Scope
YY/T 0619-2017 is the English-translated version of 医用内窥镜硬性电凝电切内窥镜.
YY/T 0619-2017 is the Chinese pharmaceutical industry standard for the rigid resectoscope, the endoscope through which tissue is resected electrosurgically, above all in the prostate and the bladder.
It gives the classification and composition, the requirements and the test methods, covering the optics and the field and direction of view, the working element and the electrode, the sheath and the irrigation channels, the insulation and the leakage current that matter where high frequency current is used, and the cleaning and sterilization of the instrument. Issued on 28 February 2017, it has applied since 1 January 2018.
Found what you need?
Document preview — YY/T 0619-2017
National Standard of the People's Republic of China
- Replacing
- YY 0619-2007
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative References5
- 3 Classification and Composition6
- 4 Requirements6
- 5 Test Methods11
1 Scope
This Standard specifies the classification and composition, requirements and test methods of rigid resectoscope.
This Standard is applicable to the rigid resectoscope (hereinafter referred to as resectoscope). The resectoscope is clinically applicable to the cutting, vaporizing and coagulating operations against the diseased tissues under the direct viewing of endoscope and using the high-frequency current thermal effect.
2 Normative References
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this document.
GB 9706.4 Medical Electrical Equipment - Part 2-2. Particular Requirements for the Safety of High Frequency Surgical Equipment
GB 9706.19 Medical Electrical Equipment-Part 2. Particular Requirements for the Safety of Endoscopic Equipment
GB/T 14233.1-2008 Test Methods for Infusion Transfusion Injection Equipment for Medical Use - Part 1. Chemical Analysis Methods
GB/T 16886.1 Biological Evaluation of Medical Devices - Part 1. Evaluation and Testing Within a Risk Management Process
YY 0068.1-2008 Medical Endoscopes - Rigid Endoscope - Part 1. Optical Properties and Test Methods
YY 0068.2-2008 Medical Endoscopes - Rigid Endoscope - Part 2. Mechanical Properties and Test Methods
YY 0068.3-2008 Medical Endoscopes - Rigid Endoscope - Part 3. Marking and Instruction Manual possible measures to minimize the undesired hazard
a) Warn the possible hazard indications, means of prevention of hazard, and the handling instructions after hazard is discovered;
b) Safe operating procedures;
c) The instructions that can be used by the operator after special training.
4 Requirements
4.2.2.6.2 In the view filed of the endoscope, the visible head end part of the surgical electrode shall have no directional reflection phenomenon that may influence the use.
4.5.4 Dissolved precipitates of polymer materials in contact with the patient 4.5.4.1 pH. compared with the same batch of blank control solution, the pH difference shall be no greater than 2.0.
4.5.4.2 Total content of soluble heavy metals. the total content of the soluble heavy metals in the dissolution solution shall not exceed 5.0µg/mL.
4.5.4.3 Potassium permanganate reducing substance. the consumption difference with the same volume of the same batch of blank control solution shall be no more than 2.0mL.
4.2.2.7 Visibility of surgical electrode
When the surgical electrode is moved to the distal position with the moving part of the manipulator, its head end shall be within the range observable by the endoscope
4.3 Conduction performance
The high-frequency connection wire and surgical electrode of resectoscope shall be well-conducted; its impedance value shall be less than 3ohm.
4.4 Label and attaching document
4.4.1 General requirements of label and attaching document
It shall meet the requirements of YY 0068.3-2009.
4.4.2 Other requirements of label and attaching document
The manufacturer shall indicate in the instruction manual the possibility of heat and other risks to the patient and operator when using the product.
a) Thermal damage to the endoscope due to electrical spark or capacitively coupling high-frequency current;
b) Image interference on the video screen caused by high-frequency electromagnetic waves;
c) Keep the working part of the charged electrode within the operator's observation range to avoid accidental high-frequency burns;
d) In order to ensure that the charged electrode in the correct position during the operation, it must be kept sufficient distance away from the head end of the endoscope; it shall avoid contacting with the metal parts and other conductors of the endoscope before exciting the high-frequency output, including the liquid vent that may be conductor; application; or information about the material and/or device is otherwise available, and the material has been previously proven to be applicable.
