YY 0585.2-2019Fluid lines for use with pressure infusion equipment and accessories for single use - Part 2: Accessories (English PDF)
压力输液设备用一次性使用液路及附件 第2部分:附件
This is a limited preview
Buy now to download the full PDF (25 pages)
Need quotation for your management?
Get a formal quotation for YY 0585.2-2019 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
May 31, 2019
Implementation date
December 1, 2020
Scope
YY 0585.2-2019 is the English-translated version of 压力输液设备用一次性使用液路及附件 第2部分:附件.
YY 0585.2-2019 is the Chinese pharmaceutical industry standard that sets the physical, chemical and biological requirements for the accessories used with pressure infusion equipment, such as the connectors, stopcocks and extension pieces that complete the fluid line.
It is the second part of the YY 0585 series and covers the dimensions and the compatibility with the line, the resistance to the working pressure and the tightness, the strength of the joints, the extractables and the biological evaluation, and sterility, together with the labelling and the information supplied. Issued on 31 May 2019, it applies from 1 December 2020.
Found what you need?
Document preview — YY 0585.2-2019
National Standard of the People's Republic of China
- Replacing
- YY 0585.2-2005
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Materials6
- 4 Physical requirements6
- 5 Chemical requirements7
- 6 Biological requirements7
- 7 Packaging7
- 8 Labelling7
- 9 Disposal9
- Appendix A (Normative) Physical test10
- Appendix B (Normative) Chemical test12
- Appendix C (Normative) Biological test13
- Appendix D (Informative) Technical differences between this part and ISO 8536-10. 2015 and their reasons14
- References16
Foreword
YY 0585 "Fluid lines for use with pressure infusion equipment and accessories for single use" includes the following parts.
--- Part 1: Fluid lines;
--- Part 2: Accessories;
--- Part 3: Filters;
--- Part 4: Check valves. This part is Part 2 of YY 0585.
This part is drafted in accordance with the rules given in GB/T 1.1-2009.
This part replaces YY 0585.2-2005. Compared with YY 0585.2-2005, appendix D is added after appendix A, appendix B, and appendix C, and the main technical changes except editorial changes are as follows.
--- The degree of standard adoption is changed from "IDT" (identical) to "MOD" (modified);
--- Chapter 3 "Marking" is deleted;
--- The requirements for Unit with Injection Site in 4.8 are modified;
--- The requirements for Unit with Check Valve in 4.9 are modified, and A.8 is deleted;
--- Chapter 6 "Biological requirements" is modified to "GB 8368 is applicable" for ease of implementation;
--- The requirements for labelling in Chapter 8 are amended, and the requirement that the symbol 2725 of ISO 7000 can be used to indicate the presence of a substance of concern is added;
--- Chapter 9 "Disposal" is added;
--- The leakage test method in A.4 is modified, among which the negative pressure test is deleted;
--- The leakage test for adapters with female and/or male conical fittings in A.5 is deleted.
This part is modified in relation to ISO 8536-10:2015 "Infusion equipment for medical use - Part 10: Accessories for fluid lines for single use with pressure infusion equipment".
There are technical differences between this part and ISO 8536-10:2015.A list of the corresponding technical differences and their reasons is given in Appendix D. The following editorial change is made in this part.
--- Appendix D (informative) is added, which gives a list of the corresponding technical differences compared with ISO 8536-10:2015 and their reasons.
1 Scope
This part of YY 0585 specifies the physical, chemical, and biological requirements for accessories used with pressure infusion equipment.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) is applicable to this standard.
GB/T 1962.21) Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings (GB/T 1962.2-2001, ISO 594-2:1998, IDT)
3 Materials
The materials from which accessories are manufactured shall meet the requirements specified in Chapter 4, Chapter 5, and Chapter 6.
4 Physical requirements
The accessories shall be impermeable to air, microorganisms and fluid. No matter where the plug is located, it shall be sealed. When the test is carried out according to A.4, there shall be no air or water leakage.
5 Chemical requirements
GB 8368 is applicable, and the test method is shown in Appendix B.
6 Biological requirements
GB 8368 is applicable, and the test method is shown in Appendix C.
7 Packaging
GB 8368 is applicable.
8 Labelling
The shelf or multi-unit package shall be labelled with at least the following information.
9 Disposal
Information on the safety and environmentally biosafety disposal of single-use infusion sets shall be given in accompanying documents.
Appendix A (Normative) Physical test
Fill the accessory to be tested with distilled water according to the usual operation method, and visually check whether there are air bubbles in the fluid pathway of the transparent components.
Appendix B (Normative) Chemical test
Take components with a total area of 100 cm².Disassemble the components that are in contact with the fluid line in the sterilized and ready-to-use accessory, and arrange them according to the same material.
Appendix C (Normative) Biological test
GB 8368 is applicable.
Appendix D (Informative)
Technical differences between this part and ISO 8536-10:2015 and their reasons
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.
Referenced standards
Normative references
GB/T 1962.21
Cited by
Similar standards
Editions of YY 0585.2
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 0585.2-2019 | Fluid lines for use with pressure infusion equipment and accessories for single use - Part 2: Accessories | current edition | Current |
| YY 0585.2-2005 | Fluid lines for use with pressure infusion equipment and accessories for single use - Part 2: Accessories | previous edition | Superseded |
This page sells the current edition, YY 0585.2-2019. Earlier editions are listed for reference only.
How to Download YY 0585.2-2019
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY 0585.3-2018 — Fluid lines for use with pressure infusion equipment and accessories for single use - Part 3: Filters
YY 0585.1-2005 — Fluid lines for use with pressure infusion equipment and accessories for single use - Part 1: Fluid lines