Scope

YY 0341.2-2020 is the English-translated version of 无源外科植入物 骨接合与脊柱植入物 第2部分:脊柱植入物特殊要求.

YY 0341.2-2020 is the Chinese pharmaceutical industry standard that sets the particular requirements for non-active surgical spinal implants, the screws, rods, hooks, plates and interbody devices with which the spine is fixed.

It is the second part of the YY 0341 series and applies in addition to the general requirements of YY/T 0640, giving the definitions, the requirements, the test methods and the rules for manufacture, sterilization, packaging and the information the manufacturer provides, covering the materials, the dimensions and the surface, the static and fatigue strength of the construct and the marking. Issued on 27 September 2020, it applies from 1 June 2022.

Found what you need?

Document preview — YY 0341.2-2020

National Standard of the People's Republic of China

Replacing
YY 0341-2009

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Terms and definitions6
  • 4 Requirements6
  • 5 Test method8
  • 6 Manufacturing11
  • 7 Sterilization11
  • 8 Packaging11
  • 9 Information provided by the manufacturer11
  • Appendix A (Informative) Relevant material standards acceptable for clinical use13
  • Appendix B (Informative) List of approved method standards for chemical analysis15
  • Appendix C (Informative) Relevant standards for design evaluation and testing16

1 Scope

This Part of YY 0341 specifies the special requirements for non-active surgical spinal implants (hereinafter referred to as "spinal implants"). In addition to the requirements that are specified in YY/T 0640, it also specifies the definition, requirements, test methods, manufacturing, sterilization, packaging and information provided by the manufacturer of spinal implants.

This Part of YY 0341 applies to non-active surgical spinal implants other than artificial intervertebral disc implants.

2 Normative references

The following documents are indispensable for the application of this document.

For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.

GB/T 4340.1, Metallic materials - Vickers hardness test - Part 1: Test method

GB/T 10610, Geometrical product specifications (GPS) - Surface texture:

Profile method - Rules and procedures for the assessment of surface texture

GB/T 14233.2, Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods

GB/T 16886.7, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals

GB 23101.2, Implants for surgery - Hydroxyapatite - Part 2: Coatings of hydroxyapatite

YY/T 0343, Liquid penetrant inspection of metallic surgical implants

YY/T 0640-2016, Non-active surgical implants - General requirements

Note: Some material standards for spinal implants that have been proven acceptable through clinical applications can be found in the informative list that is given in Appendix A.

4 Requirements

4.2 Hardness

The hardness of metal spinal implants shall meet the requirements of the corresponding product standards.

4.3 Static and/or dynamic mechanical properties

The manufacturer shall combine the characteristics of the spinal implant and its clinical use to specify its static and/or dynamic mechanical properties. If there are relevant product standards, the mechanical properties of spinal implants shall meet the requirements of the corresponding standards. Due to the different characteristics of spinal implants, the test standards may not yet exist or the existing test standards may be modified as needed.

4.4 Corrosion resistance

The pitting potential (Eb) on the surface of the final product of the stainless-steel spinal implant shall not be less than 800 mV.

4.5 Surface quality

4.5.1 Surface defects

The surface of the non-coated metal spinal implant shall not have any discontinuity defects.

4.5.2 Surface roughness

The manufacturer shall specify the roughness Ralpha value of the main surface of the spinal implant.

If there are relevant product standards, the roughness Ralpha value of spinal implants shall meet the requirements of the corresponding standards.

4.5.3 Appearance

The surface of the spinal implant shall be free of defects that affect the function of the implant, as well as inlays, final processing deposits or other contaminants.

Note: Defects that affect the function include oxide skin, knife marks, small nicks, scratches, cracks, depressions, sharp edges (except for cutting edges), burrs and other defects.

4.5.4 Anodized surface treatment

implants, as well as the physical and chemical properties of polymer materials, are tested according to the method that is specified in the material standard or the recommended method in Appendix B.

Note: If the final product cannot be sampled due to structure or size, samples of the same batch of raw materials and the same manufacturing process can also be used. […]

5 Test method

5.9.1 The sterility test shall be carried out according to the method that is specified in GB/T 14233.2.

Note: Other validated methods are also acceptable.

5.9.2 The residual amount of ethylene oxide shall be carried out according to the method that is specified in GB/T 16886.7.

Note: Other validated methods are also acceptable.

5.2 Hardness

It is performed according to the method of GB/T 4340.1.

Note: Other validated methods are also acceptable.

5.3 Static and/or dynamic mechanical properties

When using static and/or dynamic load tests to evaluate spinal implants, existing inspection standards can be used (see Appendix C for related information), or a custom test model that considers the characteristics of the implant can be used; the quantity of the samples required for tests shall be determined according to the selected test method standard.

Note 1: The test method can be selected according to different test levels.

Note 2: Preclinical evaluation shall consider the existing data of similar implants or design characteristics.

Note 3: Tests can be performed on specific implants under specific loads and/or specific environmental conditions to evaluate their performance.

5.4 Corrosion resistance

It is performed according to the method that is specified in YY/T 1074. The number of samples shall be no less than 3.

Note: If the final product cannot be sampled due to structure or size, samples of the same batch of raw materials and the same manufacturing process can also be used.

5.5 Surface quality

5.5.1 Surface defects

It is performed according to the method that is specified in YY/T 0343. The number of samples shall be no less than 3.

6 Manufacturing

It shall meet the requirements of Chapter 8 in YY/T 0640-2016.

7 Sterilization

It shall meet the requirements of Chapter 9 in YY/T 0640-2016.

8 Packaging

It shall meet the requirements of Chapter 10 in YY/T 0640-2016.

9 Information provided by the manufacturer

9.1 General principles

Chapter 11 of YY/T 0640-2016 and the following specific terms apply.

9.2 Instruction manual

It shall comply with the provisions of 11.3 in YY/T 0640-2016; the manufacturer should also provide the following information:

Restrictions on any modification after delivery of the implant: size, shape, surface condition.

9.3 Marking

9.3.1 The 11.5, 11.6 in YY/T 0640-2016 or the following terms apply.

9.3.2 The spinal implant's marking consists of material code, manufacturer name code, production batch number, etc. The material code is as specified in […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY 0341.2

EditionTitleRevisionStatus
YY 0341.2-2020Non-active surgical implants—Osteosynthesis and spinal implants - Part 2: Particular requirements for spinal implantscurrent editionCurrent
YY 0341-2009Non-active surgical implants—Osteosynthesis and spinal implants - Part 2: Particular requirements for spinal implantsprevious editionSuperseded

This page sells the current edition, YY 0341.2-2020. Earlier editions are listed for reference only.

How to Download YY 0341.2-2020

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
25 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 0341.2-2020

$195.00