Valid

YY 0948-2015Cardiopulmonary bypass systems—Arteriovenous cannula for single use (English PDF)

心肺转流系统一次性使用动静脉插管

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for YY 0948-2015 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Mandatory

Issue date

March 2, 2015

Implementation date

January 1, 2017

Scope

YY 0948-2015 is the English-translated version of 心肺转流系统一次性使用动静脉插管.

YY 0948-2015 is the Chinese standard "Cardiopulmonary bypass systems. Arteriovenous cannula for single use".

Its scope clause reads: This standard specifies the classification and structure, requirements, testing methods, marks, labels, instructions for use, packaging, transportation and storage of aseptic disposable arteriovenous cannula (hereinafter referred to as Arteriovenous cannula).

This standard is suitable for arteriovenous cannula supporting cardiopulmonary bypass system. It is used for drainage or perfusion of blood during direct vision cardiac operation of cardiopulmonary bypass (CPB).

Its clauses include testing method; marks, labels, instructions for use. It was issued by the China Food and Drug administration on MARCH 02, 2015 and took effect on JANUARY 01, 2017.

Found what you need?

Document preview — YY 0948-2015

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Classification and structure5
  • 4 Requirements5
  • 5 Testing method8
  • 6 Marks, labels, instructions for use11
  • 7 Packaging, transportation, storage12

1 Scope

This standard specifies the classification and structure, requirements, testing methods, marks, labels, instructions for use, packaging, transportation and storage of aseptic disposable arteriovenous cannula (hereinafter referred to as Arteriovenous cannula).

This standard is suitable for arteriovenous cannula supporting cardiopulmonary bypass system. It is used for drainage or perfusion of blood during direct vision cardiac operation of cardiopulmonary bypass (CPB).

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 191-2008 Packaging - Pictorial marking for handling of goods

GB/T 9969-2008 General principles for preparation of instructions for use of industrial products

GB/T 14233.1-2008 Test methods for infusion, transfusion, injection equipment for medical use - Part 1: Chemical analysis methods

GB/T 14233.2-2005 Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods

GB/T 16886.1 Biological evaluation of medical devices - Part 1 - Evaluation and testing within a risk management process

GB 18279 Medical devices - Validation and routine control of ethylene oxide sterilization

GB 18280 Sterilization of health care products - Requirement for validation and routine control

4 Requirements

4.2 Physical performance

4.2.1 No leakage

There shall be no leakage in the connection of arteriovenous cannula.

4.2.2 Connection strength

The connections of arteriovenous cannula shall be firm.

4.2.3 Temperature adaptability

The arteriovenous cannula shall not be deformed or broken in the temperature range of 0°C ~ 50°C.

4.2.4 Bending resistance

When the venous cannula is bent, the inner wall shall not be adhered.

Note: This test is not required for arterial cannula.

4.3 Biological performance

4.3.1 Biological evaluation

The arteriovenous cannula shall have no biological hazards.

4.3.2 Sterile

The arteriovenous cannula shall go through confirmed sterilization process to ensure product sterility.

4.3.3 No pyrogen

The arteriovenous cannula shall be no pyrogen.

4.4 Chemical performance

4.4.1 Reducing substance

The difference between the volumes of the potassium permanganate solution [c(KMnO4) = 0.002mol/L] consumed by the 20mL test solution and the same batch of blank control solution shall not exceed 2.0mL.

4.4.2 Heavy metals

5 Testing method

5.1 Appearance

Use visual inspection. Requirements of 4.1 shall be met.

5.2 Physical performance

5.2.1 Leakage test

Connect artery cannula channel, seal all outlets, inject air or nitrogen gas which is 50kPa higher than atmospheric pressure, put it into water, keep for 10 min, observe carefully whether there are bubbles escaping from each junction. It shall meet the specifications in 4.2.1.

Fill water in venous cannula, seal all outlets, connect to vacuum pump, keep the pressure 20 kPa below the atmospheric pressure, maintain for 10 min, observe carefully whether there are bubbles entering the venous cannula at each junction. It shall meet the specifications in 4.2.1.

Note: In order to prevent water from entering the vacuum pump, a pipeline without injected water can be used to connect the sample with the machine.

5.2.2 Connection strength test

Apply 15 N axial static tension to each of the junctions of arteriovenous cannula and keep for 15 s. They shall not be separated and shall conform to the specifications in 4.2.2.

5.2.3 Temperature adaptability test

Place the arteriovenous cannula in the environment of 0°C for 3 min, then put in the environment of 50°C for 3 h, and then take back to room temperature for observation. Perform the leakage test of 5.2.1 and the requirements in 4.2.3 shall be met.

5.2.4 Bending resistance test

The length of the test shall be 200 mm. If the length of venous cannula is less than 200 mm, the test length shall be the whole length of the product. Bend the venous cannula as shown in figure 2 and remain for 10 s, visually check it in the bent state. The requirements in 4.2.4 shall be met.

Take products with expiration of no more than 1 month (arbitration, preferred), or age according to YY/T 0681.1-2009. Test 4.2, 4.3.2, 4.3.3, and the results shall conform to 4.7.

6 Marks, labels, instructions for use

6.1 Mark

6.1.1 Product individual package

The individual packaging of each product shall be provided with the following marks.

a) The name of the manufacturer;

b) Product name, model specification;

c) Production batch number, validity period;

d) Words or marks such as "One-time use", "Aseptic", " Prohibit of use in case of packaging breakage", "Please read the instructions before use".

6.1.2 Product external package

The product external packaging shall be provided with the following marks.

a) The name and address of the manufacturer;

b) Product name, model specification;

c) Implementation of the standard number;

d) Product registration number;

e) Production license number;

f) Production batch number;

g) The words or marks of "one-time use" etc.;

h) Sterilization methods; I) The period of validity;

j) Packaging quantity, volume (length×width×height);

k) The words or marks such as "careful and light", "do not press heavily", "fear of moisture", etc., shall conform to the relevant provisions in GB/T […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY 0948-2015

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 0948-2015

$125.00