GB/T 40291-2021Nuclear instrumentation - High-purity germanium crystals for radiation detectors - Measurement methods of basic characteristics (English PDF)
Also coversGBT40291-2021
Open the GB/T 40291-2021 preview as PDF
This is a limited preview
Buy now to download the full PDF
Issued by
State Administration for Market Regulation, China National Standardization Administration
Level / Type
National · Recommended
Issue date
May 21, 2021
Implementation date
December 1, 2021
Scope
GB/T 40291-2021 (Nuclear instrumentation - High-purity germanium crystals for radiation detectors - Measurement methods of basic characteristics) is available as an English-translated PDF.
GB/T 40291-2021 — This document describes the terminology and test methods for measuring the basic characteristics of high-purity germanium crystals. Including the net concentration of electroactive impurities (NA-ND), Deep-level impurity center concentration and crystallographic characteristics. This document applies to high-purity germanium crystals for gamma-ray and X-ray radiation detectors. The net concentration of the electroactive impurity center of this germanium single crystal Less than 1011cm-3, usually on the order of 1010cm-3. The test methods listed in this document are not legal, but they are widely used in the industry and provide a The verified information satisfies the manufacturer's needs. In GB/T 7167-2008 and GB/T 11685-2003, a complete test method for assembled germanium detectors is given.
Document preview — GB/T 40291-2021
National Standard of the People's Republic of China
- Classification
- F 88
Issued by: State Administration for Market Regulation, China National Standardization Administration
Contents
- 1 Scope1
- 2 Normative references1
- 3 Terms, definitions and symbols1
- 3.1 Terms and definitions1
- 3.2 Symbols, abbreviations, quantities and units2
- 3.3 Quantity and unit4
- 4 Net electroactive impurities (NA-ND) measurement4
- 4.1 Sample preparation of Van der Pauw method4
- 4.2 Measuring the Net Concentration of Electroactive Impurities (NA-ND)6
- 5 Determination of impurity center concentration by deep-level transient spectroscopy technique11
- 5.1 Overview11
- 5.2 Deep-level transient spectroscopy equipment11
- 5.3 Sample selection and preparation for deep-level transient spectrum11
- 5.4 Method for measuring impurity center concentration12
- 5.5 P-type high purity germanium majority carrier deep energy level15
- 5.6 Deep energy levels of majority carriers in n-type high purity germanium16
- 5.7 Report16
- 6 Crystallographic properties16
- 6.1 Overview16
- 6.2 Crystal orientation17
- 6.3 Sample preparation17
- 6.4 Report18
- Appendix A (informative) Hall factor of p-type and n-type high purity germanium19
Foreword
This document is in accordance with the provisions of GB/T 1.1-2020 "Guidelines for Standardization Work Part 1.Structure and Drafting Rules of Standardization Documents"
Drafting.
The translation method used in this document is equivalent to IEC 61435.2013 "Basic characteristics of high-purity germanium crystals for nuclear instrumentation radiation detectors"
Methods of Measuring Sexuality.
This document has made the following editorial changes.
---Correct the incorrect content in the "Commonly Used Symbol Definition Comparison Table" in the original text of the international standard, including. the definitions of the symbols "f" and "F" are mutually exclusive
Change, the symbol "K" is changed to "k", the symbol "P" is changed to "rho", the symbol "T" [time, unit is second (s)] is changed to "t", the symbol is deleted
"()" and its definition, the average value is represented by the upper horizontal line (example taur2).
This document was proposed and managed by the National Standardization Technical Committee for Nuclear Instruments (SAC/TC30).
Drafting organizations of this document. Nuclear Industry Aerial Survey and Remote Sensing Center, Shenzhen Institute of Metrology and Quality Inspection.
