

YY/T 1160-2009 is the English-translated version of 癌胚抗原(CEA)定量测定试剂(盒)(化学发光免疫分析法).

YY/T 1160-2009 is the Chinese standard "Carcinoembryonic Antigen Quantitative Detection Reagent(Kit) (Chemiluminescent Immunoassay)".

Its scope clause reads: This Standard specifies the terms and definitions, classification, requirements, test methods, inspection rules, marks, labels, operating instructions, packaging, transportation, and storage of carcinoembryonic antigen (CEA) quantitative determination reagent (kit) (chemiluminescent immunoassay).

This Standard applies to the quantitative determination of human carcinoembryonic antigen (CEA) quantitative detection reagent (kit) (hereinafter referred to as "CEA reagent (kit)") based on the principle of chemiluminescent immunoassay.

It includes enzymatic and non-enzymatic chemiluminescent immunoassay detection reagent (box) with carriers such as microplates, pipes, magnetic particles, microbeads, and plastic beads.

The Standard is not applicable to the requirements to the calibration products and quality control products in the kit. Its clauses include terms and definitions; product inspection rules.

Found what you need?

Document preview — YY/T 1160-2009

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions4
  • 4 Classifications6
  • 5 Requirements6
  • 6 Test methods7
  • 7 Product inspection rules9
  • 8 Marks, labels and instructions10
  • 9 Packaging, transportation and storage12
  • Bibliography13

1 Scope

This Standard specifies the terms and definitions, classification, requirements, test methods, inspection rules, marks, labels, operating instructions, packaging, transportation, and storage of carcinoembryonic antigen (CEA) quantitative determination reagent (kit) (chemiluminescent immunoassay).

This Standard applies to the quantitative determination of human carcinoembryonic antigen (CEA) quantitative detection reagent (kit) (hereinafter referred to as "CEA reagent (kit)") based on the principle of chemiluminescent immunoassay. It includes enzymatic and non-enzymatic chemiluminescent immunoassay detection reagent (box) with carriers such as microplates, pipes, magnetic particles, microbeads, and plastic beads.

The Standard is not applicable to the requirements to the calibration products and quality control products in the kit.

2 Normative references

The articles contained in the following documents have become part of this Standard when they are quoted herein. For the dated documents so quoted, all the modifications (excluding corrections) or revisions made thereafter shall not be applicable to this Standard. For the undated documents so quoted, the latest editions shall be applicable to this Standard.

GB/T 21415-2008 In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials (ISO 17511:2003, IDT)

3 Terms and definitions

For the purpose of this Standard, the following terms and definitions apply.

3.1 Chemiluminescence, CL

It refers to the luminescence phenomenon – because the chemical reaction can generate the substances of which the electronic energy level is in the excited state; the electrons can generate photons and release energy through transition; therefore it causes the luminescence […]

6.6 Repeatability

Each sample of 10ng/mL±2ng/mL and 100ng/mL±20ng/mL is repeatedly tested for 10 times.

CALCULATE the mean M and standard deviation SD of the 10 measurement results. Coefficient of variation CV is calculated from Equation of CV=SD/M×100%. The results shall comply with the requirements of 5.6.

6.7 Difference between batches

Repeatedly USE 3 batches of kits to test the same-sample with concentration of

100ng/mL±20ng/mL for 10 times. CALCULATE the mean M and standard deviation SD of 30-times measurement results. CALCULATE the coefficient of variation CV from Equation CV=SD/M×100%.

The results shall comply with the requirements of 5.7.

6.8 Stability

a) Stability within the validity: TAKE the after-expired sample. PERFORM the detection according to the methods of 6.3, 6.4, 6.5 and 6.6. It shall comply with the requirements of 5.8 a).

b) Thermal stability test: TAKE the within-validity sample. PLACE for 3 days at 37°C.

PERFORM the detection according to the methods of 6.3, 6.4, 6.5 and 6.6. It shall comply with the requirements of 5.8 b).

7 Product inspection rules

7.1 General provisions

CEA reagent (kit) must be submitted for inspection in batches. Inspection is divided into ex-factory inspection and type inspection.

7.2

Reagent (kit) must be delivered to the Quality Inspection Department for inspection. It can be sold only after inspection is qualified.

7.3

Reagent (kit) must be submitted for ex-factory inspection in batches. Each material-feeding batch is deemed as 1 batch of products.

7.4 Ex-factory inspection

7.4.1 Amount of sampling

The amount of sampling of ex-factory inspection shall be 3 times of the amount of inspection items, including the amount of inspection, amount of re-inspection and the amount of retained sample.

8.2 Each constituent of reagent (kit)

It shall contain at least the following contents:

a) Product name and specifications;

b) Name or marks of manufacturing enterprise;

c) Product batch number;

d) Period of validity.

8.3 Operating instructions of reagent (kit)

It shall contain at least the following contents:

a) Product name;

b) Packaging specifications;

c) Intended use;

d) Inspection principle;

e) Main compositions;

f) Storage conditions and period of validity;

g) Applicable instrument;

h) Sample requirements;

i) Inspection methods;

j) Reference value (reference range);

k) Interpretation of the inspection results;

l) Limitations of inspection methods;

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Similar standards

How to Download YY/T 1160-2009

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1160-2009

$115.00