YY/T 0681.12-2022Test methods for sterile medical device package - Part 12: Flex durability of flexible barrier materials (English PDF)

无菌医疗器械包装试验方法 第12部分:软性屏障材料抗揉搓性

Preview PDF

This is a limited preview

Buy now to download the full PDF (18 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0681.12-2022 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

May 18, 2022

Implementation date

June 1, 2023

Scope

YY/T 0681.12-2022 is the English-translated version of 无菌医疗器械包装试验方法 第12部分:软性屏障材料抗揉搓性.

YY/T 0681.12-2022 is the Chinese pharmaceutical industry standard that gives the test methods for the flex durability of the flexible barrier materials used in sterile medical device packaging, that is how well a film or laminate keeps its barrier after being repeatedly folded and twisted as it is in handling and transport.

It is the twelfth part of the YY/T 0681 series and sets out the specimens, the flexing apparatus and its cycles, and the examination that follows, in which the flexed material is checked for pinholes and loss of barrier.

It replaces YY/T 0681.12-2014, was issued on 18 May 2022 and has applied since 1 June 2023.

Found what you need?

Document preview — YY/T 0681.12-2022

National Standard of the People's Republic of China

Replacing
YY/T 0681.12-2014

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • Introduction6
  • 1 Scope8
  • 2 Normative references8
  • 3 Terms and definitions8
  • 4 Overview of test methods8
  • 5 Application9
  • 6 Test instruments9
  • 7 Specimen preparation11
  • 8 Conditioning11
  • 9 Procedures11
  • 10 Report12
  • Annex A (informative) Pinhole counting test14
  • Annex B (informative) Gas and/or water vapor transmission rate test16
  • Bibliography17

Foreword

This document was drafted in accordance with the rules given in GB/T 1.1-2020 "Directives for standardization - Part 1: Rules for the structure and drafting of standardizing documents".

This document is Part 12 of YY/T 0681 "Test methods for sterile medical device package". The following parts of YY/T 0681 have been issued.

- Part 1: Test guide for accelerated aging;

- Part 2: Seal strength of flexible battier materials;

- Part 3: Internal pressurization failure resistance of unrestrained packages;

- Part 4: Detecting seal leaks in porous packages by dye penetration;

- Part 5: Detecting gross leaks in medical packaging by internal pressurization (bubble test);

- Part 6: Evaluation of chemical resistance of printed inks and coatings on flexible packaging materials;

- Part 7: Evaluating inks or coating adhesion to flexible packaging materials using tape;

- Part 8: Coating/adhesive weight determination;

- Part 9: Burst testing of flexible package seals using internal air pressurization weight restraining plates;

- Part 10: Test for microbial barrier ranking of porous package material;

- Part 11: Determining integrity of seals for medical packaging by visual inspection;

- Part 12: Flex durability of flexible barrier films;

- Part 13: Slow rate penetration resistance of flexible barrier films and laminates;

- Part 14: Testing the microbial barrier for porous packaging materials under moist conditions and with passage of air;

- Part 15: Performance testing of shipping containers and systems;

- Part 16: Test for climatic stressing of packaging system;

- Part 17: Testing the microbial barrier performance of porous package materials using aerosol filtration method;

- Part 18: Nondestructive detection of leaks in packages by vacuum decay method.

This document replaces YY/T 0681.12-2014 "Test methods for sterile medical device package. Part 12: Flex durability of flexible barrier films". Compared with YY/T 0681.12-2014, in addition to the structural adjustments and editorial changes, the main technical changes in this document are as follows.

a) The scope of the standard is changed (see Chapter 1 of this Edition; Chapter 1 of Edition 2014);

b) "Terms and definitions" is added (see Chapter 3 of this Edition);

c) The description of evaluation methods for breathable materials such as paper and polyolefin nonwovens is added (see 4.2 of this Edition; 3.2 of Edition 2014);

d) Articles 4.1 and 4.5 of "Significance and application" are deleted (see 4.1 and 4.5 of Edition 2014);

e) The dimensional accuracy of "flexing tester" is added (see 6.1 of this Edition; 5.1 of Edition 2014);

f) "Determination of flexing test results" is added (see 9.3 of this Edition);

g) "Pinhole counting" is deleted (see 8.3 of Edition 2014);

h) The content of "Report" is modified (see Chapter 10 of this Edition; Chapter 9 of Edition 2014);

i) "Annex A Precision and bias" is deleted (Annex A of Edition 2014);

j) "Pinhole counting test" is added (see Annex A of this Edition);

k) "Gas and/or water vapor transmission rate test" is added (see Annex B of this Edition).

