YY 0068.1-2008Medical Endoscopes—Rigid Endoscope - Part 1: Optical properties and test methods (English PDF)
医用内窥镜 硬性内窥镜 第1部分:光学性能及测试方法
This is a limited preview
Buy now to download the full PDF (43 pages)
Need quotation for your management?
Get a formal quotation for YY 0068.1-2008 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
April 25, 2008
Implementation date
December 1, 2009
Scope
YY 0068.1-2008 is the English-translated version of 医用内窥镜 硬性内窥镜 第1部分:光学性能及测试方法.
YY 0068.1-2008 is the Chinese pharmaceutical industry standard that sets the optical property requirements and the test methods for rigid endoscopes used in medicine.
It is the first part of the YY 0068 series and covers the field of view and the direction of view, the angular magnification and the distortion, the resolving power and the depth of field, the light transmission and the illumination of the field, and the colour rendering, together with the way each of these is measured. Issued on 25 April 2008, it has applied since 1 December 2009.
Found what you need?
Document preview — YY 0068.1-2008
National Standard of the People's Republic of China
- Replacing
- YY 0068-1992
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- Introduction5
- 1 Scope8
- 2 Normative references8
- 3 Terms and definitions8
- 4 Requirements17
- 5 Test methods22
- Appendix A28
- Appendix B31
- Appendix C33
- Appendix D37
- Appendix E40
- Appendix F43
1 Scope
This Part of YY 0068 specifies the optical property requirements and test methods of the rigid endoscopes.
This Part applies to the rigid endoscopes for medical purposes.
2 Normative references
The provisions in the following documents become the provisions of this Part through reference in this Part. For dated references, the subsequent amendments (excluding corrections) or revisions do not apply to this Part.
However, parties who reach an agreement based on this Part are encouraged to study if the latest versions of these documents are applicable. For undated references, the latest versions apply to this Part.
GB/T 5702 Methods of measuring the color of light sources
GB 10810 Uncut finished spectacle lenses
GB 11244-2005 General requirements for the medical endoscope and endoscope accessories
JB/T 7403-1994 Illuminance meter
JB/T 9328-1999 Resolving power test target
ISO 10526:1999 CIE S 005 CIE standard illuminants for colorimetry
CIE 13.3-1995 Method of measuring and specifying color rendering properties of light sources
CIE 15.2004 Colorimetry
3 Terms and definitions
The following terms and definitions are applicable to this Part of YY 0068.
3.1 Endoscope
The endoscopes refer to the medical devices with the means of observation, which enter into the human body cavities via the natural orifices or the surgical incisions created for medical purposes.
3.2 Rigid endoscope
The rigid endoscopes refer to the endoscopes, of which the parts entering into the human bodies cannot curve along the natural orifices or along the created surgical incisions or along other instrument channels.
3.3 Rigid optical endoscope
Optical endoscope It refers to a rigid endoscope containing an optical transmission and illumination path as well as an optical imaging system. The external illumination light can be brought into the human body cavities via the optical illumination path. The human body cavities can conduct external imaging or direct visual observation via the imaging system.
3.4 Rigid optical observation endoscope
Optical observation endoscope The rigid optical observation endoscopes refer to the optical endoscopes whose imaging systems contain eyepieces. It is likely to conduct direct visual observation via the eyepieces.
3.5 Distal
It refers to the position on the endoscope - whose distance, relative to the user, is farther than that of other reference points.
SEE 3 in Figure 1.
3.6 Working length
It refers to the maximum length of the optical endoscope's part that is inserted into the human body.
4 Requirements
4.1 General
The optical endoscopes used for medical purposes shall conform to the following requirements. Other rigid endoscopes can be selected and used according to the applicability.
4.2 Field of view and direction of view
4.2.1 Field angle, 2W
The manufacturer shall express the design value of the field angle of the endoscope in any possible forms. The permissible error of the design value of the field angle shall be ± 15%.
4.3 Image-quality
4.3.1 Angular resolution, ra (d)
4.3.1.1 The manufacturer shall give the nominal value of the angular resolution in the center of the field of view, and the corresponding design working distance d in the accompanying files.
4.3.1.2 The permissible error of the nominal value of the angular resolution in the center of the field of view shall be -10% (excluding upper limit).
4.3.2 Range of the effective depth of field
If the manufacturer claims that the optical endoscope has depth of field, the optical endoscope's range of the effective depth of field shall be given in the accompanying files. Otherwise, the fact that there is no depth of field shall be claimed in the accompanying files.
4.3.3 Field quality
The field of view shall have no effects, such as double image or ghost, flashing, etc. It shall also have no defects, such as visible impurities, bubbles, etc.
