YY/T 0916.7-2024Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications (English PDF)
医用液体和气体用小孔径连接件 第7部分:血管内或皮下应用连接件
This is a limited preview
Buy now to download the full PDF (16 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0916.7-2024 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
July 8, 2024
Implementation date
July 20, 2026
Scope
YY/T 0916.7-2024 is the English-translated version of 医用液体和气体用小孔径连接件 第7部分:血管内或皮下应用连接件.
YY/T 0916.7 is the Chinese adoption of the small-bore connector standard for intravascular and hypodermic use, the part of the ISO 80369 family that governs the Luer-type connection.
Its purpose is to make wrong-route connections physically impossible: a family of incompatible geometries so that an enteral line cannot be joined to an intravenous one, a class of error that has killed patients and that no amount of labelling had prevented.
At 32,760 words it is one of the longest documents in the series, because the geometry has to be specified to a tolerance that guarantees both connection and non-connection.
It is cited normatively by device standards including YY 0781 on blood pressure transducers. For any manufacturer of infusion, injection or vascular access devices for China, this is a governing connector standard.
Found what you need?
Document preview — YY/T 0916.7-2024
National Standard of the People's Republic of China
Issued by: National Medical Products Administration
Contents
- 4 Introduction...5
- 1 Scope...6
- 2 Normative references...7
- 3 Terms and definitions...7
- 4 General requirements...9
- 4.1 General requirements for Luer connectors...9
- 4.2 Type testing...10
- 5 Dimensional design requirements for Luer connectors...10
- 6 Performance requirements...11
- 6.1 Fluid leakage...11
- 6.2 Sub-atmospheric-pressure air leakage...11
- 6.3 Stress cracking...12
- 6.4 Resistance to axial load separation...12
- 6.5 Resistance to unscrewing torque separation...12
1 Scope
YY/T 0916.7 is the Chinese adoption of the small-bore connector standard for intravascular and hypodermic use, the part of the ISO 80369 family that governs the Luer-type connection. Its purpose is to make wrong-route connections physically impossible: a family of incompatible geometries so that an enteral line cannot be joined to an intravenous one, a class of error that has killed patients and that no amount of labelling had prevented. At 32,760 words it is one of the longest documents in the series, because the geometry has to be specified to a tolerance that guarantees both connection and non-connection. It is cited normatively by device standards including YY 0781 on blood pressure transducers. For any manufacturer of infusion, injection or vascular access devices for China, this is a governing connector standard.
This document specifies the dimensions and requirements for the design and functional characteristics of small-bore connectors, which are intended for use in connections for intravascular applications or hypodermic injection applications involving medical devices and accessories. Examples. Hypodermic syringes and needles or intravascular (hereinafter referred to as IV) cannulae with male and female Luer non-locking (slip) connectors and Luer-lock connectors. Note
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
YY/T 0916.1-2021 Small bore connections for medical liquids and gases - Part 1. General requirements ( ISO 80369-1.2018, IDT)
YY/T 0916.6-2022 Small bore connectors for medical liquids and gases - Part 6. Connectors for axonal applications ( ISO 80369-6.2016, IDT)
YY/T 0916.20-2019 Small bore connections for medical liquids and gases - Part 20. General test methods ( ISO 80369-20.2015, IDT)
ISO 14971.2019 Medical devices - Application of risk management to medical devices
GB/T 42062-2022 Medical devices - Application of risk management to medical devices
IEC 62366-1.2015 Medical devices - Part
3 Terms and definitions
The terms and definitions defined in YY/T 0916.1-2021, YY/T 0916.20-2019, ISO 14971.2019, IEC 62366-1.2015 (as shown in Appendix J), as well as the following terms and definitions, apply to this document.
3.1 Auxiliary dimension Dimension derived from other dimensions, provided for reference only. [Source: ISO 10209.2012, 4.2] 3.2 [Source: IEC 60601-1.2005, 3.97]
3.7 Rigid material Material with a flexural modulus or tensile modulus greater than 3433 MPa. Examples. Metals, glass, certain fiber-reinforced polymers, high-performance polymers.
3.8 Semi-rigid material Material with a flexural modulus or tensile modulus 700 MPa ~ 3433 MPa. Example. Thermoplastics.
4 General requirements
4.1 General requirements for Luer connectors Luer connectors intended to be manufactured in accordance with this document shall comply with the general requirements of YY/T 0916.1-2021, unless otherwise specified in this document. In some tolerance fits, at the least material condition (LMC), the inner diameter of the fluid lumen of a male Luer connector may contact the sealing surface of the N1 male connector (N1) specified in YY/T 0916.6-2022; consequently, these connectors may misconnect when evaluated against the non-interconnectable characteristics test in Appendix B of YY/T 0916.1-2021.Additional information is provided in G.2.2. Reference connectors for evaluating non-interconnectable characteristics are described in Appendix C (Figures C.1, C.2, C.4, C.5, as appropriate). If the design of a medical device or accessory provides the characteristics of the Luer connectors specified in this document, these characteristics shall be included in the verification of compliance with this document. Where necessary, small-bore connectors shall be fitted to the medical device or accessory to verify compliance with Appendix B of YY/T 0916.1-2021.
Note 1: A summary of medical devices incorporating these connectors and their attributes is provided in Appendix D.
Note 2: A summary of usability requirements for Luer connectors is provided in Appendix E.
Note 3: A summary of criteria and requirements for Luer connectors is provided in Appendix F.