1 Scope

This document describes test methods for the flex durability of flexible barrier materials.

This document applies to the testing of the flex durability of flexible barrier materials.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 2918, Plastics - Standard atmospheres for conditioning and testing

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

3.1 Pinhole

A small opening of no particular shape or size that passes completely through all layers of a flexible barrier material.

4 Overview of test methods

4.1 Unless otherwise specified, the flexing test is performed on specimens of flexible barrier materials under standard atmospheric conditions.

4.2 The degree of damage to the structural and/or mechanical properties of the material is judged by the flexing test. The properties to be evaluated in the flexing test determine the appropriate level of test conditions.

4.3 The various test conditions are summarized as follows.

5 Application

5.1 The various conditions described in this test are to prevent the occurrence of too many pinholes that are inconvenient to count and meaningless when testing a specimen structure, and the occurrence of too few pinholes is also meaningless. Generally, the number of pinholes on each sample shall be between 5~50.Material construction, purpose of the test, and agreement between interested parties are important factors to consider when selecting the level of test conditions.

5.2 This test method does not measure any part of wear associated with flux-to-break.

6 Test instruments

6.1 Flexing tester. is designed to be set up according to the specifications listed in Chapter 9: The instrument shall mainly consist of a (90±1)mm diameter fixed shaft and a (90±1)mm diameter moving shaft. When the moving shaft is at the initial position of the stroke (that is, the maximum distance), the two shafts face to face are separated by (180±2)mm. Both shafts shall have vents to protect the sample from pressure. The width

1) Test methods such as tensile strength and/or air permeability.

6.2 Flexible double-sided pressure-sensitive tapes. the width shall not exceed 13mm.

6.3 Templates. used for cutting samples of 200mmx280mm.

7 Specimen preparation

7.1 The samples are cut into sheets of 200mm x 280mm. The 200mm dimension is the test direction. This is also the direction of the flexing tester shaft.

7.2 Four samples are flexed in each of their machine direction and cross direction. In addition, four flexed samples are taken from positions adjacent to the samples in two directions as control samples.

8 Conditioning

According to the provisions of GB/T 2918, the sample shall be conditioned for at least 24h under the conditions that the relative humidity is (50±5)% and the temperature is (23±2)°C (unless otherwise specified between the supplier and the purchaser).

9 Procedures

9.1 Test environment

Unless otherwise specified, the flexing test is carried out under the conditions described in Chapter 8.

9.2 Flexing conditions

9.2.1 Condition A

9.2.2 Condition B

The test conditions are the same as Condition A. The samples are flexed for 20min at 45 cycles/min. (that is, a total of 900 cycles under full flexing and swirling action).

9.2.3 Condition C

The test conditions are the same as Condition A. The samples are flexed for 6min at 45 cycles/min. (that is, a total of 270 cycles under full flexing and swirling action).

9.2.4 Condition D

The test conditions are the same as Condition A. The samples are flexed for 20 times at 45 cycles/min. (that is, a total of 20 cycles under full flexing and swirling action).

9.3 Determination of flexing test results

9.3.1 Remove the flexible barrier material sample from the flexing tester. Mark a 150mm x 200mm area in the center. 150mm is the axial dimension of the tester. For samples after flex durability testing, whether for pinhole or layer integrity inspection, samples shall be taken in this area.

10 Report

10.1 The sample conditioning and test environment used.

10.2 Sample specifications (thickness/gram weight), structure (if appropriate).

10.3 Flexing test conditions.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY/T 0681.12

EditionTitleRevisionStatus
YY/T 0681.12-2022Test methods for sterile medical device package - Part 12: Flex durability of flexible barrier materialscurrent editionCurrent
YY/T 0681.12-2014Test methods for sterile medical device package - Part 12: Flex durability of flexible barrier materialsprevious editionSuperseded

This page sells the current edition, YY/T 0681.12-2022. Earlier editions are listed for reference only.

How to Download YY/T 0681.12-2022

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
18 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0681.12-2022

$210.00