Introduction
The detector manufacturer needs the specific parameters of the crystal to evaluate the characteristics of the coaxial detector. Since the physical characteristics are changed, only pass
The manufacturer's measurement results cannot deduce a complete detector performance index. This document defines the basic parameters used to determine crystals (such as electrical activity)
The method of measuring the net concentration of sexual impurities, the concentration of deep-level impurity centers and the crystallographic characteristics of crystals.
When high purity germanium (HPGe) detectors are used to detect ionizing radiation, the size and purity of the germanium crystals need to be determined. Due to the high resistance of the material
(When the temperature is 77K, the resistivity is about 10kohm.cm), and the degree of impurity compensation causes special problems in the production of germanium crystals.
And how to reasonably describe the distribution of large-volume crystal impurities in a single device, etc., the existing standards do not cover.
One of the most important characteristics of high purity germanium is the net electroactive impurity concentration (NA-ND), which determines the depletion voltage when the detector is working.
The van der Pauw method indicating n-type or p-type is usually used to determine (NA-ND), and the transfer measurement is carried out on the sheet samples immersed in liquid nitrogen. Pick
When using this technique, (NA-ND) can be calculated by resistivity or Hall coefficient. Through a series of measurements on the sample to get the resistivity or
Hall coefficient.
High-purity germanium crystal for nuclear instrument and radiation detector
Measurement methods of basic characteristics
1 Scope
This document describes the terminology and test methods for measuring the basic characteristics of high-purity germanium crystals. Including the net concentration of electroactive impurities (NA-ND),
Deep-level impurity center concentration and crystallographic characteristics.
This document applies to high-purity germanium crystals for gamma-ray and X-ray radiation detectors. The net concentration of the electroactive impurity center of this germanium single crystal
Less than 1011cm-3, usually on the order of 1010cm-3.
The test methods listed in this document are not legal, but they are widely used in the industry and provide a
The verified information satisfies the manufacturer's needs.
In GB/T 7167-2008 and GB/T 11685-2003, a complete test method for assembled germanium detectors is given.
2 Normative references
The contents of the following documents constitute the indispensable clauses of this document through normative references in the text. Among them, dated quotations
Only the version corresponding to that date is applicable to this document; for undated reference documents, the latest version (including all amendments) is applicable to
This document.
GB/T 2900.66-2004 Electrotechnical terminology for semiconductor devices and integrated circuits (IEC 60050-521.2002, IDT)
GB/T 2900.97-2016 Electrotechnical terminology nuclear instrument. physical phenomena, basic concepts, instruments, systems, equipment and detectors
(IEC 60050-395.2014, IDT)1)
1) The original international standards IEC 60050-393.2003 and IEC 60050-394.2007 have been replaced by IEC 60050-395.2014.
3 Terms, definitions and symbols
3.1 Terms and definitions
The following terms and definitions apply to this document.
3.1.1
Semiconductor
The total conductivity caused by the two carriers is usually a material between the conductor and the insulator. The carrier concentration in this material varies with the external
Ministry conditions change and change.
Note. The term semiconductor generally applies to materials whose carriers are electrons or holes.
[Source. GB/T 2900.66-2004,521-02-01]
3.1.2
High-purity semiconductor detector
Semiconductor detectors using high-purity (such as high resistivity) semiconductor materials.
[Source. GB/T 2900.97-2016,395-03-48]
3.1.3
Hall effect
The electric field intensity generated in conductors and semiconductors that is proportional to the vector product of current density and magnetic induction.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.
Referenced standards
Normative references
GB/T 2900.66-2004 · IEC 60050 · GB/T 2900.97-2016
How to Buy GB/T 40291-2021
- 1Add to cart. Click the "Buy GB/T 40291-2021" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 46830.2-2026 — Plastics - Liquid crystal polymer (LCP) moulding and extrusion materials - Part 2: Preparation of test specimens and determination of properties
GB/T 28593-2026 — Sand and dust weather warning levels
GB/T 19608.3-2026 — Classification of special environmental conditions - Part 3: Plateau
Secure payment via Stripe
Payments accepted
GB/T 40291-2021
$395.00