Scope
YY/T 1161-2009 is the English-translated version of 肿瘤相关抗原CA125定量测定试剂(盒)(化学发光免疫分析法).
YY/T 1161-2009 is the Chinese pharmaceutical industry standard for kits that measure the tumour-associated antigen CA125 by chemiluminescent immunoassay. It applies to both enzymatic and non-enzymatic chemiluminescent kits, whatever the carrier - microplates, tubes, magnetic particles, microbeads or plastic beads.
After defining chemiluminescence, chemiluminescent immunoassay, the minimum detectable concentration, analytical specificity and linearity, it sets the performance requirements of the CA125 kit and the methods used to test its detection limit, linearity, repeatability, batch-to-batch difference and stability.
It also gives the ex-factory and type inspection rules, and the marking, labels, instructions, packaging, transport and storage. The calibrators and control materials supplied in the kit are not covered. Issued on 30 December 2009, it has applied since 1 June 2011.
Found what you need?
Document preview — YY/T 1161-2009
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Classification6
- 5 Requirements6
- 6 Test methods7
- 7 Inspection rules of products9
- 8 Marks, labels and instructions11
- 9 Packaging, transportation and storage12
- Bibliography13
Foreword
The preparation of this standard follows the basic regulations of the GB/T 1.1-2000 Directives for Standardization Part 1. Rules for the Structure and Preparation of Standards.
It is the basis of product quality to evaluate tumor associated antigen CA125 quantitative determination reagent (kit) (chemiluminescence immunoassay).
1 Scope
This standard specifies the terms and definitions, classification, requirements, test methods, inspection rules, identification, labels, instructions, packaging, transportation and storage of tumor associated antigen CA125 quantitative determination reagent (kit) (chemiluminescence immunoassay).
This standard is applicable to quantitative detection of tumor associated antigen (CA125) reagent (kit) [hereinafter referred to as "CA125 reagent (kit)"], based on the principles of chemiluminescence immunoassay. It includes the enzymatic and non-enzymatic chemiluminescence immunoassay detection reagent (kit) with carriers of micro-plates, tubes, magnetic particles, micro-beads and plastic beads.
This standard is not applicable to the requirements of calibrators and quality control products in the kit.
2 Normative references
The articles contained in the following documents have become part of this document when they are quoted herein. For the dated documents so quoted, all the modifications (Including all corrections) or revisions made thereafter shall be applicable to this document.
GB/T 21415-2008 In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
3 Terms and definitions
For the purpose of standard, the following terms and definitions shall apply.
3.1 Chemiluminescence, CL
Because that the chemical reaction can generate the substances of electronic energy level to be in the excited state, the latter can generate photon by transition release energy, thereby to cause the luminescence phenomenon.
3.2 Chemiluminescent immunoassay, CLIA
The luminescence phenomenon that is caused by - because of chemical reactions, the substances with generated electron energy-level at excited state unleash energy and produce the photons through transition.
3.3 Minimum detectable concentration / detection limit, limit of detection
It refers to the measured minimum value that can be declared as varying from zero at a certain probability in the sample. [ISO/DIS 18113-1]
Note 1. Sometimes it is incorrectly referred to as a sensitivity of analysis.
Note 2. The minimum detection limit in this standard is the minimum concentration which differs from zero and is at the confidence interval of not less than 95%.
3.4 Analytical specificity
The capability of measurement procedure that only measures the sample to-be-measured. [GB/T 19702-2005/ISO 15193:2002, 3.8]
Note 1. Lack of specificity may be referred to the analytical interference.
Note 2. The lack of specificity may be referred as due to cross reaction in the immunochemical measurement procedure.
Note 3. The specificity of the measurement procedure shall not be confused with the diagnostic specificity.
3.5 Linearity of a measuring system
The directly proportional capacity between the given measuring result and the measured value in the sample. [ISO/DIS 18113-1]
Note 1. For in vitro diagnostic (IVD) medical devices, the linearity-related results of measurement are the measurement values after correction or linearization. can only be sold after the inspection is qualified.
7 Inspection rules of products
7.3 Reagent (kit) must be submitted to ex-factory inspection in batches. Each material-feeding batch is deemed as one batch of products.
7.4 Ex-factory inspection
7.4.4 The retained-sample reagent (kit) is used for reinspection in particular cases, such as quality complaints by the users. If the retained-sample is not used for reinspection, then the reagent (kit) shall be destroyed in two months after the validity period.
7.4.1 Sampling quantity
The sampling quantity of ex-factory inspection shall be 3 times as the quantity of inspection items. It includes inspection quantity, reinspection quantity and retained-sample quantity.
7.4.2 Inspection items
The items of ex-factory inspection shall be clear.
7.4.3 Eligibility determination
If any item is unqualified in the process of inspection, reinspection shall be carried out. If any item is unqualified in the reinspection, the batch of reagent (kit) is deemed as unqualified.
7.5 Type inspection
7.5.1 Type inspection shall be conducted in case of one of the following conditions.
a) Production of new products;
b) When there is significant change to the material, formula or process;
c) Not less than once a year during continuous production;
d) When production is resumed after the long-term shutdown;
e) When it is required by the contract provision or the administrative department.
7.5.2 Sampling quantity
The sampling quantity of type inspection shall be three times as the quantity of inspection items. It includes the inspection quantity, reinspection quantity and retained-sample quantity.
7.5.3 Inspection items
For type inspection, it shall conduct full-inspection. All inspection results shall be qualified.
7.5.4 Eligibility determination
If all inspection items are qualified, the type inspection is deemed as qualified. If the type inspection is not qualified, it must not start mass-production.
8 Marks, labels and instructions
8.1 Outer package of reagent (kit)
It shall contain at least the following contents.
a) Product name and specification;
b) Name, address and contact details of the manufacturer;
c) Number of medical device registration certificate number and product standard;
d) Product batch number;
e) The period of validity; f) Storage conditions.
8.2 Each component of reagent (kit)
It shall contain at least the following contents.
a) Product name and specification;
b) Name or trademark of the manufacturer;
c) Product batch number;
d) The period of validity.
8.3 Instructions of reagent (kit)
It shall contain at least the following contents.
a) Product name;
b) Packaging specifications;
c) Intended use;
d) Inspection principle; […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1161-2009
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 21415-2008 — In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0463-2026 — Dentistry - Refractory investment and die materials
YY/T 1161-2009
$115.00