YY/T 1253-2015Low density lipoprotein(LDL) cholesterol assay kit (English PDF)
低密度脂蛋白胆固醇测定试剂(盒)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1253-2015 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 20, 2015
Implementation date
January 1, 2016
Scope
YY/T 1253-2015 is the English-translated version of 低密度脂蛋白胆固醇测定试剂(盒).
YY/T 1253-2015 is the Chinese pharmaceutical industry standard for the low density lipoprotein cholesterol assay kit.
It specifies the requirements, the test methods, the marking, the labelling, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the blank absorbance, the traceability of the calibrator, the accuracy, the linear range, the precision, the interference by bilirubin, haemoglobin and triglycerides and the stability of the reagents. Issued on 20 March 2015, it has applied since 1 January 2016.
Found what you need?
Document preview — YY/T 1253-2015
National Standard of the People's Republic of China
- ICS
- 11.100
- Classification
- C 44
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Requirements
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
1 Scope
This standard specifies the low-density lipoprotein cholesterol assay reagent (kit) requirements, test methods, marking, labeling and instructions for use, package Loading, transportation and storage.
This standard applies to the use of the direct method in human serum or plasma low-density lipoprotein cholesterol quantitative detection reagent (box), the base Spectrophotometry in principle, including manual reagents and on the semi-automatic biochemical analyzer used.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
Symbols - Part 1 YY/T 0466.1 medical equipment used with medical device labels, labeling and information. General requirements
3 Requirements
3.1 Appearance
3.1.1 Kit components should be complete, complete, no liquid leakage.
3.1.2 Packaging label text symbols should be clear.
3.2 PACKING
Not less than the indicated value.
3.3 Reagent blank absorbance
When adding reagent blank sample test, the absorbance of the reagent blank should be < 0.05.
3.4 Sensitivity Analysis
Determination of 1.00mmol/L samples, the absorbance difference (deltaA) should be > 0.03.
3.5 Linear
Reagent (box) within the linear [0.30,10.00] mmol/L range.
a) The linear correlation coefficient (r) shall be not less than 0.995;
b) within [0.30,10.00] mmol/L range, linearity deviation should not exceed ± 0.30mmol/L or no more than ± 10%.
3.6 Precision
3.6.1 Repeatability
Repeat the test (2.50 ± 0.50) mmol/L and (5.00 ± 1.00) mmol/L in the sample, the results of the coefficient of variation (CV) should not 3%.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1253-2015
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0466.1-2009 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2016 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2023 — Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
YY/T 1253-2015