YY 1271-2016Cardiopulmonary bypass systems—Suction catheter for single use (English PDF)
心肺转流系统一次性使用吸引管
This is a limited preview
Buy now to download the full PDF (16 pages)
Need quotation for your management?
Get a formal quotation for YY 1271-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
March 23, 2016
Implementation date
January 1, 2018
Scope
YY 1271-2016 is the English-translated version of 心肺转流系统一次性使用吸引管.
YY 1271-2016 is the Chinese pharmaceutical industry standard for the sterile single-use suction catheter of a cardiopulmonary bypass system, through which blood is aspirated from the operative field.
It specifies the classification and structure, the requirements, the test methods, the markings, the labels, the instructions for use and the rules for packaging, transport and storage, covering the dimensions and the eyelets, the flow and the resistance, the strength of the tube and of its connections, the surface and the freedom from particles, the materials and their biological evaluation and sterility. Issued on 23 March 2016, it applies from 1 January 2018.
Found what you need?
Document preview — YY 1271-2016
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Classification and structure5
- 4 Requirements6
- 5 Test methods8
- 6 Marking, instruction manual11
- 7 Packaging, transport, storage12
1 Scope
This standard specifies the classification and structure, requirements, test methods, markings, labels, instructions for use, packaging, transportation, and storage of sterile suction catheter for single use (hereinafter referred to as suction catheter).
This standard applies to suction catheters, for supporting cardiopulmonary bypass system, suction catheter for cardiovascular ventilation in the left heart, suction decompression or reduction of left heart load, to attract the intracardiac blood and other fluids.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard.
GB/T 191-2008 Packaging - Pictorial marking for handling of goods
GB/T 9969-2008 General principles for preparation of instructions for use of industrial products
GB/T 14233.1-2008 Test methods for infusion transfusion injection equipment for medical use - Part 1. Chemical analysis methods
GB/T 14233.2-2005 Test methods for infusion, transfusion, injection equipment for medical use - Part 2. Biological test methods
GB/T 16886.1 Biological evaluation of medical devices - Part 1. Evaluation and testing within a risk management process
GB 18279 Medical devices - Validation and routine control of ethylene oxide sterilization
4 Requirements
4.2.1 Leakage-free
Suction catheter shall be no leakage free.
4.2.2 Connection strength
Suction catheter shall be firmly connected.
4.2.3 Temperature adaptability
The suction catheter shall be free from deformation and rupture in the temperature range of 0 °C ~ 50 °C.
4.2.4 Negative pressure tolerance
Under room temperature conditions, when a negative pressure of 20 kPa (150 mmHg) is applied for 15 seconds, the suction catheter body shall not be flattened.
4.3 Biological properties
4.3.1 Biological evaluation
Suction catheters shall be free of biological hazards.
4.3.2 Sterility
The suction catheter shall be sterilized by the sterilization process that has been confirmed.
4.3.3 Pyrogen-free
The suction catheter shall be pyrogen free.
4.4 Chemical properties
4.4.1 Reducing substances (easily oxides)
The difference between the 20 mL of test solution and the volume of potassium permanganate solution [c(KMnO4) = 0.002 mol/L] consumed by the same batch of blank control solution shall not exceed 2.0 mL.
4.4.2 Heavy metals
When measured by atomic absorption spectrophotometry (AAS) or equivalent method, the total content of bismuth, chromium, copper, lead, and tin in the test solution shall not exceed 1 µg/mL. The content of cadmium shall not exceed 0.1 µg/mL.
When it is subject to visual observation, the suction catheter shall meet the requirements of 4.1.
5 Test methods
5.2 Physical properties
5.2.1 Leakage-free
CONNECT the suction catheter passage, LEAD in air or nitrogen gas at a pressure of 50 kPa above atmospheric pressure, PLACE it in water, MAINTAIN it for 10 min, there shall be free from air bubble escape under careful observation. It shall comply with the requirements of 4.2.1.
5.2.2 Connection strength
APPLY the axial static tensile force of 15 N at each joint of the suction catheter for 15 s. It shall be free from separation. It shall comply with the requirements of 4.2.2.
5.2.3 Temperature adaptability
PLACE the suction catheter in a 0°C environment for 3 minutes, then PLACE it in a 50°C environment for 3 hours. After removal, RESTORE it to room temperature for observation, MAKE it subject to the leakage free test of 5.2.1, it shall comply with the requirements of 4.2.3.
5.2.4 Negative pressure tolerance
CONNECT the suction end of the suction catheter to a vacuum source, BLOCK the suction hole, APPLY a negative pressure of 20 kPa for 15 seconds at room temperature. The suction catheter body shall not be flattened, it shall comply with the provisions of 4.2.4.
5.3 Biological properties
5.3.1 Biological evaluation
The biological properties shall be evaluated in accordance with the provisions of GB/T 16886.1.
5.3.2 Sterility test
The sterilization process shall be confirmed in accordance with relevant standards such as GB 18279 or GB 18280.
Sterility test is performed in accordance with the provisions of GB/T 14233.2-2005, the method should not be used for exit-factory inspection.
5.5 Corrosion resistance
Take out the metal parts, make it subject test in accordance with the sodium chloride solution test method in YY/T 0149-2006, the results shall comply with the provisions of 4.5.
5.6 Particle contamination
The test shall be conducted in accordance with the provisions in Appendix A of GB 19335-2003, it shall comply with the provisions of 4.6.
5.7 Validity period test
Take the products that have expired not exceeding one month (arbitration method, which is preferred), or perform aging in accordance with the provisions of YY/T 0681.1-2009, to test the items of 4.2, 4.3.2, 4.3.3, the results shall comply with the requirements of 4.7.
6 Marking, instruction manual
6.1 Marking
6.1.1 The product packaging shall have the following markings.
a) The name of the manufacturer;
b) The product name, model specifications;
c) The production lot number, validity period;
d) Words or illustrations such as "for single-use", "sterility", "do not use if package broken", "please read manual before use", etc.
6.1.2 Product packaging shall have the following markings.
a) Manufacturer's name, address;
b) Product name, model specifications;
c) Implemented standard number;
d) Product registration number;
e) Production license number;
f) Production lot number;
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191-2008Graphical symbols marking for handling and storage of packages
- GB/T 9969-2008General principles for preparation of instructions for use of industrial products
- GB/T 14233.1-2008Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods
- GB/T 14233.2-2005Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- GB 18279Sterilization of health-care products—Ethylene oxide—Requirements for the development,validation and routine control of a sterilization process for medical devices
Similar standards
How to Download YY 1271-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 18279.1-2015 — Sterilization of health care products—Ethylene oxide—Part 1: Requirements for development,validation and routine control of a sterilization process for medical devices
GB 18279-2023 — Sterilization of health-care products—Ethylene oxide—Requirements for the development,validation and routine control of a sterilization process for medical devices
GB/T 14233.1-2008 — Test Methods for Infusion, Transfusion, Injection Equipment for Medical Use — Part 1: Chemical Analysis Methods