YY/T 1481-2016Guide for the check of Ultrasonic Doppler fetal monitor (English PDF)

超声多普勒胎儿监护仪核查指南

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1481-2016 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

July 29, 2016

Implementation date

June 1, 2017

Scope

YY/T 1481-2016 is the English-translated version of 超声多普勒胎儿监护仪核查指南.

YY/T 1481-2016 is the Chinese pharmaceutical industry standard that guides the checking of the ultrasonic Doppler fetal monitor.

It specifies the technical requirements, the test methods and the way the results are expressed, covering the accuracy of the fetal heart rate and its response to a change, the sensitivity of the ultrasound transducer and its acoustic output, the uterine contraction channel where fitted, the recorder and the alarms, and the electrical safety of the equipment. Issued on 29 July 2016, it has applied since 1 June 2017.

Found what you need?

Document preview — YY/T 1481-2016

National Standard of the People's Republic of China

Classification
C 41

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Preface
  • Introduction
  • 1 Scope
  • 2 normative reference document
  • 3 Terms and definitions
  • 4 Verification
  • 5 Test methods
  • Appendix A (informative) Type of test description
  • Appendix B (Informative Appendix) Inspection Record Form

Foreword

This standard is drafted in accordance with the rules given in GB/T 1.1-2009.

Introduction

The purpose of this standard is for medical institutions in use of ultrasound Doppler fetal monitor performance verification, the establishment of standardized methods, Technical requirements and test methods provide guidance.

To ensure that the ultrasound Doppler fetal monitor in a normal state of use, the need for regular verification. This standard specifies the acceptance check Inspection and periodic inspection of two forms of verification. Receive inspection check the performance of the equipment at the time of delivery and/or equipment repair; the cycle check is performed at the manufacturer The required time intervals to carry out the verification; according to the different cycles can have different verification content. This standard also provides periodic inspection reports Of the example.

The usual cycle test is undertaken by the tester of the medical institution. Receive the test, the longer the interval of the cycle test by the skilled testers, Such as biomedical engineers, medical physicists, medical device maintenance personnel, professional testers, inspection bodies, or manufacturers. Ultrasound Doppler Fetal Monitor Verification Guide

1 Scope

This standard specifies the technical requirements, test methods and test results of the Ultrasound Doppler Fetal Monitor (hereinafter referred to as "Equipment"). The expression of the fruit.

The method described in this standard is used to assist the user of the equipment to verify the performance of such equipment, mainly for medical practitioners, biomedical Engineers, medical physicists, medical equipment maintenance personnel, professional testers, inspection agencies, or manufacturers. Part of the safety involved According to YY/T 0841-2011 implementation.

2 Normative reference documents

The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.

YY/T 0841-2011 Medical electrical equipment Medical electrical equipment Periodic testing and post-repair testing

3 Terms and definitions

The following terms and definitions apply to this document.

3.1 Testers

After professional training, with the appropriate qualifications, authorized to carry out performance testing equipment or testing personnel, including medical practitioners, students Medical engineers, medical physicists, medical equipment maintenance personnel, professional testers, inspection agencies, or manufacturers.

3.2 Reference value

The value used to evaluate the subsequent measurements.

The data of the periodic test is compared with the reference number to determine whether the device is in normal working condition.

Note 1. The result of the equipment acceptance test can be used as a reference value after the new or restored equipment.

Note 2. Change from 3.30 in YY/T 0841-2011.

3.3 Check

System actions to ensure the safety and effectiveness of medical equipment, including acceptance testing and periodic inspection.

4 Verification

4.1 Acceptance test

4.1.1 Perform the first test after the equipment is delivered to the user, verifying the performance of the equipment at the time of delivery and/or equipment repair.

4.1.2 The contents of the test shall include at least the contents of the annual inspection in addition to the contract.

4.1.3 Test results of the receiving test as a reference for the performance of the equipment (see Appendix A).

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1481-2016

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1481-2016

$145.00