YY/T 1206-2013Total cholesterol kit(COD-PAP method) (English PDF)
总胆固醇测定试剂盒(氧化酶法)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1206-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY/T 1206-2013 is the English-translated version of 总胆固醇测定试剂盒(氧化酶法).
YY/T 1206-2013 is the Chinese pharmaceutical industry standard for the total cholesterol kit based on the COD-PAP method.
It specifies the principle of the measurement, the requirements, the test methods, the marking, the labelling, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the blank absorbance, the traceability of the calibrator, the accuracy, the linear range, the precision, the interference by bilirubin, haemoglobin and ascorbic acid and the stability. Issued on 21 October 2013, it has applied since 1 October 2014.
Found what you need?
Document preview — YY/T 1206-2013
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Requirement4
- 4 Test Method5
- 5 Mark, Label and Instructions7
- 6 Packaging, Transport and Storage9
Foreword
This Standard was drafted in accordance with the rules specified in GB/T 1.1-2009.
1 Scope
This standard specifies the determination principle, requirements, test method, marks, instructions, packaging, transport and storage, etc. of total cholesterol kit (COD-PAP method).
This standard applies to the quality control of liquid-type total cholesterol kit (COD-PAP method) of single reagent and double reagent used for quantitative inspection in medical laboratory.
2 Normative References
The following document is indispensable for the application of this Document. For dated references, only dated edition applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
YY/T 0466.1 Medical Devices. Symbols to Be Used with Medical Device Labels, Labeling and Information to Be Supplied. Part 1. General Requirements
3 Requirement
3.1 Appearance
It shall meet the normal appearance requirements specified by the manufacturer.
3.2 Installed capacity
The installed capacity of liquid reagent shall not be less than the labeled amount.
3.3 Absorbance of reagent blank
At the wavelength of 500nm~550nm, for double-reagent kits, it shall not be greater than 0.080; for single-reagent kits, it shall not be greater than 0.100.
3.4 Linear interval
The linear correlation coefficient (r) of theoretical concentration and measured concentration shall not be less than 0.990, within the linear interval specified by the manufacturer.
3.5 Accuracy
6 Packaging, Transport and Storage
6.1 Packaging
The packaging-pictorial marking for handling of goods shall meet the requirements of YY/T 0466.1. The packaging container shall be sealed well, intact, and free of leakage and damage.
6.2 Transport
The kit shall be transported according to the requirements of the manufacturer, and shall be free of moisture, stress, sunshine, rain, snow, acid, alkaline and damage of exterior and interior packaging during the transport.
6.3 Storage
Test kit shall be stored under the conditions specified by the manufacturer. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1206-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0466.1-2009 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2016 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2023 — Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
YY/T 1206-2013