GB/T 12417.1-2024Non-active surgical implants - Osteosynthesis and joint replacement implants - Part 1: Particular requirements for osteosynthesis implants (English PDF)
This is a limited preview
Buy now to download the full PDF (11 pages)
Need quotation for your management?
Get a formal quotation for GB/T 12417.1-2024 in 30 minutes.
Issued by
State Administration for Market Regulation
Level / Type
National · Recommended
Issue date
December 31, 2024
Implementation date
January 1, 2026
Scope
GB/T 12417.1-2024 (Non-active surgical implants - Osteosynthesis and joint replacement implants - Part 1: Particular requirements for osteosynthesis implants) is available as an English-translated PDF.
GB/T 12417.1-2024 is the Chinese standard "Non-active surgical implants - Osteosynthesis and joint replacement implants - Part 1: Particular requirements for osteosynthesis implants".
Its scope clause reads: This document specifies particular requirements for non-active surgical implants for osteosynthesis (hereinafter referred to as the “implants”).
This document provides particular requirements regarding intended performance, design attributes, materials, design evaluation, manufacturing, sterilization, packaging, and information supplied by the manufacturer.
Its clauses include terms and definitions; intended performance; design attributes; materials; design evaluation; manufacturing; sterilization; packaging; information supplied by the manufacturer.
It was issued by the State Administration for Market Regulation on 2024-12-31 and took effect on 2026-01-01.
Found what you need?
Document preview — GB/T 12417.1-2024
National Standard of the People's Republic of China
- ICS
- 11.040.40
Issued by: State Administration for Market Regulation
Contents
- Foreword...3
- Introduction...5
- 1 Scope...6
- 2 Normative References...6
- 3 Terms and Definitions...6
- 4 Intended Performance...6
- 5 Design Attributes...9
- 6 Materials...9
- 7 Design Evaluation...9
- 8 Manufacturing...10
- 9 Sterilization...11
- 10 Packaging...11
- 11 Information Supplied by the Manufacturer...11
1 Scope
This document specifies particular requirements for non-active surgical implants for osteosynthesis (hereinafter referred to as the “implants”).
This document provides particular requirements regarding intended performance, design attributes, materials, design evaluation, manufacturing, sterilization, packaging, and information supplied by the manufacturer.
2 Normative References
The contents of the following documents constitute indispensable clauses of this document through normative references in the text. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.
YY/T 0640-2016 Non-active surgical implants - General requirements (ISO 14630.2012, IDT)
3 Terms and Definitions
The terms and definitions defined in YY/T 0640-2016, as well as the following terms and definitions, are applicable to this document.
3.1 Non-Active Surgical Implants for Osteosynthesis
Non-active surgical implants intended to provide support for bone, cartilage, tendon, or ligament structures.
4 Intended Performance
4.1 General
The intended performance shall comply with the requirements of Clause 4 of YY/T 0640-2016, while also taking into account the requirements of 4.2 to 4.4.
NOTE. Due to the diversity of anatomical sites, fracture locations, and applications, osteosynthesis
5 Design Attributes
Clause 5 of YY/T 0640-2016 and the following particular requirements apply.
a) When an implant is part of a connection system, its shape, dimensions, and tolerances shall be such that its intended use and function are not impaired.
NOTE 1.When an implant consists of two or more components, the design must take into account potential wear as well as electrolytic and corrosive effects [compare with Items 5 a), c), h), k), and l) of YY/T 0640-2016].
b) Where applicable, the dimensions of the implant shall be consistent with the anatomical characteristics of the human body into which it is to be implanted.
NOTE 2.The design of the implants must take into account factors such as anatomical structure, the type of tissue defect, bone rehabilitation, and fusion rates; the final design may represent a compromise to satisfy these varying requirements.
c) To minimize use errors, implant designers consider the required operative techniques as well as appropriate maintenance and handling of the implant, provided that these do not interfere with the implant’s intended use and function.
NOTE 3.Appendix B provides an informative list of common implant designs; these specific applications have been proven acceptable in practice.
6 Materials
Clause 6 of YY/T 0640-2016 applies.
NOTE. Standards concerning materials for implants, the suitability of which has been demonstrated through clinical use, are provided in Appendix C.
7 Design Evaluation
7.1 General
For preclinical evaluation, the implant shall be evaluated in accordance with Clause 7 of YY/T
0640-2016 and the following particular requirements.
NOTE. Appendix D lists standards that can be used to assess whether different types of implants meet the aforementioned requirements.
