YY/T 1201-2013Urea assay kit(Enzyme coupling kinetic method) (English PDF)

尿素测定试剂盒(酶偶联监测法)

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1201-2013 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

October 21, 2013

Implementation date

October 1, 2014

Scope

YY/T 1201-2013 is the English-translated version of 尿素测定试剂盒(酶偶联监测法).

YY/T 1201-2013 is the Chinese pharmaceutical industry standard for the urea assay kit based on the enzyme coupling kinetic method.

It specifies the principle of the measurement, the requirements, the test methods, the marking, the labelling, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the blank absorbance, the traceability of the calibrator, the accuracy, the linear range, the precision, the interference and the stability. Issued on 21 October 2013, it has applied since 1 October 2014.

Found what you need?

Document preview — YY/T 1201-2013

National Standard of the People's Republic of China

ICS
11.100
Classification
C 44

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 Principle of Measurement
  • 4 Requirements

Foreword

This standard was drafted in accordance with GB/T 1.1-2009 given rules.

1 Scope

This standard specifies the urea assay kit (enzyme-coupled monitoring method) measurement principle, requirements, test methods, marking, labeling and instructions for use Books, packaging, transportation and storage.

This standard applies to urea assay kit (enzyme-coupled monitoring method) quality control.

2 Normative references

The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.

GB/T 191 Packaging - Pictorial signs

3 Principle of Measurement

Urea at a urea enzyme hydrolysis NH4 CO2, under and NH4, NADH and alpha-ketoglutarate glutamate dehydrogenase Generates CO2, glutamate and NAD. Since NADH is oxidized to the NAD, reducing the absorbance at 340nm, the absorbance value decreased with It is proportional to the concentration of urea.

4 Requirements

4.1 Appearance

Compatible with the manufacturer's normal appearance requirements.

4.2 PACKING

Net content of the liquid reagent to be less than the labeled amount.

4.3 Reagent blank

4.3.1 Reagent blank absorbance

It should not be less than 1.0 (a wavelength of 340nm, light trails 1.0cm).

4.3.2 Reagent blank absorbance change rate

Should not be greater than 0.04/min.

4.4 Linear range

In 0.9mmol/L ~ 35.7mmol/L range, the linear theory of concentration and measured concentration shall be not less than the correlation coefficient r 0.9900.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1201-2013

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1201-2013

$165.00