YY 0068.4-2009Medical endoscopes - Rigid endoscopes - Part 4: Fundamental requirements (English PDF)
医用内窥镜 硬性内窥镜 第4部分:基本要求
Open the YY 0068.4-2009 preview as PDF
This is a limited preview
Buy now to download the full PDF (13 pages)
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
May 1, 2023
Implementation date
June 1, 2011
Scope
YY 0068.4-2009 is the English-translated version of 医用内窥镜 硬性内窥镜 第4部分:基本要求.
YY 0068.4 sets the fundamental requirements for rigid medical endoscopes, the laparoscopes, arthroscopes, cystoscopes and sinuscopes through which a surgeon sees inside the body. It is Part 4 of the Chinese series and carries the requirements that apply across the class: the mechanical construction and its integrity, the materials, the resistance to the repeated autoclaving these instruments undergo between patients, the sealing that keeps fluid out of the optical channel, and the marking. A rigid endoscope that admits moisture fogs internally and cannot be repaired in theatre, which turns a scheduled operation into a cancelled one. Being a YY standard without the /T suffix, compliance is compulsory in China. The standard carries Amendment 1 of 2023.
Document preview — YY 0068.4-2009
National Standard of the People's Republic of China
Issued by: National Medical Products Administration
Contents
- Foreword...3
- 1 Scope...4
- 2 Normative references...4
- 3 Terms and definitions...5
- 4 General provisions...5
- 5 Optical and mechanical properties...6
- 6 Electrical safety...6
- 7 Biocompatibility...7
- 8 Interface security...7
- 9 Manufacturing...8
- 10 Disinfection and sterilization...8
- 11 Packaging...9
- 12 Marking and instruction manual...9
- 13 Documentation...9
Foreword
YY 0068 "Medical endoscopes - Rigid endoscope" is divided into four parts. - Part
1 Scope
YY 0068.4 sets the fundamental requirements for rigid medical endoscopes, the laparoscopes, arthroscopes, cystoscopes and sinuscopes through which a surgeon sees inside the body. It is Part 4 of the Chinese series and carries the requirements that apply across the class: the mechanical construction and its integrity, the materials, the resistance to the repeated autoclaving these instruments undergo between patients, the sealing that keeps fluid out of the optical channel, and the marking. A rigid endoscope that admits moisture fogs internally and cannot be repaired in theatre, which turns a scheduled operation into a cancelled one. Being a YY standard without the /T suffix, compliance is compulsory in China. The standard carries Amendment 1 of 2023.
This part of YY 0068 specifies the fundamental requirements for the rigid endoscope for medical purposes.
2 Normative references
The provisions in following documents become the provisions of this part through reference in this part. For the dated references, the subsequent amendments (excluding corrections) or revisions do not apply to this part; however, parties who reach an agreement based on this part are encouraged to study if the latest versions of these documents are applicable. For undated references, the latest edition of the referenced document applies.
GB 9706.1-2007 Medical electrical equipment - Part
3 Terms and definitions
The terms and definitions in the other parts of YY 0068 apply to this part.
4 General provisions
The safety and performance of rigid endoscopes shall be subjected to both preclinical and clinical evaluations, including appropriate risk analysis in accordance with YY/T 0316.
5 Optical and mechanical properties
The manufacturer shall ensure that the rigid endoscope meets the requirements of YY 0068.1 and 0068.2.If the endoscope also has other properties, the manufacturer shall provide the requirements AND provide the appropriate test methods for the intended purpose of the performance.
6 Electrical safety
Hard endoscopes shall comply with the applicable requirements of GB 9706.19. For endoscope with eyepiece cover, if its eyepiece cover is claimed to be electrically isolated from the insertion part, the dielectric strength of the isolated part shall pass a test at a sinusoidal voltage of
50 Hz with a maximum current not exceeding
0.03 mA at 1500 V. Test method shall be as specified in
20.4 of GB 9706.1-2007.
7 Biocompatibility
Materials of the part of the rigid endoscope inserted into the human body shall be subject to the biosafety assessment in accordance with the principles and requirements of GB/T 16886.1 to demonstrate good biocompatibility.
8 Interface security
As for the rigid endoscope with a cable interface for illumination, particularly an endoscope with multiple interfaces other than an optical cable interface, in view of the possibility of other interfaces being incorrectly connected to the endoscope interface, the endoscope manufacturers shall implement the risk management procedures in accordance with YY/T 0316, in order to evaluate the physical possibility of incorrect connection to the endoscope interface, the likelihood of the operation occurring, and the severity of the potential harm to the patient.
9 Manufacturing
Endoscopes shall be produced in a manner that guarantees design characteristics.
10 Disinfection and sterilization
10.1 Endurance of repeatable disinfection or sterilization product For rigid endoscopes that can be repeatedly disinfected or sterilized, the method of disinfection or sterilization shall neither impair the functioning of the product NOR cause corrosion.
11 Packaging
Endoscopic packaging shall protect the endoscope from the effects of adverse storage and shipping conditions.
12 Marking and instruction manual
Rigid endoscope product markings and instruction manual shall comply with the provisions of other parts of YY 0068.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 13 pages — is available in the English PDF.
Referenced standards
Normative references
IEC 60601 · GB 9706.19-2000 · GB 11244-2005 · GB 18278-2000 · ISO 11134.1994 · GB 18279-2000 · ISO 11135.1994 · GB 18280-2000 · ISO 11137.1995
Similar standards
YY 0068.1|YY 0068.2|YY 0068.3|YY/T 0071
How to Buy YY 0068.4-2009
- 1Add to cart. Click the "Buy YY 0068.4-2009" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY 0017-2026 — Implants for osteosynthesis - Metallic bone plates
YY 0018-2026 — Implants for osteosynthesis - Metallic bone screws
YY 0118-2026 — Joint replacement implants - Hip joint prostheses - Fundamental requirements
Secure payment via Stripe
Payments accepted
YY 0068.4-2009
$60.00