YY/T 1163-2009Total Prostate Specific Antigen(t-PSA) Quantitative Detection Reagent(Kit) (Chemiluminescent Immunoassay) (English PDF)
总前列腺特异性抗原(t-PSA)定量测定试剂(盒)(化学发光免疫分析法)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1163-2009 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
December 30, 2009
Implementation date
June 1, 2011
Scope
YY/T 1163-2009 is the English-translated version of 总前列腺特异性抗原(t-PSA)定量测定试剂(盒)(化学发光免疫分析法).
YY/T 1163-2009 is the Chinese pharmaceutical industry standard for total prostate specific antigen (t-PSA) reagent kits that work by chemiluminescent immunoassay on human blood or other body fluids.
It lays down how laboratories and manufacturers judge a t-PSA kit: the terms used, the classification by enzymatic or non-enzymatic chemistry and by carrier, the requirements, and the test methods for minimum detection limit, linearity, repeatability over ten measurements, between-batch difference and stability.
It then fixes the sampling and conformance rules for ex-factory and type inspection, as well as marking, labelling, instructions, packaging, transport and storage.
The calibrators and quality control material in the kit are excluded. Issued on 30 December 2009, it has applied since 1 June 2011.
Found what you need?
Document preview — YY/T 1163-2009
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Classification6
- 5 Requirements6
- 6 Test methods8
- 7 Product inspection rules10
- 8 Marking, labeling, and instructions12
- 9 Packaging, transportation, and storage13
- References14
Foreword
The preparation of this Standard follows the basic specifications of GB/T 1.1-2000 Directives for Standardization. Part 1: Standard Structure and Compilation Rules. It is the basis for evaluating the product quality of total prostate specific antigen (t-PSA) quantitative detection reagent (kit) (chemiluminescent immunoassay).
1 Scope
This Standard specifies the terms and definitions, classification, requirements, test methods, inspection rules, marks, labels, operating instructions, packaging, transportation, and storage of total prostate specific antigen (t-PSA) quantitative detection reagent (kit) (chemiluminescent immunoassay).
This Standard applies to the quantitative detection of total prostate specific antigen (t-PSA) quantitative detection reagent (kit) (herein after referred to as "t-PSA reagent (kit)") of the human blood matrix or other body fluid components based on the principle of chemiluminescent immunoassay. It includes the enzymatic and non-enzymatic chemiluminescence immunoassay detection reagent (kit) with carriers such as microplates, tubes, magnetic particles, microbeads, and plastic beads.
This Standard does not apply to the requirements of the calibrator and quality control material in reagent kits.
2 Normative references
The provisions in following documents become the provisions of this Standard through reference in this Standard. For dated references, the subsequent amendments (excluding corrections) or revisions do not apply to this Standard, however, parties who reach an agreement based on this Standard are encouraged to study if the latest versions of these documents are applicable. For undated references, the latest edition of the referenced document applies.
GB/T 21415-2008 In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrator and control materials (ISO 17511:2003, IDT)
3 Terms and definitions
The following terms and definitions apply to this Standard.
6 Test methods
6.4 Minimum detection limit
USE zero-concentration calibrator or sample diluent as the sample to perform the detection. REPEAT measurement for 20 times. OBTAIN the RLU value (relative luminescence value) of the 20 measurement results. CALCULATE the mean (M) and standard deviation (SD). OBTAIN the M+2SD. CONDUCT two-points regression fitting, based on the concentration - RLU value result BETWEEN zero-concentration calibrator AND adjacent calibrator, to obtain the linear formula. SUBSTITUTE the RLU value of M+2SD into the above formula. OBTAIN the corresponding concentration value, that is, the minimum detection limit. The result shall comply with the requirements of 5.4.
6.5 Linearity
DILUTE the samples with high-value close to the upper limit of linear range INTO at least 5 concentrations according to certain proportion. In which, the samples with low-value concentration shall be close to the lower limit of the linear range. OPERATE according to the instructions of reagent (kit). For each concentration, repeatedly DETECT for 2 times. CALCULATE the mean. USE the least square method to perform the straight-line fitting FOR result's mean and dilution proportion. And CALCULATE the linear correlation coefficient r. The result shall comply with the requirements of 5.5.
6.6 Repeatability
USE samples of 4ng/ML ± 0.8ng/mL and 30ng/mL ± 6ng/mL to repeatedly detect for 10 times. CALCULATE the mean M and standard deviation SD of 10-time measured results. CALCULATE coefficient of variation CV according to the formula CV = SD/M × 100%. The results shall comply with the requirements of 5.6.
6.7 Between-batch difference
USE 3 batches of reagent kits to detect one sample with concentration of (30±6) ng/mL respectively. REPEAT for 10 times. CALCULATE the mean M and standard deviation SD of 30-time measured results. CALCULATE coefficient of variation CV according to the formula CV = SD/M × 100%. The results shall comply with the requirements of 5.7.
6.8 Stability
a) Stability within the period of validity. TAKE the expired sample to inspect according to the methods of 6.3, 6.4, 6.5, and 6.6. It shall comply with the requirements of 5.8 a).
7 Product inspection rules
7.4.4 The retained sample reagent (kit) is used for re-inspection in special cases, such as quality complaints from users. If it is not used in re-inspection, then the reagent (kit) shall be destroyed in two months after the period of validity.
7.4.1 Amount of sampling
The amount of sampling of factory-exit inspection shall be three times of the amount of inspection items, including the amount of inspection, amount of re-inspection and the amount of sample.
7.4.2 Inspection items
The items of factory-exit inspection shall be clarified.
7.4.3 Conformance determination
During the process of inspection, if there is 1 item unqualified, then it shall conduct re-inspection. If there is 1 item unqualified in the re-inspection, then this batch of reagent (kit) is deemed as unqualified.
7.5 Type inspection
7.5.1 When there is one of the following conditions, it shall conduct type inspection.
a) Launch of new products;
b) When there is significant changes to the material, formula and process;
c) Not less than once a year in continuous production;
d) When production is resumed after the long-term shutdown;
e) When it is required by the contract provisions or the administrative department.
7.5.2 Amount of sampling
The amount of sampling of type inspection shall be three times of the amount of inspection items, including the amount of inspection, amount of re-inspection and the amount of sample.
7.5.3 Inspection items
When conducting type inspection, it shall conduct full-inspection. All inspection results shall be qualified.
7.5.4 Conformance determination
If all inspection items are qualified, the type inspection is passed. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1163-2009
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 21415-2008 — In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0463-2026 — Dentistry - Refractory investment and die materials
YY/T 1163-2009
$115.00