Scope
YY/T 0110-2009 is the English-translated version of 医用超声压电陶瓷材料.
YY/T 0110-2009 is the Chinese pharmaceutical industry standard for the piezoelectric ceramic materials used in medical ultrasound.
It specifies their classification, the requirements, the test methods and the inspection rules, covering the dielectric and the piezoelectric constants, the electromechanical coupling coefficient, the mechanical quality factor, the Curie temperature, the density and the stability of these properties with time and temperature. Issued on 15 November 2009, it has applied since 1 December 2010.
Found what you need?
Document preview — YY/T 0110-2009
National Standard of the People's Republic of China
- ICS
- 11.040.50
- Classification
- C 41
- Replacing
- YY/T 0110-1993
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms, definitions and symbols GB/T
- 5 Requirements
Foreword
Instead of the standard YY/T 0110-1993 "medical ultrasound piezoelectric ceramic material." The main changes in the standard of YY/T 0110-1993 compared as follows.
a) According to the latest national standards updated normative references.
b) According to the development of medical ultrasound piezoelectric ceramic material, although a lot of new materials to meet the old standard of freedom relative permittivity to Seeking, but piezoelectric strain constant, the mechanical quality factor and other parameters of the old standards do not match, especially in the lead titanate is defined as the power The rate of emission material less accurate. Therefore, according to GB/T 3388 provisions of the material features, combined with medical ultrasonic piezoelectric ceramic material Material characteristics of medical ultrasonic piezoelectric ceramic materials have been reclassified.
c) main performance requirements for medical ultrasonic piezoelectric ceramic material, increasing the longitudinal electromechanical coupling coefficient k 33, the longitudinal length of the stretching vibration Frequency constant N3 and the longitudinal sound velocity Vp 3D, thickness of the sound velocity Vp tD. error.
e) Amendment of the inspection rules, according to medical ultrasonic piezoelectric ceramics production characteristics, test methods for materials added to the sample system For, taking into account the small number of samples, so the test acceptance criteria were direct statements.
This standard by the standardization of medical ultrasound equipment for medical appliances National Standardization Technical Committee Technical Committee (SAC/TC10/SC2) Centralized.
This standard is mainly drafted by. Kunshan Risheng Electronics Co., Ltd., the State Food and Drug Administration of Quality Supervision, Inspection Hubei Medical Devices Test center.
1 Scope
This standard specifies the medical ultrasonic piezoelectric ceramic material classification, requirements, test methods and inspection rules.
This standard applies to medical ultrasonic piezoelectric ceramic material.
2 Normative references
The following criteria are contained in the terms of this standard reference and become the standard terms. For cited documents with dates, with its After all the modifications (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement Research on whether to use the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.
GB/T 2413-1981 piezoelectric ceramic material bulk density measurement method
GB/T 2414.1-1998 Test methods for properties of piezoelectric ceramics discs radially stretching vibration mode
GB/T 3388-2002 piezoelectric ceramics type designation
GB/T 3389.1-1996 ferroelectric ceramics vocabulary Test GB/T 3389-2008 properties of piezoelectric ceramics test method performance parameters
GB/T 6427-1999 piezoelectric ceramic oscillator frequency temperature stability test methods
GB/T 11310-1989 properties of piezoelectric ceramics test methods relative permittivity of free test temperature characteristics
GB/T 15750-2008 properties of piezoelectric ceramics test method aging test
3 Terms, definitions
and symbols GB/T terms, definitions and symbols established 3389.1 apply to this standard. Category 4 Materials
4.1 Classification of the materials
According to GB/T 3388 provisions of the material features, combining the characteristics of medical ultrasonic piezoelectric ceramic material, this standard specifies nine kinds of piezoelectric Ceramic material.
4.1.1 P-42
in the power transmission type, and can be used to send and receive use either.
4.1.2 P-43
in the power transmission type.
4.1.4 PT-71-type
high-frequency, dual-purpose transceiver. Generally used to make high-frequency probe unit.
4.1.5 P-51
high-sensitivity, commonly used for signal reception to be used along with small-signal transceiver, typically used to make the probe unit.
4.1.6 P-52
high-sensitivity, commonly used for signal reception to be used along with small-signal transceiver, typically used to make the probe unit; free phase epsilonr3T of permittivity greater than the P-51.
4.1.7 P-53
high-sensitivity, commonly used for signal reception to be used along with small-signal transceiver, typically used to make low-frequency linear array transducer; from epsilonr3T by the relative permittivity greater than P-52.
4.1.8 P-54
high-sensitivity, commonly used for signal reception to be used along with small-signal transceiver, typically used to make linear or convex array probe.
4.1.9 P-55
high-sensitivity, commonly used for signal reception to be used along with small-signal transceiver, particularly suitable for the production of high frequency and high-density probe.
5 Requirements
The main performance should meet the requirements in Table 1.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 2414.1-1998Test methods for the properties of piezoelectric ceramics--Radial extension vibration mode for disk
- GB/T 3388-2002Designations for types of piezoelectric ceramics
- GB/T 3389.1-1996Vocabulary for ferroelectric and piezoelectric ceramics
- GB/T 6427-1999Test method for frequency temperature stability of piezoelectric ceramic vibrator
- GB/T 11310-1989Test methods for the properties of piezoelectric ceramics. Test methods for temperature characteristics of relative free dielectric constants
- GB/T 15750-2008Test methods for the properties of piezoelectric ceramics - Test for the ageing properties
GB/T 2413-1981
Similar standards
Editions of YY/T 0110
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0110-2009 | Piezoelectric materials for medical ultrasound | current edition | Current |
| YY/T 0110-1993 | Piezoelectric materials for medical ultrasound | previous edition | Superseded |
This page sells the current edition, YY/T 0110-2009. Earlier editions are listed for reference only.
How to Download YY/T 0110-2009
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 11310-1989 — Test methods for the properties of piezoelectric ceramics. Test methods for temperature characteristics of relative free dielectric constants
GB/T 15750-2008 — Test methods for the properties of piezoelectric ceramics - Test for the ageing properties
GB/T 2414.1-1998 — Test methods for the properties of piezoelectric ceramics--Radial extension vibration mode for disk