GB/Z 40842-2021General technical requirenments for electrolarynx (English PDF)
电子人工喉通用技术要求
Open the GB/Z 40842-2021 preview as PDF
This is a limited preview
Buy now to download the full PDF (16 pages)
Issued by
State Administration for Market Regulation; Standardization Administration of China
Level / Type
National · Recommended
Issue date
October 11, 2021
Implementation date
October 11, 2021
Scope
GB/Z 40842-2021 is the English-translated version of 电子人工喉通用技术要求.
GB/Z 40842-2021 is the Chinese national standard on general technical requirenments for electrolarynx, in the field of metrology and measurement. The /Z suffix marks it as a guiding technical document: it does not prescribe requirements that can be certified against, but sets out the technique, the method or the state of the art that the standards bodies recommend following. It was issued on 11 October 2021 by the State Administration for Market Regulation; Standardization Administration of China. As a guiding technical document it carries no separate date of entry into force: it applies from publication. Classification: ICS 17.140.50, CCS L13. This page is published from the official record of the standard held by the Chinese standards administration: the identification, the dates, the classification and the issuing body are taken from there. The clause text, the tables and the numeric limits are in the document itself, which is delivered complete in English translation.
Document preview — GB/Z 40842-2021
National Standard of the People's Republic of China
- ICS
- 17.140.50
- Classification
- L13
Issued by: State Administration for Market Regulation; Standardization Administration of China
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Technical requirements...
- 4.6 Biological evaluation
- 5 Inspection methods
- 5.2 Performance test method
- 5.2.1 Sound pressure level
- 5.6 Biological evaluation test
- 6 Inspection rules...
- 7 Marks and labels
- 8 Inspection report...
1 Scope
GB/Z 40842-2021 is the Chinese national standard on general technical requirenments for electrolarynx, in the field of metrology and measurement. The /Z suffix marks it as a guiding technical document: it does not prescribe requirements that can be certified against, but sets out the technique, the method or the state of the art that the standards bodies recommend following. It was issued on 11 October 2021 by the State Administration for Market Regulation; Standardization Administration of China. As a guiding technical document it carries no separate date of entry into force: it applies from publication. Classification: ICS 17.140.50, CCS L13. This page is published from the official record of the standard held by the Chinese standards administration: the identification, the dates, the classification and the issuing body are taken from there. The clause text, the tables and the numeric limits are in the document itself, which is delivered complete in English translation.
This document specifies the technical requirements, test methods, inspection rules, marks, labels, packaging, transportation, and storage of the electrolarynx. This document applies to the production, use, and distribution of the electrolarynx. It is suitable for patients who have received laryngectomy and lost vocal function due to diseases, accidents, and other reasons.
2 Normative references
The following documents contain the provisions which, through normative reference in this document, constitute the essential provisions of this document. For the dated referenced documents, only the versions with the indicated dates are applicable to this document; for the undated referenced documents, only the latest version (including all the amendments) is applicable to this document.
GB/T 191 Packaging - Pictorial marking for handling of goods
GB 9706.1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
GB/T 10111 Generation of random numbers and procedures applied to sampling inspection for product quality
GB/T 14710 The environmental requirement and test methods for medical electrical equipment
GB/T 16886.5-2017 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
GB/T 16886.10-2017 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
YY 0505 Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 electrolarynx A voice recovery device powered by a battery.
Note: It can transmit the vibrating sound signal to the pharynx through the patient's neck tissue, and the signal is modulated by the upper part of the vocal tract to form an electronic artificial laryngeal voice at the lip end; it belongs to the speech compensation rehabilitation aids.
3.2 external electrolarynx A voice recovery device whose vibration device is located outside the body, and the vibration signal is transmitted to the oral cavity through the neck tissue or the sound transmission tube.
3.3 Internal electrolarynx A voice recovery device whose vibration device is located in the body, and the vibration signal directly forms in the mouth.
3.4 vibrating membrane A component of an electrolarynx, which can be pressurized by electromechanical control and cause vibration and sound.
3.5 volume adjuster A device that can adjust the sound intensity of an electrolarynx.
3.6 pitch adjuster A device that can adjust the vibration frequency of a vibrating membrane.
3.7 laser measuring micro-displacement instrument An instrument that can directly measure the vibration frequency of a vibrating membrane.
3.8 vibration measuring instrument An instrument that uses mechanical means, optical means, vibration sensors, etc. to measure the vibration amplitude, acceleration, speed, and frequency of objects.
Note: The vibration measurement accuracy should be less than 100 µm, and the measurement range
4.3.2 After the test sample is subjected to the button life test according to the method specified in 5.3.2, the fundamental frequency of the sample shall not exceed ±10% of the fundamental frequency before the test.
