YY/T 0809.2-2020Implants for surgery—Partial and total hip joint prostheses - Part 2: Articulating surfaces made of metallic, ceramic and plastics materials (English PDF)
外科植入物 部分和全髋关节假体 第2部分:金属、陶瓷及塑料关节面
This is a limited preview
Buy now to download the full PDF (18 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0809.2-2020 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
September 27, 2020
Implementation date
September 1, 2021
Scope
YY/T 0809.2-2020 is the English-translated version of 外科植入物 部分和全髋关节假体 第2部分:金属、陶瓷及塑料关节面.
YY/T 0809.2-2020 is the Chinese pharmaceutical industry standard that sets the requirements for the articulating surfaces of total and partial hip joint prostheses of the ball and socket kind.
It is the second part of the YY/T 0809 series and covers the head and the bearing surface it runs against in the materials used for them - metal, ceramic and polymer - fixing the dimensions and tolerances, the clearance between head and cup, the surface roughness and sphericity, and the tests by which each is verified. Issued on 27 September 2020, it has applied since 1 September 2021.
Found what you need?
Document preview — YY/T 0809.2-2020
National Standard of the People's Republic of China
- Replacing
- YY/T 0809.2-2010
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope6
- 2 Normative references6
- 3 Terms and definitions7
- 4 Requirements and methods7
- Annex A (normative) Determination of radial separation value11
1 Scope
This part of YY/T 0809 specifies requirements for the articulating surfaces of those types of total and partial hip joint prostheses that provide a joint replacement of ball and socket configuration, as follows:
a) sphericity and surface finish requirements for metallic and ceramic femoral prostheses for partial hip joint replacement that are in accordance with classification a) of YY/T 0809.1-2010;
b) sphericity and surface finish requirements for bipolar heads with plastics inner surfaces which articulate on femoral components that are in accordance with classification a) of YY/T 0809.1-2010 and with metallic or ceramic outer surfaces which articulate on the biological acetabulum;
c) sphericity and surface finish requirements and dimensional tolerances for plastics acetabular components that are in accordance with classification b) of YY/T 0809.1-2010;
d) sphericity and surface finish requirements and dimensional tolerances for metallic or ceramic femoral components of total hip joint prostheses that are in accordance with classification c) of YY/T 0809.1-2010.
This Part is applicable to articulating surfaces of all types of total and partial hip joint prostheses that provide a joint replacement of ball and socket configuration.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 3505, Geometrical product specifications (GPS) - Surface texture: Profile method - Terms definitions and surface texture parameters (GB/T 3505-2009, ISO 4287:1997, IDT)
GB/T 10610-2009, Geometrical product specifications (GPS) - Surface texture:
Profile method - Rules and procedures for the assessment of surface texture (ISO 4288:1996, IDT)
YY/T 0809.1-2010, Implants for surgery - Partial and total hip joint prostheses - Part 1: Classification and designation of dimensions (ISO 7206-1:2008, IDT)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in YY/T 0809.1- 2010 and the following apply.
3.1 Spherical pole
Point created by the intersection of the axis of symmetry of the partial sphere of the head or cup with the spherical surface of the head or cup.
4 Requirements and methods
4.5 Metallic and ceramic acetabular components of total hip joint prostheses
4.1 Femoral components of total hip joint prostheses
Note: This subclause refers to spherical articulating surfaces of femoral components of total hip joint prostheses in accordance with classification c) of YY/T 0809.1- 2010.
4.1.1 Sphericity
When measured in accordance with the method given in A.1, the departure from sphericity of the spherical articulating surface of a femoral component shall have a radial separation value not greater than 10 µm, as shown in Formula (A.1) of Appendix A.
4.1.2 Surface finish
The principles for the surface finish measurements of hip joint prostheses components are given in GB/T 3505. Surface finish measurements shall be performed according to the rules and procedures described in GB/T 10610-2009.
The spherical articulating surfaces of metallic and ceramic components shall have Ra values not greater than 0.05 µm and 0.02 µm, respectively. GB/T 10610-2009, Table 1, requires using a cut-off value of 0.25 mm for the 0.05 µm requirement and requires using a cut-off value of 0.08 mm for the 0.02 µm requirement. The Rt value shall not be greater than 1.0 µm. The cut-off value used to determine the Rt value shall be in accordance with GB/T 10610-2009, Table 2.
b) position of measurement on specimen;
c) average Ra of the five locations measured.
