GB/T 35894-2018Determination of 10 forbidden glycol ethers and esters in cosmetics—GC-MS (English PDF)
化妆品中10种禁用二元醇醚及其酯类化合物的测定 气相色谱-质谱法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 35894-2018 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
February 6, 2018
Implementation date
September 1, 2018
Scope
GB/T 35894-2018 is the English-translated version of 化妆品中10种禁用二元醇醚及其酯类化合物的测定 气相色谱-质谱法.
GB/T 35894-2018 is the Chinese national gas chromatography-mass spectrometry method for determining ten glycol ethers and their esters that are banned in cosmetics, among them 2-methoxyethanol, 2-ethoxyethanol, ethylene glycol dimethyl ether and triethylene glycol dimethyl ether.
The sample is extracted ultrasonically with an ethyl acetate-methanol mixed solvent, analysed by GC-MS in selected ion monitoring mode on a VF-WAX column and quantified by the external standard method, with separate treatment for waxy and non-waxy samples.
The standard covers reagents, instruments, sample treatment, blank test, qualitative confirmation, result calculation, recovery and precision, and the allowable deviation between two results.
Annexes give detection and quantitation limits, compound data and selected ions. It is used by cosmetics testing laboratories. It was issued on 6 February 2018 and has applied since 1 September 2018.
Found what you need?
Document preview — GB/T 35894-2018
National Standard of the People's Republic of China
- ICS
- 71.100.70
- Classification
- Y 42
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- Introduction4
- 1 Scope5
- 2 Principle5
- 3 Reagents and Materials5
- 4 Reagents and Materials6
- 5 Analytical Procedures6
- 6 Result Calculation8
- 7 Recovery Rate and Precision9
- 8 Allowable Deviation9
- Appendix A (Informative) Detection Limit and Quantitation Limit of This Method10
- Appendix B (Informative) Chinese Name, English Name, CAS No., Molecular Formula, Relative Molecular Mass and Chemical Structural Formula of Forbidden Glycol Ethers and Esters10
- Appendix C (Informative) Quantitative and Qualitative Selected Ion Parameters of 10 Forbidden Glycol Ethers and Esters12
- Appendix D (Informative) Gas Chromatography-Mass Spectrometry Selected Ion Chromatogram of Standard Substances of 10 Forbidden Glycol Ethers and Esters13
1 Scope
This Standard specifies reagents and materials, instruments and equipment, analytical procedures, result calculation, recovery rate and precision, and allowable deviation of gas chromatography-mass spectrometry method of determining 10 forbidden glycol ethers (ethylene glycol dimethyl ether, 2-methoxy ethanol, 2-methoxy propanol, 2- ethoxyethanol, 2-methoxy propyl acetate, 2-methoxyethyl acetate, 2-ethoxy ethyl acetate, dimethoxy diglycol, 2-(2-methoxye-thoxy) ethanol, triethylene glycol dimethyl ether) and their ester compounds in cosmetics.
This Standard is applicable to the determination of the content of 10 forbidden glycol ethers and their ester compounds in cosmetics.
Please refer to Table A.1 in Appendix A for the detection limit and quantitation limit of the method in this Standard.
2 Principle
Sample goes through ultrasonic extraction with ethyl acetate-methanol mixed solvent.
Adopt gas chromatograph-mass spectrometer for determination; quantify through the external standard method.
3 Reagents and Materials
Unless it is otherwise indicated, all reagents being adopted shall be analytical pure.
3.1 Ethyl acetate. chromatographic purity.
3.2 Methanol. chromatographic purity.
3.3 Anhydrous sodium sulfate.
3.4 20% ethyl acetate methanol solution (V/V, volume ratio). measure-take 200 mL of ethyl acetate (3.1), then, mix it up with 800 mL of methanol (3.2).
3.5 Standard samples/substances of 10 forbidden glycol ethers and esters. purity shall be not less than 98%.
5 Analytical Procedures
5.2 Determination
Since the determination result depends on the instrument being adopted, it is impossible to provide general parameters of gas chromatography-mass spectrometry analysis. The set parameters shall guarantee that the tested components and the other components can be effectively separated in chromatographic determination. The reference conditions of gas chromatography-mass spectrometry are as follows.
a) Chromatographic column. VF-WAX column, 60 m x 0.25 mm (i.d.) x 0.25 µm, or other equivalents;
b) Inlet temperature. 220 °C;
c) Transmission line temperature. 280 °C;
d) Inlet mode. split injection, split ratio. 5.1;
e) Carrier gas. helium, purity >= 99.999%; mode of control. constant flow; flow rate. 1.0 mL/min;
f) Column temperature. initial temperature. 50 °C, maintain for 7 min; rise to 140 °C at 5 °C/min, maintain for 1 min; rise to 240 °C at 20 °C/min, maintain for 2 min. Post-treatment temperature. 245 °C, duration. 3 min.
g) Inlet volume. 1 µL.
h) Ionization source. EI source;
i) Ionization energy. 70 eV;
j) Quadrupole temperature. 150 °C;
k) Ionization source temperature. 230 °C;
l) Solvent delay time. 7.8 min;
m) Mode of monitoring. select ion scanning mode (SIM); please refer to Table C.1 in Appendix C for monitoring ions.
5.3 Blank Test
Other than sample-weighing, comply with procedure 5.1 and procedure 5.2.
5.4 Qualitative Confirmation
Under the instrument conditions in 5.2, the selected ion chromatographic peak of the tested sample and the standard solution emerges in the same retention time […]
6 Result Calculation
The calculation result shall retain 3 significant figures.
7 Recovery Rate and Precision
Non-waxy based sample shall be within the fortified concentration range of 2.0 mg/kg ~ 30 mg/kg; within the recovery rate of 80% ~ 110%; its relative standard deviation shall be less than 10%.
Waxy-based sample shall be within the fortified concentration range of 3.0 mg/kg ~ 60 mg/kg; within the recovery rate of 80% ~ 110%; its relative standard deviation shall be less than 10%.
8 Allowable Deviation
The absolute difference of two independent determination results obtained under repeatability conditions shall not exceed 10% of the arithmetic mean value. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Similar standards
How to Download GB/T 35894-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 10003-2026 — Biaxially oriented polypropylene (BOPP) film
GB/T 10004-2026 — Plastic/Plastic laminated films and pouches for packaging
GB/T 10007-2026 — Rigid cellular plastics - Determination of shear properties
GB/T 35894-2018
$105.00