YY/T 1207-2013Uric acid assay kit (Uricase-PAP method) (English PDF)
尿酸测定试剂盒(尿酸酶过氧化物酶偶联法)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1207-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY/T 1207-2013 is the English-translated version of 尿酸测定试剂盒(尿酸酶过氧化物酶偶联法).
YY/T 1207-2013 is the Chinese pharmaceutical industry standard for the uric acid assay kit based on the uricase-peroxidase coupled method.
It specifies the requirements, the test methods, the marking, the labelling, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the blank absorbance, the traceability of the calibrator, the accuracy, the linear range, the precision, the interference and the stability of the reagents. Issued on 21 October 2013, it has applied since 1 October 2014.
Found what you need?
Document preview — YY/T 1207-2013
National Standard of the People's Republic of China
- ICS
- 11.100
- Classification
- C 44
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Requirements
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
1 Scope
This standard specifies the uric acid assay kit (uricase peroxidase conjugated Act) requirements, test methods, marking, labeling and use of said Specification, packaging, transportation and storage.
This standard applies to uric acid assay kit (uricase peroxidase conjugated method) quality control.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
Symbols - Part 1 YY/T 0466.1 medical equipment used with medical device labels, labeling and information. General requirements
3 Requirements
3.1 Appearance
Compatible with the manufacturer's appearance requirements.
3.2 PACKING
Liquid reagent loading amount not less than the labeled amount.
3.3 Reagent blank absorbance
At 500nm ~ 550nm wavelength, dual reagent blank absorbance of uric acid assay kit should be less than 0.200, a single measurement of uric acid reagent Blank absorbance given the kit should be less than 0.300.
3.4 Linear Interval
Within the linear range specified by the manufacturer, and the theoretical concentration measured concentrations of the linear correlation coefficient (r) shall not be less than 0.9900.
3.5 Accuracy
Use standard product has traceability was measured relative deviation of the measured value and the value declared shall be within ± 10.0%.
3.6 Sensitivity Analysis
When test kits given concentration of analyte, change in absorbance (deltaA) shall comply with the manufacturer's given range.
3.7 Precision
3.7.1 Within run precision
Intra-assay precision should be less than 4.0%.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1207-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0466.1-2009 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2016 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2023 — Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
YY/T 1207-2013
$125.00