4.4 Color resolution and reducibility
The optical observation endoscopes shall have good color resolution and reducibility. When using the spectra of the A and D65 standard illuminants specified in the ISO 10526:1999 CIE S 005, and outputting after transmitting via the illumination light path and imaging system, the optical observation endoscope's output spectrum shall still be able to maintain good color rendering properties. The manufacturer shall give the nominal value of this output spectrum's color rendering index Ra in the accompanying files, and shall also explain its meanings. The optical observation endoscope's measured value of the color rendering index under standard illuminants shall not be less than the nominal value.
4.5 Illumination
4.5.1 Illumination change rate
After the sterilization or disinfection test, the integral light energy transmittance of the optical endoscope's illumination light path shall maintain steady. USE output luminous flux for measurement. The change rate of the luminous flux shall not be greater than 20%.
Note: This Clause does not apply to the disposable optical endoscopes.
4.6 Synthetical relative effect of light energy
The manufacturer shall give the nominal value of the synthetical relative self-effect of light energy at Wp's 90% of the field of view in the accompanying files, and shall also explain its meanings. The measured value of this effect of light energy shall not be less than the nominal value.
4.7 Light energy transfer efficiency - Effective luminosity rate
EXPRESS the light energy transfer efficiency of optical endoscopes as effective luminosity rate. For the optical observation endoscopes, USE the symbol DM to express this effective luminosity rate; DEFINE it as the ratio OF the average luminous exitance in object plane TO the average illuminance in fundus image plane. For the non-visual optical observation endoscopes, DEFINE this effective luminosity rate as the discernible minimum luminosity in object plane (Lmin) of the critical display grayscale in image plane.
5 Test methods
5.6 Determination of the light energy transfer efficiency (expressed as effective luminosity rate)
5.1 Determination of the field of view and direction of view
SEE Appendix A (normative) of this Part.
5.2 Determination of the image-quality
5.2.1 Determination of the angular resolution
SEE Appendix B (normative) of this Part.
5.2.2 Determination of the effective depth of field
SELECT the worst working distance within the range of the effective depth of field for testing. TEST the angular resolution in the center of the field of view at this working distance according to the method specified in the Appendix B (normative) of this Part.
5.3 Determination of the color resolution and reducibility
SEE Appendix C (normative) of this Part.
5.4 Determination of the illumination property
5.4.1 Equipment
USE the following equipment.
5.4.3 Test for the illumination change rate
ADJUST the light intensity of the test light source to the intermediate value.
USE a light flux meter or other equivalent devices to measure the stable value of the luminous flux output by the test light resource via special optical fiber. DENOTE the stable value as the luminous flux Phi-i.
5.4.4 Illumination effectiveness test
USE the test light source to illuminate an integrating sphere. PASTE the cosine translucent material at the light inlet of the optical endoscope's illumination light path. ENTER the small hole on the integrating sphere. The diameter of this small hole shall not be greater than 10% of the diameter of the integrating sphere.
5.5 Determination of the synthetical relative effect of light energy
SEE Appendix D (normative) of this Part.
5.6.1 Equipment
USE the following equipment.
5.6.2 Environmental conditions
The environmental conditions are as follows. […]
5.6.3 Procedures
5.6.3.1 PLACE the cosine radiation screen perpendicular to the endoscope's axis of view. ADJUST the endoscope via axial movement so that the screen area is greater than the area of the observation field of view.
5.6.3.2 For non-visual optical observation endoscopes.
5.6.4 Calculation
CALCULATE according to the Equation (13).
5.7 Determination of the unit relative distortion
SEE Appendix E (normative) of this Part. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 43 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 5702Method of measuring the color rendering properties of light sources
- GB 11244-2005General requirements for the medical endoscope and endoscope accessories
- JB/T 7403-1994Illuminance meters
- JB/T 9328-1999Resolving power test target
GB 10810 · ISO 10526:1999
Cited by
- YY/T 1028-2023Medical endoscopes - Fibre endoscopes
- YY/T 1587-2018Medical endoscopes—Video endoscopes
- SN/T 0869-2017Determination of ascorbic acid in beverage for export
- YY/T 0619-2017Medical endoscopes - Rigid resectoscope
- YY 1298-2016Medical endoscopes—Capsule endoscopes
- YY/T 1298-2016Medical endoscopes - Capsule endoscopes
Similar standards
Editions of YY 0068.1
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 0068.1-2008 | Medical Endoscopes—Rigid Endoscope - Part 1: Optical properties and test methods | current edition | Current |
| YY 0068-1992 | Medical Endoscopes—Rigid Endoscope - Part 1: Optical properties and test methods | previous edition | Superseded |
This page sells the current edition, YY 0068.1-2008. Earlier editions are listed for reference only.
How to Download YY 0068.1-2008
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 11244-2005 — General requirements for the medical endoscope and endoscope accessories
GB/T 5702-2019 — Method of measuring the color rendering properties of light sources
JB/T 7403-1994 — Illuminance meters
Secure payment via Stripe
Payments accepted
YY 0068.1-2008
$135.00