GB/T 12417.2-2025Non-active surgical implants - Osteosynthesis and joint replacement implants - Part 2: Particular requirements for joint replacement implants (English PDF)

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for GB/T 12417.2-2025 in 30 minutes.

Issued by

State Administration for Market Regulation

Level / Type

National · Recommended

Issue date

February 28, 2025

Implementation date

March 1, 2026

Scope

GB/T 12417.2-2025 (Non-active surgical implants - Osteosynthesis and joint replacement implants - Part 2: Particular requirements for joint replacement implants) is available as an English-translated PDF.

GB/T 12417.2-2025 is the Chinese standard "Non-active surgical implants - Osteosynthesis and joint replacement implants - Part 2: Particular requirements for joint replacement implants".

Its scope clause reads: This document specifies particular requirements for total and partial joint replacement implants, artificial ligaments, and bone cement (hereinafter referred to as the “implants”).

This document specifies requirements regarding intended performance, design attributes, materials, design evaluation, manufacturing, sterilization, packaging, and information supplied by the manufacturer.

In this document, artificial ligaments and related fixation devices are included within the term “implants”, hereinafter also referred to as the “implants”. This document is applicable to total and partial joint replacement implants, artificial ligaments, and bone cement.

NOTE. Some tests required to demonstrate compliance with this document are included in the Level 3 standard or its references. Its clauses include terms and definitions; intended performance; design attributes; materials; design evaluation; manufacturing and inspection.

It was issued by the State Administration for Market Regulation on 2025-02-28 and took effect on 2026-03-01.

Found what you need?

Document preview — GB/T 12417.2-2025

National Standard of the People's Republic of China

ICS
11.040.40

Issued by: State Administration for Market Regulation

Contents

  • Foreword...3
  • Introduction...6
  • 1 Scope...7
  • 2 Normative References...7
  • 3 Terms and Definitions...8
  • 4 Intended Performance...8
  • 5 Design Attributes...9
  • 6 Materials...10
  • 7 Design Evaluation...11
  • 8 Manufacturing and Inspection...12
  • 9 Sterilization...13
  • 10 Packaging...13
  • 11 Information Supplied by the Manufacturer...13
  • Appendix A (Informative) Approved Materials for Implant Manufacturing...15

1 Scope

This document specifies particular requirements for total and partial joint replacement implants, artificial ligaments, and bone cement (hereinafter referred to as the “implants”). This document specifies requirements regarding intended performance, design attributes, materials, design evaluation, manufacturing, sterilization, packaging, and information supplied by the manufacturer. In this document, artificial ligaments and related fixation devices are included within the term “implants”, hereinafter also referred to as the “implants”.

This document is applicable to total and partial joint replacement implants, artificial ligaments, and bone cement.

NOTE. Some tests required to demonstrate compliance with this document are included in the

Level 3 standard or its references.

2 Normative References

The contents of the following documents constitute indispensable clauses of this document through normative references in the text. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.

GB/T 3505 Geometrical product specifications (GPS) - Surface texture. Profile method - Terms, definitions and surface texture parameters (GB/T 3505-2009, ISO 4287.1997, IDT)

YY/T 0640-2016 Non-active surgical implants - General requirements

YY/T 0651.1 Implants for surgery - Wear of total hip-joint prostheses - Part 1.Loading and displacement parameters for wear-testing machines and corresponding environmental conditions for test

YY/T 0651.2 Implants for surgery - Wear of total hip-joint prostheses - Part 2.Methods of measurement

YY/T 0809.4 Implants for surgery - Partial and total hip joint prostheses - Part 4.Determination of endurance properties and performance of stemmed femoral components

YY/T 0810.1 Joint replacements - Total knee-joint prostheses - Part 1.Determination of endurance properties of knee tibial trays

YY/T 1426.1 Implants for surgery - Wear of total knee-joint prostheses - Part 1.Loading and displacement parameters for wear-testing machines with load control and corresponding environmental conditions for test

YY/T 1426.2 Implants for surgery - Wear of total knee-joint prostheses - Part 2.Methods of measurement

YY/T 1426.3 Implants for surgery - Wear of total knee-joint prostheses - Part 3.Loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test

3 Terms and Definitions

The terms and definitions defined in YY/T 0640-2016, as well as the following terms and definitions, are applicable to this document.

3.1 Artificial Ligament

Implants used to strengthen or replace the body’s natural ligaments, including their necessary fixation devices.

3.2 Joint Replacement Implants

Implants used to provide similar function to natural human joints and connect with the corresponding bone, including auxiliary implant components and materials.

3.3 Mean Centre

The position inside the sphere head that has the smallest average distance from a set of points evenly distributed on the surface of the sphere.

3.4 Radial Separation Value

The maximum difference between the average radius of the sphere and the distance from a point on the sphere to the mean centre.

NOTE. The unit is micrometers (µm).

4 Intended Performance

For the purposes of this document, the intended performance of the implant shall meet the requirements of Clause 4 of YY/T 0640-2016.Additionally, the design shall also consider the following issues.

a) the intended minimum and maximum relative angular motion between the bones fixed by the joint replacement implants;

b) the maximum intended loads (forces and moments) transmitted to the bony tissue fixed by the joint replacement implants;

c) the dynamic response of the human body to the shape/stiffness of the implant;

d) the intended wear of the articular surfaces;

e) the suitability of the implant’s dimensions and shape relative to the intended user population;

f) the adhesion strength and durability of surface treatments or surface coatings.

NOTE 1.The indications and contraindications for using a specific implant are complex issues.

When selecting an implant for a specific patient, surgeons analyze and consider factors based on their personal judgment and experience.

NOTE 2.The lifespan of an implant depends on the interaction of a variety of factors, some of which are the responsibility of the manufacturer; some are the responsibility of the surgeon, such as the implantation technique; and some are related to the patient, such as the patient’s biological and physiological response to the implant, the patient’s medical condition, and the patient’s behavioral characteristics, such as weight gain, lifting heavy objects, and high activity levels.

5 Design Attributes

5.1 General

To achieve the intended performance, the manufacturer shall formulate design attributes in accordance with the requirements of Clause 5 of YY/T 0640-2016, and also consider the following aspects.

a) the adhesion strength and durability of surface treatments and surface coatings;

b) wear on articular surfaces and other surfaces;

c) the stability of the implant when the minimum and maximum relative movements between bones are permitted;

d) avoidance of cutting or rubbing tissue while performing functions, except in cases of implantation and removal of implants;

e) creep resistance and fracture characteristics, especially when related to ligaments.

NOTE 1.Methods for assessing wear on articular surfaces and other surfaces are described in standards such as YY/T 0651.2, YY/T 1426.1, YY/T 1426.2, and YY/T 1426.3.

NOTE 2.More specific requirements may appear in other standards, such as standards for hip replacement implants.

5.2 Surface Finishing of Metallic or Ceramic Implants Actively Connected to

Ultra-High Molecular Weight Polyethylene (UHMWPE)

For the surfaces of metallic or ceramic components that are actively connected to the UHMWPE in total joint replacements, the surface roughness (Ra) shall not exceed 0.1 µm (according to

7.2.2).

5.3 Surface Finishing of Metallic or Ceramic Implants for Partial Joint

Replacements

For the articular surfaces of metallic or ceramic components in partial joint replacements, the surface roughness (Ra) shall not exceed 0.5 µm (according to 7.2.2).

5.4 Convex Spherical Metallic or Ceramic Surfaces Actively Connected to

UHMWPE

For metallic or ceramic components that are actively connected to the UHMWPE in total joint replacements, if the articular surface is convex spherical, its surface roughness (Ra) shall not exceed 0.05 µm, and the radial separation value shall not exceed 10 µm (according to 7.2.2 and

7.2.3).

5.5 Surfaces of Spherical Metallic or Ceramic Implants for Partial Joint

Replacements

For spherical metallic or ceramic components in partial joint replacements, the articular surface roughness (Ra) shall not exceed 0.5 µm, and the radial separation value shall not exceed 100

µm (according to 7.2.2 and 7.2.3).

5.6 Surfaces of Concave Spherical UHMWPE Components

For the surfaces of concave spherical UHMWPE components in total joint replacements, the articular surface roughness (Ra) shall not exceed 2 µm, and the radial separation value shall not exceed 200 µm (according to 7.2.2 and 7.2.3).

6 Materials

6.1 General

The requirements of Clause 6 of YY/T 0640-2016, and the particular requirements of 6.2, apply to this document.

NOTE 1.Appendix A lists materials that have been validated and approved for use in manufacturing implants or for use with implants.

NOTE 2.Appendix B lists mating materials that have been validated, approved and unapproved for use as articular surfaces of implants.

NOTE 3.Clause 6 of YY/T 0640-2016 stipulates that materials should be selected from those that have proven suitable for clinical use in similar application scenarios. For the purposes of this document, proof of material availability must be provided, along with implantation records of at least 500 implants and at least 5 years of successful clinical use data.

6.2 Different Metals or Alloys

In certain application scenarios, where two different metals or alloys, or two identical metals or alloys in different metallographic states, are brought into contact, and non-articular surface contact is intended, the material combination used shall not produce unacceptable galvanic effects.

NOTE. Appendix C lists the approved and unapproved metallic combinations for the non-articular contact surfaces of implants.

7 Design Evaluation

7.1 General

Joint replacement implants shall be evaluated to demonstrate that they achieve the intended performance. The evaluation shall comply with the particular requirements of Clause 7 and 7.2 to 7.4 of YY/T 0640-2016.The evaluation shall be conducted using components that fully represent the final implantation state.

7.2 Preclinical Evaluation

7.2.1 General

Preclinical evaluation considerations include.

a) biocompatibility of previously unused materials;

b) mechanical loads and related motion conditions of the implant when it functions as described in YY/T 0640-2016;

c) fatigue testing of high-stress components meeting the requirements of YY/T 0809.4 and YY/T 0810.1;

d) wear testing of joint contact surfaces meeting the requirements of YY/T 0651.1, YY/T

0651.2, YY/T 1426.1, etc.;

e) the suitability of the implant’s dimensions and shape to the intended population;

NOTE. The suitability of the implant’s dimensions and shape to the intended population is demonstrated by the following methods. cadaver implantation; imaging systems such as

X-ray, computed axial tomography (CAT), and magnetic resonance imaging; or by referring to corresponding implants that have been clinically validated (see NOTE 3 in 6.1).

f) the adhesion strength and durability of the coating if a coating is present.

7.2.2 Surface roughness measurement

The surface roughness shall be measured according to one of the methods given in GB/T 3505.

7.2.3 Sphericity measurement

The radial separation value shall be measured according to a method proven to be accurate and repeatable.

7.3 Clinical Studies

Clinical studies shall be conducted in accordance with the general requirements of ISO 14155.

7.4 Post-Market Follow-Up

Post-market performance of joint replacement implants shall be evaluated.

NOTE. Appropriate methods include survival analysis (with revision as the endpoint) and clinical evaluation. If available, relevant information from joint replacement registrations shall be considered.

8 Manufacturing and Inspection

8.1 General

The requirements of Clause 8 of YY/T 0640-2016, and the particular requirements of 8.2 to 8.4, apply to this document.

8.2 Metallic Surfaces

All metallic components shall be polished using iron-free polishing agents. Clean, degrease, rinse, and dry all components, and visually inspect the joint surfaces with normal or corrected vision. The surface shall be free of any defects that could affect implant functions, and shall also be free of finishing deposits and other contaminants.

NOTE. Defects affecting functions include oxide scale, stab marks, small notches, scratches, cracks, dents, burrs, and other defects.

8.3 Plastic Surfaces

The articular surfaces of plastic components shall not be prepared using non-removable abrasives or polishing agents. Clean, degrease (if necessary), rinse, and dry the components,

......

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Normative references

How to Download GB/T 12417.2-2025

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 12417.2-2025

$355.00