4.4 Packaging requirements The test sample (with packaging) shall be able to work normally after being tested according to the method in 5.4.
4.5 Environmental adaptability The environmental test of the electrolarynx shall meet the requirements of GB/T 14710.
4 Technical requirements...
4.6 Biological evaluation
4.6.1 In vitro cytotoxicity Cytotoxicity shall be less than grade 2.
4.6.2 Skin irritation The primary irritation score shall be less than 1.
4.6.3 Delayed-type hypersensitivity reaction There shall be no delayed-type hypersensitivity.
4.7 Electrical safety and electromagnetic compatibility The test results shall meet the requirements of GB 9706.1 and YY 0505.
5 Inspection methods
5.1 Inspection methods for appearance and structure The method can be visual inspection and touch detection, which shall meet the requirements of 4.1.
5.2.1 Sound pressure level
5.2.1.1 Install the test sample on the fixture and place the sound level meter at a distance of 1000 mm in front of it (see Figure 1).
5.2.1.2 Apply a certain contact pressure to the control switch of the test sample to generate sound.
5.3.1 Service life test method Install the test sample on the fixture, so that it can simulate the button action in the normal working mode; press once every 5 s and hold for 3 s; operate intermittently, and control the button with a load value of greater than 30 N; the "press + release" counts as once, and repeat it for 3×104 times. After the test, the sample shall meet the requirements of 4.3.1.
5.3.2 Button life test method Use different samples of the same batch for the test; after turning off the power supply or taking out the battery, install the test sample on the fixture, and load the button pressure of greater than 30 N cyclically; the frequency shall be less than or equal to
1 Hz, and the number of tests shall be greater than or equal to 1×105 times. After the test, the sample shall meet the requirements of 4.3.2.
5.4 Packaging test Drop the packaged test sample freely onto the cement floor from a height of 1000 mm; carry out 2 times tests for each of the 6 surfaces. After the test, the sample shall meet the requirements of 4.4.
5.5 Environmental test The test shall be carried out according to the relevant provisions in GB/T 14710, and the results shall meet the requirements of 4.5.
5.6 Biological evaluation test
5.6.1 In vitro cytotoxicity Take the cell complete medium as the extraction medium, according to the ratio of
0.2 g/mL, and extract for 24 h under the condition of (37±1) °C; the test shall be carried out according to the liquid extraction method specified in GB/T 16886.5-2016. The results shall meet the requirements of 4.6.1.
5.6.2 Skin irritation The test shall be carried out in accordance with the primary skin irritation method specified in GB/T 16886.10-2017, and the results shall meet the requirements of 4.6.2.
5.6.3 Delayed-type hypersensitivity reaction Take normal saline and vegetable oil as the extraction medium respectively, according to the ratio of 0.2g/mL, and extract for (72±2) h under the condition of (37±1) °C; the test shall be carried out according to the maximum dose method specified in GB/T 16886.10-2017, and the results shall meet the requirements of 4.6.3. If any performance required by electrical safety does not qualify, such as the protectiveness to continuity of the earthed circuit, insulation diagram, and withstand voltage performance, or if two or more of the other performances do not qualify, then the type inspection is judged as failed; if one of the other performances does not qualify, it is allowed to recheck multiple selected prototypes for the re-inspection of the unqualified item (if there is only one prototype, it is allowed to be repaired); if the result is still unqualified, then the type inspection is judged as failed.
7 Marks and labels
7.1 The following information shall be marked in the appropriate position of an electrolarynx:
a) Manufacturer's name, trademark;
b) Product name and model;
c) Product number or date of manufacture;
d) Supply voltage;
e) The fundamental frequency range over which the product operates;
f) The medical device registration certificate number or filing document number.
7.2 There shall be a certificate of quality in the packing box, and the certificate of quality shall contain the following contents:
a) Name of the manufacturer;
b) Product name and model;
c) Date of inspection;
d) Product number;
e) Inspector code;
f) The word "Qualified".
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB 9706.1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 10111Generation of random numbers and procedures applied to sampling inspection for product quality
- GB/T 14710The environmental requirement and test methods for medical electrical equipment
- GB/T 16886.10-2017Biological evaluation of medical devices - Part 10: Tests for skin sensitization
GB/T 16886.5-2017 · YY 0505
Similar standards
Editions of GB/Z 40842
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/Z 40842-2021 | General technical requirenments for electrolarynx | current edition | Current |
This page sells the current edition, GB/Z 40842-2021. Earlier editions are listed for reference only.
How to Buy GB/Z 40842-2021
- 1Add to cart. Click the "Buy GB/Z 40842-2021" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
GB/T 10111-2008 — Generation of random numbers and procedures applied to sampling inspection for product quality
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
Secure payment via Stripe
Payments accepted
GB/Z 40842-2021
$170.00