YY/T 0751-2009Ultrasonics—Dental descaler systems—Measurement and declaration of the output characteristics (English PDF)
超声洁牙设备 输出特性的测量和公布
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0751-2009 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
November 25, 2009
Implementation date
December 1, 2010
Scope
YY/T 0751-2009 is the English-translated version of 超声洁牙设备 输出特性的测量和公布.
YY/T 0751-2009 adopts IEC 61205:1993 without technical change and deals with ultrasonic dental scalers, the devices used to remove calculus from teeth.
It identifies their basic non-thermal output characteristics, such as the primary tip vibration excursion and the vibration frequency, and describes how to measure them, including a method based on the width of the mark the vibrating tip scores on a slide.
Measurements are made under the operating conditions recommended by the manufacturer, taking account of ambient temperature, tip irrigation water and excitation frequency.
It also states which output characteristics the manufacturer must declare. Safety and the biological effects of the ultrasound output are not addressed. Issued on 25 November 2009, it has applied since 1 December 2010.
Found what you need?
Document preview — YY/T 0751-2009
National Standard of the People's Republic of China
- ICS
- 11.040.01; 17.140.50
- Classification
- C 41
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions
- 4 Symbol
- 5 General measurement conditions
Foreword
This standard is equivalent to using the International Electrotechnical Commission standard IEC 61205:1993 "ultrasonic scaler device output characteristic measurement and disclosure." The standard IEC 61205:1993 only made a very small number of editorial changes did not affect the degree of consistency. The Standard Annexes A and B are informative appendices.
This standard by the standardization of medical ultrasound equipment for medical appliances National Standardization Technical Committee Technical Committee (SAC/TC10/SC2) Centralized.
Introduction
Ultrasonic scaler equipment widely used in spot and remove calculus from the teeth, although the mode of action of such devices are not fully carry out special Zheng explained, but the scavenging effect ultrasonic scaler device depends on the amplitude of the vibration. The purpose of this standard is the fundamental vibration ultrasonic scaler device regulations Characteristics of standard measurement methods and technical requirements. In this standard does not consider security issues and biological effects of ultrasound output. Ultrasonic scaler equipment Measuring and publishing output characteristics
1 Scope
This standard specifies.
--- Ultrasonic scaler device basic non-thermal output characteristics;
--- Measurement equipment ultrasonic scaler output performance;
--- By the ultrasonic scaler equipment manufacturer's published output characteristics.
2 Normative references
The following documents contain provisions which, through reference in this standard and become the standard terms. For dated references, subsequent Amendments (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement are based on research Whether the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.
IEC 60782. 1984 ultrasonic magnetostrictive transducers measure
3 Terms and Definitions
The following terms and definitions apply to this standard.
3.1 Blocking force
Tip must be in contact with the slides, the applied force, so that the tip of the main vibrating 0.15N ± 0.02N offset falling under the conditions of contact force 50% lower.
3.2 Primary tip vibration excursion
In the role of the head from the tip of no more than 1mm from the point of maximum amplitude in the direction of the peak to peak displacement of the application head.
3.3 Tip vibration frequency
Effect tip of the fundamental frequency of vibration.
4 Symbol
Ju main vibrating tip offset. When Wo shutdown, slides on the score mark width. When Wp power, slides on the score mark width.
5 General measurement conditions
5.1 Working conditions
Measurements shall be recommended by the manufacturer under the conditions set parameters, the parameters considered are as follows.
--- Ambient temperature;
--- Probe tip flushing water;
--- Excitation frequency;
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
IEC 60782
Similar standards
How to Download YY/T 0751-2009
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
YY/T 0492-2026 — Implantable leads
Secure payment via Stripe
Payments accepted
YY/T 0751-2009
$205.00