NOTE2. If metallic material uses the medical metallic material suitable for the application range of the national and industry standards, biological test may not be repeated.
4.6 Corrosion resistance (applicable to the non-disposable products)
The martensitic and austenitic stainless-steel materials are tested as per the boiling water test method specified in YY/T 0149-2006, which shall meet the requirements of Class-b; the remaining materials shall be performed once disinfection and sterilization method that is most unfavorable for the device specified by the manufacturer in the instruction manual, which shall meet the requirements of Class-b.
4.7 Disinfection and sterilization
4.7.1 Resistance of the re-disinfected and re-sterilized products
It shall meet the requirements of YY 0068.4-2009.
4.7.2 Sterile supply product
It shall meet the requirements of YY 0068.4-2009.
4.8 Package
It shall meet the requirements of YY 0068.4-2009.
4.9 Electrical safety performance
4.9.1 The electrical safety performance requirements of endoscope shall meet the requirements of YY 0068.4-2009.
4.9.2 The electrical safety performance requirements of surgical electrode shall meet the requirements of GB 9706.4.
5 Test Methods
5.5.4 Test on the dissolved precipitate of polymer material in contract with patient 5.5.4.1 Preparation of test solution. prepare according to the sixth method in Table 1 of GB/T 14233.1-2008.
5.5.4.3 Heavy metal. follow the method specified in 5.6 of GB/T 14233.1-2008.
5.5.4.4 Potassium permanganate reducing substance. follow the method specified in 5.2 of GB/T 14233.1-2008.
5.5.2 Material quality requirement test
The main structure of the polymer material shall be confirmed by verification data provided by the manufacturer.
Chemical composition test of the metallic material. take the method with accuracy greater than or reaching the tolerance or 1/3 limit value to carry out the chemical composition analysist test of the material.
5.5.3 Biocompatibility test
All tests are preferred to use the relevant standards of GB/T 16886 and perform the test as per their methods.
5.6 Corrosion resistance (applicable to the non-disposable products) test
The martensitic and austenitic stainless-steel materials shall be tested as per the boiling water test method in YY/T 0419-2006. The remaining materials shall be disinfected and sterilized once according the specified method in the instruction manual; make them cool off to the room temperature in the air; then check the corrosion marks on the product surface; the test on the corrosion degree shall be evaluated according to the 5.4 of YY/T 0149-2006.
5.7 Disinfection and sterilization test
5.7.1 Resistance test for the re-disinfected and re-sterilized products
Follow the method specified in YY 0068.4-2009; verify the conformity of 4.7.1 requirements.
5.7.2 Sterile supply product test
Fellow the method specified in YY 0068.4-2009; verify the conformity of 4.7.2 requirements.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 19 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.4Medical electrical equipment -- Part 2-2: Particular requirements for the safety 0f high frequency surgical equipment
- GB 9706.19Medical electrical equipment. Part 2: Particular requirements for the safety of endoscopic equipment
- GB/T 14233.1-2008Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- YY 0068.1-2008Medical Endoscopes—Rigid Endoscope - Part 1: Optical properties and test methods
- YY 0068.2-2008Medical endoscopes—Rigid endoscope - Part 2: Mechanical properties and test methods
Similar standards
Editions of YY/T 0619
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0619-2017 | Medical endoscopes - Rigid resectoscope | current edition | Current |
| YY 0619-2007 | Medical endoscopes - Rigid resectoscope | previous edition | Superseded |
This page sells the current edition, YY/T 0619-2017. Earlier editions are listed for reference only.
How to Download YY/T 0619-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.19-2000 — Medical electrical equipment. Part 2: Particular requirements for the safety of endoscopic equipment
GB 9706.4-2009 — Medical electrical equipment -- Part 2-2: Particular requirements for the safety 0f high frequency surgical equipment
GB/T 14233.1-2008 — Test Methods for Infusion, Transfusion, Injection Equipment for Medical Use — Part 1: Chemical Analysis Methods