When examined by normal or corrected vision, the articulating surface shall be free from embedded particles and from scratches and score marks other than those arising from the finishing process.
4.2.3 Dimensional tolerances
The spherical socket shall have a diameter equal to the nominal diameter within a tolerance of +0.1 mm ~ +0.3 mm at a temperature of 20 °C ± 2 °C (i.e., it shall be oversized within the given tolerance).
4.3 Femoral prostheses for partial joint replacements
Note: This subclause refers to spherical articulating surfaces of femoral prostheses for partial joint replacements in accordance with classification a) of YY/T 0809.1- 2010.
4.3.1 Sphericity
When measured in accordance with the method given in A.1, the departure from sphericity of the spherical articulating surface shall have a radial separation value no greater than 100 µm, as shown in Formula (A.1) of Appendix A.
4.3.2 Surface finish
The principles for the surface finish measurements of hip joint prostheses components are given in GB/T 3505. Surface finish measurements shall be performed according to the rules and procedures described in GB/T 10610-2009.
The spherical articulating surface of the implant shall have an Ra value not greater than 0.5 µm using a cut-off value of 0.8 mm. The Rt value shall not be greater than 1.0 µm.
The cut-off value used to determine the Rt value shall be in accordance with GB/T 10610-2009, Table 2.
The measurements shall be taken at five locations on the spherical surface. One measurement shall be taken in each of four quadrants approximately 30° from the spherical pole and one at the spherical pole. The following details shall be reported along with the measured values:
a) stylus tip radius;
b) position of measurement on specimen;
c) average Ra of the five locations measured.
When examined by normal or corrected vision, the articulating surface shall be free from embedded particles and from scratches and score marks other than those arising from the finishing process.
4.3.3 Dimensional tolerances
The tolerance for the spherical head shall be ±0.5 mm on the nominal diameter.
4.4 Bipolar heads Note:
This subclause refers to bipolar heads for femoral prostheses consisting of a concave (inner) spherical surface in a plastics component intended to articulate with a femoral component of a partial or total hip joint. The bipolar head also has a convex (outer) spherical surface which is intended to articulate with the biological acetabulum.
4.4.1 Inner articulating surface
The inner articulating surface of the plastics part of a bipolar head shall conform to 4.2.
4.4.2 Outer articulating surface
The outer articulating surface of a bipolar head shall conform to 4.3.
Note: This subclause refers to spherical articulating surfaces of acetabular components of total hip joint prosthesis with classification c) of YY/T 0809.1-2010.
4.5.1 Surface finish
The principles for the surface finish measurements of hip joint prostheses components are given in GB/T 3505. Surface finish measurements shall be performed according to the rules and procedures described in GB/T 10610-2009.
The spherical articulating surfaces of metallic and ceramic components shall have an Ra value not greater than 0.05 µm using a cut-off value of 0.25 mm.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 3505Geometrical product specifications (GPS) -- Surface texture: Profile method -- Terms, definitions and surface texture parameters
- GB/T 10610-2009Geometrical product specifications (GPS) -- Surface texture: Profile method -- Rules and procedures for the assessment of surface texture
- YY/T 0809.1-2010Implants for surgery-Partial and total hip joint prostheses - Part 1: Classification and designation of dimensions
Similar standards
Editions of YY/T 0809.2
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0809.2-2020 | Implants for surgery—Partial and total hip joint prostheses - Part 2: Articulating surfaces made of metallic, ceramic and plastics materials | current edition | Current |
| YY/T 0809.2-2010 | Implants for surgery—Partial and total hip joint prostheses - Part 2: Articulating surfaces made of metallic, ceramic and plastics materials | previous edition | Superseded |
This page sells the current edition, YY/T 0809.2-2020. Earlier editions are listed for reference only.
How to Download YY/T 0809.2-2020
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 10610-2009 — Geometrical product specifications (GPS) -- Surface texture: Profile method -- Rules and procedures for the assessment of surface texture
GB/T 3505-2009 — Geometrical product specifications (GPS) -- Surface texture: Profile method -- Terms, definitions and surface texture parameters
YY/T 0809.1-2010 — Implants for surgery-Partial and total hip joint prostheses - Part 1: Classification and designation of dimensions