YY/T 1155-2009Automatic Luminescence Immunoassay Analyzer (English PDF)
全自动发光免疫分析仪
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1155-2009 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
December 30, 2009
Implementation date
June 1, 2011
Scope
YY/T 1155-2009 is the English-translated version of 全自动发光免疫分析仪.
YY/T 1155-2009 is the Chinese standard "Automatic Luminescence Immunoassay Analyzer". Its scope clause reads: This standard specifies the requirements, test methods, labels, markings and instructions for use, packaging, transportation, and storage of automatic luminescence immunoassay analyzers.
This standard applies to automatic luminescence immunoassay analyzers (hereinafter referred to as analyzers) used in medical laboratories.
Analyzers use luminescent systems and immunoassay methods to quantitatively or qualitatively detect various analytes in human serum, plasma, or other body fluids, including luminescence immunoassay analyzers based on chemiluminescence, electrochemiluminescence, fluorescence, and other principles.
This standard does not apply to luminescence immunoassay analyzers based on image recognition and automatic luminescence immunoassay analyzers for point-of-care testing (POCT).
Its clauses include terms and definitions; requirements; test methods; labels, markings, and instructions for use; packaging, transportation, and storage.
Found what you need?
Document preview — YY/T 1155-2009
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Terms and definitions6
- 4 Requirements6
- 5 Test methods9
- 6 Labels, markings, and instructions for use15
- 7 Packaging, transportation, and storage15
- standard atmospheric pressure16
- References17
1 Scope
This standard specifies the requirements, test methods, labels, markings and instructions for use, packaging, transportation, and storage of automatic luminescence immunoassay analyzers.
This standard applies to automatic luminescence immunoassay analyzers (hereinafter referred to as analyzers) used in medical laboratories. Analyzers use luminescent systems and immunoassay methods to quantitatively or qualitatively detect various analytes in human serum, plasma, or other body fluids, including luminescence immunoassay analyzers based on chemiluminescence, electrochemiluminescence, fluorescence, and other principles.
This standard does not apply to luminescence immunoassay analyzers based on image recognition and automatic luminescence immunoassay analyzers for point-of-care testing (POCT).
2 Normative references
The following documents are essential for the application of this document. For any referenced document with a date, only the version with the date applies to this document.
For any referenced document without a date, the latest version (including all amendments) applies to this document.
GB/T 191 Packaging - Pictorial marking for handling of goods
GB 4793.1 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1.General requirements
GB 4793.6 Safety requirements for electrical for measurement, control, and laboratory use - Part 6.Particular requirements for laboratory equipment for the heating of materials
GB 4793.9 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 9.Particular requirements for automatic and semi- automatic laboratory equipment for analysis and other purposes
GB/T 14710 Environmental requirement and test methods for medical electrical equipment
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 luminescence immunoassay
A method that combines a luminescent system with an immune response to detect antigens or antibodies.
3.3 automatic luminescence immunoassay analyzer
A luminescence immunoassay analyzer by which all analytical processes including the injection of samples and reagents, the provision of the immune binding reaction environment, data measurement, result calculation and output are implemented automatically.
4 Requirements
4.1 Accuracy and repeatability of sample addition
The minimum and maximum sample loading volumes and the minimum and maximum reagent loading volumes specified on the instrument shall be tested and shall comply with the requirements of Table 1.
4.3 Light detection device
4.3.1 Instrument noise
The manufacturer's specifications shall not be exceeded.
4.3.2 Linearity of luminescence values
In the range of no less than 3 luminescence value orders of magnitude, the linear correlation coefficient (r) shall be >=0.99.
4.3.3 Repeatability of luminescence values
4.3.4 Stability of luminescence values
4.4 Carryover
The carryover rate shall be <=10-5.For analyzers that can only report qualitative test results, negative samples cannot be tested as positive after high-concentration positive samples are tested.
4.5 Intra-batch precision of clinical items
Select at least one clinical item and conduct the intra-batch precision test using the calibrators and kits specified by the manufacturer. The intra-batch precision shall comply with the requirements of the corresponding national standards and industry standards.
4.6 Main functions of the analyzer
The analyzer shall have the following main functions.
4.7 Appearance
The appearance of the analyzer shall meet the following requirements.
4.8 Environmental test requirements
It shall meet the requirements of climate environment group I and mechanical environment group I in GB/T 14710.
4.9 Security requirements
It shall comply with the requirements of applicable clauses in GB 4793.1, GB 4793.6,
GB 4793.9, and YY 0648.
5 Test methods
5.1 Normal working environment conditions
5.1.1 Power supply voltage
220 V±22 V; 50 Hz±1 Hz;
5.2 Accuracy and repeatability of sample addition
There are two types of determination methods. colorimetry and weighing method.
Manufacturers can choose either method.
5.2.1 Weighing method
The weighing method is carried out according to the following steps.
5.2.2 Colorimetry
The colorimetric method is carried out according to the following steps.
5.3 Accuracy and fluctuation of temperature control in the reaction zone
Place the probe of a temperature detector with a resolution of not less than 0.1 °C, or a calibrated special temperature measuring tool with the same accuracy provided by the analyzer manufacturer, at the location specified by the manufacturer.
5.4 Light detection device
5.4.1 Instrument noise
After the analyzer is turned on and in a stable working state, use a blank sample to test for 20 consecutive times, record the luminescence value of each test, calculate the arithmetic mean 𝐼 of the luminescence values, and calculate the instrument noise IB according to formula (5), which shall meet the requirements of 4.3.1.
5.4.2 Linearity of luminescence values
One of the following two methods can be chosen.
5.6 Intra-batch precision of clinical items
Using calibrators and kits for the clinical test item specified by the manufacturer, test the intra-batch precision in accordance with the corresponding national standards, industry standards, or test kit technical requirements, which shall comply with the provisions of 4.5.
5.7 Main functions of the analyzer
The verification is performed according to the instructions; the functions shall meet the requirements of 4.6.
5.8 Appearance
Visual inspection is carried out; the appearance shall comply with the requirements of
4.7.
5.9 Environmental test requirements
The test is carried out according to the method specified in GB/T 14710.The results shall meet the requirements of 4.8.
5.10 Security requirements
The test is carried out according to the methods specified in GB 4793.1, GB 4793.6,
GB 4793.9, and YY 0648, and the results shall meet the requirements of 4.9.
6 Labels, markings, and instructions for use
They shall comply with the requirements of GB/T 29791.3.
7 Packaging, transportation, and storage
7.1 Packaging
The packaging of the analyzer shall comply with the following provisions.
7.2 Transportation
Transportation shall be carried out in accordance with the requirements specified by the manufacturer.
7.3 Storage
Storage shall be carried out in accordance with the manufacturer's instructions.
YY/T 1162-2009 | YY/T 1161-2009 | YY/T 1164-2021 | […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Similar standards
How to Download YY/T 1155-2009
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0463-2026 — Dentistry - Refractory investment and die materials
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
YY/T 1155-2009
$85.00