YY/T 0586-2016Medical polymer products - Test method for radiopacity (English PDF)
医用高分子制品 X射线不透性试验方法
This is a limited preview
Buy now to download the full PDF (25 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0586-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
January 26, 2016
Implementation date
January 1, 2017
Scope
YY/T 0586-2016 is the English-translated version of 医用高分子制品 X射线不透性试验方法.
YY/T 0586-2016 is the Chinese pharmaceutical industry standard that gives the test methods for the radiopacity of medical polymer products and materials, using X-ray techniques including fluoroscopy, radiography, angiography, computed tomography and dual energy absorptiometry.
It sets out the specimens and the reference wedges, the exposure conditions, the reading of the image and the expression of the result as an equivalent thickness of aluminium, so that a claim that a device is visible under X-rays can be verified. Issued on 26 January 2016, it has applied since 1 January 2017.
Found what you need?
Document preview — YY/T 0586-2016
National Standard of the People's Republic of China
- Replacing
- YY/T 0586-2005
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Terms and definitions5
- 4 Summary of test methods7
- 5 Significance and use7
- 6 Apparatus7
- 7 Test specimens8
- 8 Imaging conditions9
- 9 Procedure9
- 10 Report11
- 11 Precision and bias11
- Annex A (informative) Description13
Foreword
This Standard is drafted in accordance with the rules given in GB/T 1.1-2009.
This Standard replaces YY/T 0586-2005 "Medical polymer products - Test methods of radiopacity". Compared with YY/T 0586-2005 "Medical polymer products - Test methods of radiopacity", in addition to editorial modifications, the main technical changes are as follows.
- DELETE Method A; COMBINE Method B and Method C;
- ADD series of terms associated with digital image analysis;
- ADD the method to measure the pixel intensity by digital image analysis tools;
- MODIFY the location of the test specimen in the X-ray image to allow it to sit at or near the middle versus only the middle;
- ADD the following procedure. as appropriate, the effect of the clinical X-ray table shall also be included with use of an appropriate material of thickness similar to that used clinically;
- ADD a description that the step wedge may be used as a user-defined standard;
- CLARIFY the X-ray exposure procedure to indicate completion of X-ray exposure using conditions typical of those used in the X-ray diagnosis of humans, the product and for the particular area of interest;
- RENEW the report requirements, by adding the report requirement for the material and thickness of X-ray table (if used).
This Standard is modified on the basis of ASTM F 640-12 "Standard Test Method for Determining Medical Radiometric Transmission" and is technically equivalent.
1 Scope
The test methods in this Standard cover the determination of the radiopacity of materials and products utilizing X-ray based techniques, including fluoroscopy, angiography, CT (computed tomography) and DEXA (dual energy X-ray absorptiometry), also known as DXA. The results of these measurements are an indication of the likelihood of locating the product within the human body.
Radiopacity is determined by qualitatively comparing image(s) of a test specimen and a user-defined standard, with or without the use of a body mimic, or quantitatively determining the specific difference in optical density or pixel intensity between the image of a test specimen and the image of a user-defined standard, with or without the use of a body mimic.
NOTE. As a method standard, this Standard does not specify the acceptance criteria for X-ray detectability, but for ease of use, Annex A recommends the acceptance criteria.
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB 3880.1-2006 Wrought aluminium and aluminium alloy plates, sheets and strips for general engineering - Part 1: Technical conditions of delivery
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 Body mimic
A piece of material, a phantom, a cadaver or an animal utilized to mimic the appropriate X-ray attenuation through a particular part of the human body.
3.3 Grayscale range
The number of levels in pixel intensity resolved in the digital image.
NOTE. This is normally 256 levels in an 8-bit grayscale image.
3.4 Optical density
The range of values of optical density as measured by a densitometer; in this test method, the expected range is 0.50 to 1.50.
3.5 Optical density difference
The difference in optical density units between two regions or objects in an image, reported to at least two digits to the right of the decimal point.
3.6 Pixel intensity
The grayscale level of a pixel between 0 and 255, as determined by the digital analysis program.
4 Summary of test methods
The test specimen is placed so it sits at or near the middle of the X-ray image area in the X-ray imaging system. X-ray images are made at specified voltages, times, and currents that are typical of those used in the X-ray diagnosis of humans
5 Significance and use
5.1 These methods are intended to determine whether a polymer product (a material, product, or part of a product) has the degree of radiopacity desired for its application as a medical device in the human body.
5.2 These methods allow for both qualitative and quantitative evaluation in different comparative situations.
6 Apparatus
6.2 X-ray film or digital image acquisition system
The film or digital imaging system shall be appropriate for the imaging conditions used. A grid may be used.
6.3 Body mimic (if used)
6.3.1 Animal
An appropriate animal, or portion of appropriate animal, with which to perform the tests may be used.
6.3.5 Step wedge
A step wedge may be used as a user-defined standard, if it has the requisite thickness steps.
6.4 Rubber blankets
Blankets incorporating X-ray absorbers may be used to mask the image area not covered by the body mimic material (this prevents undercutting). Lead sheets may also be used for masking.
6.5 Back-scatter protection
As appropriate with the specific X-ray imaging system.
6.7 Step tablet
For calibrating densitometers. This is not required if digital analysis is used.
7 Test specimens
7.1 Material
The material may be in any form. For comparing results between materials, best results will be obtained by utilizing the same form and dimensions for each material.
8 Imaging conditions
8.1 The test shall be performed at appropriate conditions for the imaging system, the product or material, and the area of the body within which the product is intended for use.
8.2 Imaging conditions shall be described in the test report.
9 Procedure
9.1 Test specimen placement
Place the test specimen(s) and the user-defined standard at or near the middle of the X-ray imaging area. If a body mimic is used, place the test specimen(s) and the user-defined standard as appropriate in, on, or under the body mimic. As appropriate, the effect of the clinical X-ray table shall also be included with use of an appropriate material of thickness similar to that used clinically.
9.3 Film development
Develop the X-ray film in accordance with the manufacturer's instructions. If a digital analysis method will be used, convert the developed film image(s) to digital format using an appropriate digital scanning or photographic method.
9.4 Qualitative analysis
Visually compare the image(s) of the test specimen and the user-defined standard to the background on the film or in the digital image (whether original or converted from film).
9.5.2 Measurement of pixel density (digital)
9.5.2.1 Background
Measure the background pixel intensity of the test specimen and the user-defined standard by digital analysis method.
10 Report
10.1 The report shall include the following information.
11 Precision and bias
11.1 The precision and bias of these test methods have not yet been established.
11.2 It is intended to develop data for the overall method and the measurement of density from a round-robin test.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.
Referenced standards
Similar standards
Editions of YY/T 0586
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0586-2016 | Medical polymer products - Test method for radiopacity | current edition | Current |
| YY/T 0586-2005 | Medical polymer products - Test method for radiopacity | previous edition | Superseded |
This page sells the current edition, YY/T 0586-2016. Earlier editions are listed for reference only.
How to Download YY/T 0586-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
YY/T 0492-2026 — Implantable leads
Secure payment via Stripe
Payments accepted
YY/T 0586-2016
$75.00