YY/T 1546-2017Testing method for SPECT imaging based on CT-attenuation correction (English PDF)
用于SPECT成像CT衰减校正的试验方法
This is a limited preview
Buy now to download the full PDF (23 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1546-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
May 2, 2017
Implementation date
April 1, 2018
Scope
YY/T 1546-2017 is the English-translated version of 用于SPECT成像CT衰减校正的试验方法.
YY/T 1546-2017 is the Chinese pharmaceutical industry standard that gives the test method for SPECT imaging with CT-based attenuation correction.
It sets out the phantoms, the acquisition and reconstruction conditions and the measurements by which the accuracy of the correction is judged, among them the uniformity and the contrast recovery of the corrected images and the registration between the SPECT and the CT data.
It applies to SPECT/CT systems and not to stand-alone SPECT. Issued on 2 May 2017, it has applied since 1 April 2018.
Found what you need?
Document preview — YY/T 1546-2017
National Standard of the People's Republic of China
- ICS
- 11.040.50
- Classification
- C 43
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 test methods
- 4.1 Overview
- 4.2 Purpose
- 4.3 Method
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the test method for CT attenuation correction SPECT imaging.
This standard applies to SPECT/CT one machine. This standard does not apply to stand-alone SPECT system.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB/T 18988.3 radionuclide imaging equipment performance and test rules Part 3. Gamma camera body imaging system
GB/T 18989 Radionuclide imaging equipment performance and test rules gamma camera
3 Terms and definitions
GB/T 18988.3, GB/T 18989 and defined by the following terms and definitions apply to this document.
3.1 Attenuation correction attenuationcorrection; AC
Correction of loss of radiant intensity after passing through the media.
4 Test methods
4.1 Overview 1
For single-photon emission tomography devices, the attenuation correction of any medium within the device's fault volume for attenuation correction should have the correct theory basis. The basis of the calibration is measuring the transmittance of the object through the object in the total field of view. The accuracy actually achieved is an important part of its quantitative ability Guarantee.
4.2 Purpose 1
In the method, SPECT and CT scanning are performed on the motif of known components, and the linear attenuation coefficient obtained by using CT image conversion Value, the SPECT image attenuation correction to assess the accuracy of SPECT attenuation correction.
Note. The results of this test are also affected by uniformity, scattering, etc.
4.3 Method 1
4.3.1 Overview
The transmission of X-ray radiation through the non-uniform attenuating medium is processed to give the attenuation correction model, which is used as the SPECT projection number It is part of the reconstruction process. The accuracy of attenuation correction is reflected in the reconciliation between the transmitted value and the true value.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1546-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 18988.3-2003 — Radionuclide imaging device - Characteristics and test conditions - Part 3: Gamma camera based - Whole body imaging systems
GB/T 18988.3-2013 — Radionuclide imaging device -- Characteristics and test conditions -- Part 3: Gamma camera based-whole body imaging systems
GB/T 18989-2003 — Radionuclide imaging device - Characteristics and test conditions - Gamma cameras
Secure payment via Stripe
Payments accepted
YY/T 1546-2017