Note 4: A summary of design assessment for Luer connectors in accordance with
6.1 of YY/T 0916.1-2021 is provided in Appendix G. Note
5.This document is intended to address the relevant essential principles of safety and performance of ISO 16142-1.2016, as shown in Appendix H. Note
6.This document is intended to address the relevant general safety and performance requirements of EU Regulation (EU) 2017/745, as shown in Appendix I.
4.2 Type testing Compliance with the requirements of this document shall be determined through type testing.
5 Dimensional design requirements for Luer connectors
Luer connectors shall comply with the dimensional and tolerance requirements specified below. - Male Luer non-locking (slip) connector (L1). Figure B.1 and Table B.1; - Female Luer non-locking (slip) connector (L1). Figure B.2 and Table B.2; - Male Luer lock connector with fixed collar (L2). Figure B.3 and Table B.3; - Male Luer lock connector with movable or rotatable collar (L2). Figure B.4 and Table B.4; - Female Luer lock connector (L2). Figure B.5 and Table B.5; - Female Luer lock connector (L2) Type A with lugs perpendicular to the axis. Figure B.6 and Table B.6; - Female Luer lock connector (L2) Type B with lugs perpendicular to the axis. Figure B.7 and Table B.7; - Female Luer lock connector (L2) Type C with lugs perpendicular to the axis. Figure B.8 and Table B.8. Compliance with the corresponding Figures and Tables is verified by confirming the dimensions and tolerances specified in Appendix B.
Note. See Appendix A.
6.1 Fluid leakage
6.1.1 Fluid leakage requirements Luer connectors shall be evaluated for leakage using either the pressure decay leakage test method or the positive pressure liquid leakage test method.
6.1.2 Pressure decay leakage When evaluating the fluid leakage performance of Luer connectors using the pressure decay leakage test method, it shall use air as the test medium; apply a pressure of 300 kPa ~ 330 kPa and maintain it for 15 s ~ 20 s. The leakage rate shall not exceed
0.005 Pa·m^3/s. Check for compliance by testing in accordance with Appendix B of YY/T 0916.20- 2019, using the leakage reference connectors specified in Appendix C (Figures C.1, C.2, C.4, C.5, as appropriate). Higher pressures may be applied.
6.1.3 Positive-pressure liquid leakage When evaluating the fluid leakage performance of Luer connectors using the positive- pressure liquid leakage test method, it shall apply a pressure of 300 kPa ~ 330 kPa and maintain it for 30 s ~ 35 s; there shall be no signs of leakage sufficient to form a falling drop. Check for compliance by testing in accordance with Appendix C of YY/T 0916.20- 2019, using the leakage reference connectors specified in Appendix C (Figures C.1, C.2, C.4, C.5, as appropriate). Higher pressures may be applied.
6.2 Sub-atmospheric-pressure air leakage Sub-atmospheric-pressure air leakage of Luer connectors shall be evaluated. When subjected to a sub-atmospheric pressure of
88.0 kPa for 15 s ~ 20 s, the leakage rate of the Luer connectors shall not exceed
0.005 Pa·m^3/s. Check for compliance by testing in accordance with Appendix D of YY/T 0916.20- 2019, using the leakage reference connectors specified in Appendix C (Figures C.1, C.2, C.4, C.5, as appropriate). Higher sub-atmospheric pressures may be applied.
6.3 Stress cracking Stress cracking of Luer connectors shall be evaluated. After being subjected to the stress specified in Appendix E of YY/T 0916.20-2019, Luer connectors shall meet the requirements of 6.1.1. Check for compliance by testing in accordance with Appendix E of YY/T 0916.20-2019, using the stress-cracking reference connectors specified in Appendix C (Figures C.1, C.2, C.4, C.5, as appropriate).
6.4 Resistance to axial load separation Axial load separation of Luer connectors shall be evaluated. When subjected to the following axial separation forces for a duration of 10 s ~ 15 s, Luer connectors shall not separate from the standard connector.
a) Luer non-locking (slip) connectors. 23 N ~ 25 N;
b) Luer lock connectors. 32 N ~ 35 N. Check for compliance by testing in accordance with Appendix F of YY/T 0916.20-2019, using the standard connectors for resistance to axial load separation specified in Appendix C (Figures C.2, C.3, C.5, C.6, as appropriate). Higher axial separation forces or longer hold times may be used.
6.5 Resistance to unscrewing torque separation Resistance to unscrewing torque separation of Luer lock connectors shall be evaluated. Luer lock connectors shall not separate from standard connectors, when subjected to an unscrewing torque of
0.018 N·m ~
0.020 N·m for a duration of 10 s ~ 15 s. Check for compliance by testing in accordance with Appendix G of YY/T 0916.20- 2019, using the standard connectors for resistance to unscrewing torque separation specified in Appendix C (Figures C.1 and C.4, as appropriate). Higher unscrewing torques or longer holding times may be used.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.
Referenced standards
Normative references
- YY/T 0916.1-2021Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
- YY/T 0916.6-2022Small-bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications
- YY/T 0916.20-2019Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
- GB/T 42062-2022Medical devices - Application of risk management to medical devices
- YY/T 1474-2016Medical devices - Application of usability engineering to medical devices
ISO 80369 · ISO 14971.2019 · IEC 62366 · IEC 62366.2007
Similar standards
How to Download YY/T 0916.7-2024
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 42062-2022 — Medical devices - Application of risk management to medical devices
YY/T 0916.1-2021 — Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
YY/T 0916.20-2019 — Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
Secure payment via Stripe
Payments accepted
YY/T 0916.7-2024
$745.00