GB/T 9247-2008Microscopes - Condensers (English PDF)
显微镜 聚光镜
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 9247-2008 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
June 20, 2008
Implementation date
January 1, 2009
Scope
GB/T 9247-2008 is the English-translated version of 显微镜 聚光镜.
GB/T 9247-2008 is the Chinese national standard for microscope condensers, and applies to condensers for transmitted-light observation on upright microscopes. It fixes the classification and basic parameters, the requirements, the test methods and the marking.
Its table 1 sets out the series - low-magnification condensers, two-lens and three-lens Abbe condensers, achromatic-aplanatic and dark-field condensers, dry and immersion - each with its numerical aperture, the range of objective magnifications it suits and whether it carries a variable aperture diaphragm; the dovetail size for the non-plug-in mounting is 37f8.
The requirements limit the shift of the front group lens axis in the focal plane to 0.02 mm, and this edition adds a manufacturing tolerance on numerical aperture and checks condenser stress with a polarizing microscope.
It replaces GB/T 9247-1996, was issued on 20 June 2008 and has applied since 1 January 2009.
Found what you need?
Document preview — GB/T 9247-2008
National Standard of the People's Republic of China
- ICS
- 37.020
- Classification
- N 32
- Replacing
- GB/T 9247-1996
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Classification and basic parameters
- 4 Requirements
Foreword
This standard is the result on the basis of GB/T 9247-1996 "microscope condenser" on amending the Supplementary made.
This standard and GB/T 9247-1996 The main differences are as follows.
--- This release adds the numerical aperture of the condenser manufacturing variations of provisions and a corresponding increase in the numerical aperture of the condenser measurement procedures;
--- This version of the 1996 version of the test used "special microscope frame" are replaced by "and the subject of a microscope equipped with a condenser frame";
Condenser stress --- for a polarizing microscope examination instead stress tester checks;
--- Cleaning requirements on quality and appearance of the condenser optics are clearly specific quality requirements. Appendix A of this standard is an informative annex. The standard proposed by China Machinery Industry Federation.
1 Scope
This standard specifies the classification microscope condenser and the basic parameters, requirements, test methods and marking.
This standard applies to upright microscope condenser transmitted light observations.
2 Normative references
The following documents contain provisions which, through reference in this standard and become the standard terms. For dated references, subsequent Amendments (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement are based on research Whether the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.
GB/T 15464 instrumentation packaging technology
3 Classification and basic parameters
3.1 Condenser series and basic parameters according to Table 1
Type the numerical aperture Apply objective Magnification range Variable aperture stop Low magnification with condenser >=0.16 1 ~ 4 - Low magnification with condenser >=0.4 2 ~ 40 - Abbe two condenser lens >=0.85 4 ~ 100 have Dark field condenser The inside of the 0.75 to 0.85 The inside of the outer >= +0.10 10 ~ 40 - Immersion Abbe three condenser lens >=1.2 4 ~ 100 have Achromatic/aplanatic condenser >=1.3 10 ~ 100 have Dark field condenser Medial 1.25 The inside of the outer >= +0.15 10 ~ 100-- 3.2 Non-plug-in connection dovetail size of the condenser is removable Ø37f8.
4 Requirements
When the swing frame front group before 4.1 times put into the condenser lens group, the lens axis of the front group in the focal plane of the condenser lens shift should not exceed 0.02mm.
4.2 condenser aperture stop should be the central axis substantially coincident.
4.3 Variable aperture stop should be basically round, adjusting lightweight and comfortable, the smallest diameter should be less than Ø2mm.
NA 4.4 condenser deviation shall be in accordance with Table 2.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
GB/T 15464
Similar standards
Editions of GB/T 9247
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 9247-2008 | Microscopes - Condensers | current edition | Current |
| GB/T 9247-1996 | Microscopes - Condensers | previous edition | Superseded |
This page sells the current edition, GB/T 9247-2008. Earlier editions are listed for reference only.
How to Download GB/T 9247-2008
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 10003-2026 — Biaxially oriented polypropylene (BOPP) film
GB/T 10004-2026 — Plastic/Plastic laminated films and pouches for packaging
GB/T 10007-2026 — Rigid cellular plastics - Determination of shear properties
Secure payment via Stripe
Payments accepted
GB/T 9247-2008
$135.00