GB/T 33865-2017Calibration method for photosynthetic active radiometer (English PDF)
光合有效辐射表校准方法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 33865-2017 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
July 12, 2017
Implementation date
February 1, 2018
Scope
GB/T 33865-2017 is the English-translated version of 光合有效辐射表校准方法.
GB/T 33865-2017 gives the method for calibrating a photosynthetic active radiometer, the hemispherical instrument that measures the solar radiation between 400 nm and 700 nm arriving from the 2 pi solid angle above a given plane.
It fixes the calibration conditions: an open site with nothing above the sensor, a clear sky with the sun at least 30° high, between 10:00 and 14:00, an air temperature of 10 °C to 30 °C, relative humidity no higher than 80 % and wind speed no higher than 5 m/s.
The reference radiometer must keep its uncertainty within 6 %, its cosine and azimuth response errors within 10 % and 5 % and its stability within 3 %. The calibration method, the uncertainty of the result and the recalibration interval follow. Issued on 12 July 2017, it has applied since 1 February 2018.
Found what you need?
Document preview — GB/T 33865-2017
National Standard of the People's Republic of China
- ICS
- 07.060
- Classification
- A 47
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Preface
- 1 Scope
- 2 terms and definitions
- 3 Calibration conditions
- 4 calibration method
- 5 Uncertainty assessment of calibration results
- 6 calibration results
- 7 re - school time interval
- Appendix A (Normative Appendix) Spectral Radiometer Specifications
- Appendix B (normative) Spectral radiometer method
- Appendix C (informative) Evaluation of Uncertainty in Calibration Results of Photosynthetically Effective Radiation Table
- Reference
Foreword
This standard is drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the calibration conditions for the photosynthetically active radiation table, the calibration method, and the uncertainty of the calibration results.
This standard is applicable to the calibration of the sensitivity of the hemisphere to the photosynthetically active radiation table.
2 Terms and definitions
The following terms and definitions apply to this document.
2.1 Photosynthetic active radiation table; hemispherical photosynthetic active radiometer
Measure the radiant table of the total solar radiation from 400 nm to 700 nm received from the upper 2pi solid angle of the given plane.
3 Calibration conditions
3.1 Environmental conditions
3.1.1 around the open, the instrument sensor surface above no obstacles.
3.1.2 sunny sky, the sun height angle of not less than 30 °, should be in place 10:00 to 14:00 between.
3.1.3 air temperature in the range of 10 °C ~ 30 °C, the relative humidity of not more than 80%, wind speed is not greater than 5m/s.
3.2 Standard and ancillary equipment
3.2.1 Standard Photosynthetic Effective Radiation Table
Should meet the following requirements.
a) The uncertainty should not be greater than 6%;
b) cosine response error (zenith angle 0 ° ~ 80 °) should not be greater than 10%;
c) azimuth response error (zenith angle 0 ° ~ 70 °) should not be greater than 5%;
d) the temperature error should not be greater than 0.3%/°C;
e) Stability should not be greater than 3%.
3.2.2 Digital instrumentation
0.05, resolution 1µV.
3.2.3 Environmental measuring instruments
Technical indicators in Table 1.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Similar standards
How to Download GB/T 33865-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 10003-2026 — Biaxially oriented polypropylene (BOPP) film
GB/T 10004-2026 — Plastic/Plastic laminated films and pouches for packaging
GB/T 10007-2026 — Rigid cellular plastics - Determination of shear properties
GB/T 33865-2017
$205.00