7.2 Preclinical Evaluation
The requirements in 7.2 of YY/T 0640-2016 apply, together with the following particular requirements.
a) In vitro simulated-use testing shall be conducted to verify the intended interactions between the implant and surgical instruments, and, where applicable, the intended interactions between interconnected implants. Where applicable, if the removal of the implant cannot be evaluated through direct comparison with existing instruments, cadaveric evaluation should be performed if possible.
b) When evaluating implants using static and/or dynamic load testing, recognized test standards (where applicable) or test models customized to the implant’s characteristics shall be used. Due to the diversity of implants and their characteristics, test standards may not exist or may require modification as needed.
c) Biological, physical, or modeling studies may be used to demonstrate that the implant achieves its intended performance, provided that they are appropriately verified.
NOTE 1.Existing data on similar implants or design characteristics are taken into account during preclinical evaluation.
NOTE 2.Test methods may involve different levels of testing.
a) Basic technical tests characterizing the implant or implant components (e.g., tension, bending, torsion);
b) Tests under expected loading conditions for installed components;
c) Biomechanical performance tests of assembled components (bone may be replaced by suitable artificial materials);
d) Static or dynamic tests (cyclic fatigue).
NOTE 3.Specific implants may be tested under specific loading and/or environmental conditions to evaluate their performance.
7.3 Clinical Evaluation
Section 7.3 of YY/T 0640-2016 applies.
7.4 Post-Market Follow-Up
Section 7.4 of YY/T 0640-2016 applies.
8 Manufacturing
Clause 8 of YY/T 0640-2016 applies.
9 Sterilization
Clause 9 of YY/T 0640-2016 applies.
10 Packaging
Clause 10 of YY/T 0640-2016 applies.
11 Information Supplied by the Manufacturer
11.1 General
Section 11.1 of YY/T 0640-2016 applies.
11.2 Label
Section 11.2 of YY/T 0640-2016 applies.
11.3 Instructions for Use
The requirements in 11.3 of YY/T 0640-2016 apply, together with the following particular requirements. The information supplied by the manufacturer shall include the following data under the corresponding conditions.
a) Restrictions on any modifications to the implant after delivery, e.g., dimensions, shape, surface condition.
b) Date of issue or latest revision of the instructions for use.
11.4 Restrictions on Combinations
Section 11.4 of YY/T 0640-2016 applies.
11.5 Implant Marking
The requirements in 11.5 of YY/T 0640-2016 apply, together with the following particular requirements.
The application of implant markings shall not impair the intended performance of the implant.
11.6 Special-Purpose Marking
Section 11.6 of YY/T 0640-2016 applies.
Appendix B (Informative)
Standards for Osteosynthesis Implants and Associated Surgical Instruments Proven
Acceptable in Clinical Use
NOTE. Informative reference standards, including other parts of this series of international standards, are listed in the bibliography of YY/T 0640-2016.
B.1 Summary of Different Types of Osteosynthesis Implants and Associated
Surgical Instruments
Relevant standards are listed in parentheses.
a) General requirements for osteosynthesis implants and associated surgical instruments (see B.2).
b) Osteosynthesis implants (see B.3).
--- Bone screws (see B.3.1);
--- Bone plates (see B.3.2);
--- Intramedullary fixation devices (see B.3.3);
--- Implants for fixation of the ends of the femur (see B.3.4);
--- Metallic skeletal pins and wires (see B.3.5);
--- Staples (see B.3.6);
--- Malleable wires for cerclages and other fixation purposes (see B.3.7);
--- Devices for external fixation of bones (see B.3.8);
--- Devices for fixation of the spine (see B.3.9);
--- Absorbable implants for internal osteosynthesis (see B.3.10).
c) ISO standards for associated surgical instruments (see B.4).
--- Drive connection instruments (see B.4.1);
--- Drilling and tapping instruments (see B.4.2).
B.2 General Requirements and Guidance
GB/T 24629 Implants for surgery - Guidance on care and handling of orthopaedic implants
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 11 pages — is available in the English PDF.
Referenced standards
Normative references
Cited by
- YY/T 0591-2011Implants for osteosynthesis - Metallic lockable intramedullary nails
- YY/T 0856-2011Implants for osteosynthesis - Metallic angled fixation devices
- YY/T 0727.1-2009Implants for surgery - Metal intramedullary nailing systems - Part 1: Intramedullary nails
- YY/T 0727.2-2009Implants for surgery - Metal intramedullary nailing systems - Part 2: Locking components
How to Download GB/T 12417.1-2024
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0640-2016 — Non-active surgical implants - General requirements
GB/T 12417.1-2008 — Non-active surgical implants -- Osteosynthesis and joint replacement implants -- Part 1: Particular requirements for osteosynthesis implants
GB/T 12417.2-2025 — Non-active surgical implants - Osteosynthesis and joint replacement implants - Part 2: Particular requirements for joint replacement implants
Secure payment via Stripe
Payments accepted
GB/T 12417.1-2024
$365.00