YY/T 0521-2018Dentistry—Implants—Dynamic fatigue test for endosseous dental implants (English PDF)
牙科学 种植体 骨内牙种植体动态疲劳试验
This is a limited preview
Buy now to download the full PDF (24 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0521-2018 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
April 11, 2018
Implementation date
May 1, 2019
Scope
YY/T 0521-2018 is the English-translated version of 牙科学 种植体 骨内牙种植体动态疲劳试验.
YY/T 0521-2018 is the Chinese pharmaceutical industry standard that gives the dynamic fatigue test for endosseous dental implants and their premanufactured prosthetic components.
The implant is loaded in a way that reproduces the worst case of function in the mouth, and the standard fixes the specimen, the embedding, the angle and point of load application, the frequency and the number of cycles and the way the result is reported, which makes it useful above all for comparing designs and sizes. Issued on 11 April 2018, it has applied since 1 May 2019.
Found what you need?
Document preview — YY/T 0521-2018
National Standard of the People's Republic of China
- Classification
- C 33
- Replacing
- YY/T 0521-2009
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
This standard replaces YY/T 0521-2009 "Dental fatigue test of dental implants in dentistry". With YY/T 0521-2009 In addition to the editorial changes, the main technical changes are as follows.
--- Added 3.3 bone implant implant components;
--- Added 4.2 requirements "In the method recommended by the manufacturer, the tightening torque should be noted";
--- Added 4.2 requirements "If the manufacturer does not indicate the tightening torque in the manual, the appropriate tightening torque should be determined before the test";
--- In 5.3.1, change "the modulus of elasticity should be greater than 3GPa" to "the modulus of elasticity (compression modulus of elasticity) should be greater than 3GPa";
--- In 5.3.1, the "intra-bone fixation of the specimen should be fixed with a rigid fixture" should be changed to "the internal fixation of the specimen should be used Rigid fixture fixing (it is recommended to embed the specimen in the embedding material and then use the rigid fixture to fix it)";
--- Change the "test can be carried out in air at 20 °C ± 5 °C" in 5.4 to "For other planting systems, the test can be Conducted in air at a temperature of 20 ° C ± 10 ° C";
--- Cancel the "determination of the critical failure point and locate the location of the damage origination" in 5.6.3 to "describe the failure mode of the affected part, such as If possible, describe the failure process (eg, screw fracture and subsequent abutment fracture);
--- Increase 5.7 "Alternative method --- ladder method";
--- Added Appendix B "Worst Case Determination Guide".
1 Scope
This standard specifies the fatigue test method for wearing a mucosal single-bone internal dental implant and its pre-repair assembly. This method is different in comparison It is especially useful when designing different sizes of intraosseous implants.
Although this standard simulates the functional load of the main body of the endodontic implant and the pre-repaired component under "worst case" conditions, it is not applicable.
To predict the in vivo performance of endodontic implants or restorations, especially for multi-pile implant restorations.
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only dated versions apply to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB/T 3075-2008 Metallic materials fatigue test axial force control method (ISO 1099:2006, MOD)
GB/T 9937-2008 Oral vocabulary (all) (ISO 1942. 1989, IDT)
GB/T 16825.1-2008 Inspection of static single-axis testing machines - Part 1. Examination of tensile force and/or pressure testing machine Inspection and calibration (ISO 7500-1:2004, IDT)
ISO 4965-1:2012 Dynamic force calibration for uniaxial fatigue testing of metallic materials - Part 1 Metallic materials - Axial load fatigue testing machines - Dynamic force calibration - Part 2. Dynamic calibration equipment (DCD) Test instrument
3 Terms and definitions
The following terms and definitions as defined in GB/T 9937-2008 apply to this document.
3.1 Endosenous dental implant system
A device consisting of multiple components, including implant preparation and placement of implants in the clinical and laboratory, and corresponding restorations Auxiliary tools and special equipment necessary for seating and seating.
Note 1. In addition to resisting the displacement of the dental restoration, the endodontic implant can also be used as an orthodontic anchorage.
Note 2. The endodontic implant can be composed of one or more parts.
Note 3. The concept of dental restorations includes crowns and fixed prostheses, active restorations.
3.2 Load cycle diagram
A curve showing the fatigue performance of the implants in the bone by showing the number of cycles that each sample is subjected to under each peak load. Figure (see Appendix A).
3.3 Endosenous dental implant assembly
The individual components of the implant are assembled together according to the manufacturer's instructions.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 3075-2008Metallic materials - Fatigue testing - Axial force-controlled method
- GB/T 9937-2008Dentistry - Vocabulary
- GB/T 16825.1-2008Metallic materials - Calibration and verification of static uniaxial testing machines - Part 1: Tension/compression testing machines - Calibration and verification of the force-measuring system
ISO 4965-1:2012
Similar standards
Editions of YY/T 0521
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0521-2018 | Dentistry—Implants—Dynamic fatigue test for endosseous dental implants | current edition | Current |
| YY/T 0521-2009 | Dentistry—Implants—Dynamic fatigue test for endosseous dental implants | previous edition | Superseded |
This page sells the current edition, YY/T 0521-2018. Earlier editions are listed for reference only.
How to Download YY/T 0521-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
YY/T 0492-2026 — Implantable leads
Secure payment via Stripe
Payments accepted
YY/T 0521-2018
$205.00