YY/T 0641-2008Standard test method for transformation temperature of nickel—Titanium alloys by thermal analysis (English PDF)
热分析法测量NiTi合金相变温度的标准方法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0641-2008 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
April 25, 2008
Implementation date
June 1, 2009
Scope
YY/T 0641-2008 is the English-translated version of 热分析法测量NiTi合金相变温度的标准方法.
YY/T 0641-2008 is the Chinese pharmaceutical industry standard test method for measuring the transformation temperatures of nickel-titanium shape memory alloys by thermal analysis with a differential scanning calorimeter.
It refers to ASTM terminology for thermal analysis and for nickel-titanium alloys, and sets out the summary and use of the method, possible interferences, the instrument, with heating and cooling rates of 10 °C/min, inert sample pans, nitrogen or helium purge and a 0.1 mg balance, and sample preparation, with 20 mg to 50 mg specimens annealed at 800 °C and cleaned of oxide and cutting fluid.
It closes with temperature calibration, the test procedure, the reading of the transformation temperatures from the curve, the report and the precision and bias. Issued on 25 April 2008, it has applied since 1 June 2009.
Found what you need?
Document preview — YY/T 0641-2008
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Term4
- 4 Summary of test methods5
- 5 Meaning and use5
- 6 Interference5
- 7 Instrument6
- 8 Sample preparation6
- 9 Temperature calibration6
- 10 Test procedure6
- 11 Conversion of graphical data7
- 12 Report8
- 13 Accuracy and deviation8
- 14 Keywords9
1 Scope
This standard specifies the procedure for measuring the transformation temperature of NiTi shape memory alloys. The value indicated in the SI unit is considered standard.
This standard does not indicate that it involves safety related to the use of the standard. It is the responsibility of users of this standard to establish special operating procedures to ensure the safety of use and personnel health before use, and to determine the applicability of the management scope.
2 Normative references
The provisions in following documents become the provisions of this Standard through reference in this Standard. For the dated references, the subsequent amendments (excluding corrections) or revisions do not apply to this Standard; however, parties who reach an agreement based on this Standard are encouraged to study if the latest versions of these documents are applicable. For undated references, the latest edition of the referenced document applies.
ASTM E 473 Standard terminology relating to thermal analysis
ASTM E 967 Standard test method for temperature calibration of differential scanning calorimeters and differential thermal analyzers
ASTM E 1142 Standard terminology relating to thermophysical properties
ASTM F 2005 Standard terminology for nickel-titanium shape memory alloy
3 Term
For the technical terms used in this test method, see ASTM E 473, ASTM E 1142, ASTM F 2005.
7 Instrument
7.1 Use the difference display thermometer which has a heating and cooling rate reaching 10 °C/min, which can automatically record the input energy difference between the measured sample and the reference sample according to the required sensitivity and accuracy.
7.2 Use a sample cuvette or pan made of aluminum or other inert materials with high thermal conductivity.
7.3 Use clean nitrogen or helium.
7.4 Use an analytical balance with a sensitivity of 0.1 mg and a measuring range of 100 mg.
8 Sample preparation
8.1 Use (20 ~ 50) mg of sample. Cut the sample, to get the largest surface in contact with the DSC sample pan.
8.2 Place it in vacuum, or in air at 800 °C for 15 min which has sufficient protection to prevent oxidation for annealing. Quickly cool the sample, to prevent phase precipitation that can change the alloy transformation temperature.
8.3 Clean the sample to remove all foreign material contamination, such as cutting fluid. If the sample is oxidized during heat treatment, grinding, polishing or corrosion, the oxide layer shall be removed. Care shall be taken to avoid cold working of the sample, because cold working will change the thermal effect (characteristic curve) of the sample. Slight oxidation is allowed, but all thick scale shall be removed.
9 Temperature calibration
In accordance with ASTM E 967, use the same heating rate, clean gas, and flow rate to calibrate the temperature axis of the instrument as used in the actual test sample.
10 Test procedure
10.1 Put the sample into the sample pan; place the sample pan on the test bench. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
ASTM E 967 · ASTM E 1142 · ASTM F 2005
Cited by
- GB/T 24627-2023Wrought nickel-titanium shape memory alloys for surgical implants
- YS/T 1136-2016Nickel-titanium shape memory alloy seamless tube for medical devices
- YS/T 1064-2015Terminology for nickel-titanium shape memory alloys
- YS/T 1076-2015Titanium-nickel alloy plate
- YS/T 971-2014Titanium-nickel shape me: ncry alloy wires
Similar standards
How to Download YY/T 0641-2008
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards