Valid

YY/T 1649.2-2019Test method for interactions of medical devices with platelets - Part 2: In vitro platelet activation products (beta-TG, PF4 and TxB2) assay (English PDF)

医疗器械与血小板相互作用试验 第2部分:体外血小板激活产物(β-TG、PF4和TxB2)的测定

Preview PDF

This is a limited preview

Buy now to download the full PDF (25 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1649.2-2019 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

October 23, 2019

Implementation date

October 1, 2020

Scope

YY/T 1649.2-2019 is the English-translated version of 医疗器械与血小板相互作用试验 第2部分:体外血小板激活产物(β-TG、PF4和TxB2)的测定.

YY/T 1649.2-2019 is the Chinese pharmaceutical industry standard that gives the in vitro platelet activation test used in the blood compatibility evaluation of a medical device or material.

It is the second part of the YY/T 1649 series and sets out the blood and its handling, the exposure of the test surface, and the markers by which activation is measured, so that the tendency of a surface to make platelets adhere and become activated can be judged. Issued on 23 October 2019, it has applied since 1 October 2020.

Found what you need?

Document preview — YY/T 1649.2-2019

National Standard of the People's Republic of China

ICS
11.040.01
Classification
C 30

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword
  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Overview
  • 5 Test principle (taking beta-TG as an example)

Foreword

YY/T 1649 `` Medical Device and Platelet Interaction Test '', including the following.

--- Part 1. In vitro platelet counting method;

--- Part 2. Determination of platelet activation products (beta-TG, PF4 and TxB2) in vitro. This part is the second part of YY/T 1649.

This section is drafted in accordance with the rules given in GB/T 1.1-2009.

Introduction

There are many methods for evaluating platelets and their activation status, but the most commonly used method for evaluating medical devices and materials in contact with blood It is a simple platelet counting method and a method of releasing platelet particulate matter. This part of YY/T 1649 is in vitro platelet activation test Specific test methods can be used as a supplement to the evaluation of the interaction between medical devices/materials and platelets in GB/T 16886.4.

As an important component of blood, platelets play a key role in preventing bleeding. Platelet particulate matter in blood after contact with device/material The increase indicates that platelets are activated, which will directly affect the blood compatibility of medical devices/materials. Described in this part of YY/T 1649 The in vitro assay of platelet activation products can be used to evaluate the potential adhesion and activation of platelets by medical devices/materials.

1 Scope

This part of YY/T 1649 specifies that when performing blood compatibility evaluation of medical devices/materials in vitro, medical devices/materials and blood Platelet Interaction in vitro Platelet Activation Test Method.

This standard applies to the adhesion and activation of platelets on the surface of medical devices/materials. Evaluation of potential.

The test system established in this section applies to human blood. If animal blood is used, it can be tested on the basis of demonstrating its suitability.

2 Normative references

The following documents are essential for the application of this document. For dated references, only the dated version applies to this document. For undated references, the latest version (including all amendments) applies to this document.

GB/T 16886.1 Biological evaluation of medical devices. Part 1. Evaluation and testing during risk management

GB/T 16886.4 Biological evaluation of medical devices. Part 4. Selection of tests for interaction with blood

GB/T 16886.12 Biological evaluation of medical devices. Part 12. Sample preparation and reference materials

3 Terms and definitions

The terms and definitions defined in GB/T 16886.1, GB/T 16886.4, and GB/T 16886.12 apply to this document.

4 Overview

When medical devices/materials come in contact with human blood, the increase in particulate matter in the blood may be caused by medical devices/materials or promote bleeding or Causes of Thrombosis. A significant increase in platelet particulate matter compared to controls indicates that the medical device/material has the potential to activate platelets can. Common platelet particulate matter mainly includes beta-thromboglobulin (beta-TG), platelet factor 4 (PF4), and thromboxane B2 (TxB2) and so on.

5 Test principle (taking beta-TG as an example)

The ELISA kit used the double antibody sandwich method to determine the human beta-TG level in the specimen. Microtiter plates were coated with purified beta-TG antibody to prepare To form a solid phase antibody, beta-TG is sequentially added to the microwells coated with the monoclonal antibody, and then combined with horseradish peroxidase (HRP) labeled beta-TG antibody.

The antibody-antigen-enzyme-labeled antibody complex was formed, and after washing thoroughly, the substrate was added with tetramethylbenzidine (TMB) to develop color. TMB in HRP enzyme It is converted into blue under the catalysis of acid, and finally into yellow under the action of acid. The shade of color is positively correlated with beta-TG in the sample.

The absorbance (OD value) was measured with a microplate reader at a wavelength of 450 nm, and the human beta-TG concentration in the sample was calculated by a standard curve.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1649.2-2019

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
25 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1649.2-2019

$125.00