YY/T 0079-2016Metallic clip (English PDF)
医用金属夹
This is a limited preview
Buy now to download the full PDF (24 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0079-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 23, 2016
Implementation date
January 1, 2017
Scope
YY/T 0079-2016 is the English-translated version of 医用金属夹.
YY/T 0079-2016 is the Chinese pharmaceutical industry standard for the metallic surgical clip, the clip applied to close a vessel or a duct.
It sets the types and the basic dimensions, the requirements, the test methods, the inspection rules, the marking, the operating instructions and the rules for packaging, transport and storage, covering the material and its corrosion resistance, the form and the finish of the clip, its closing and holding force, the retention on the tissue and the compatibility with its applier. Issued on 23 March 2016, it has applied since 1 January 2017.
Found what you need?
Document preview — YY/T 0079-2016
National Standard of the People's Republic of China
- Replacing
- YY/T 0079-2006
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Type and basic dimension6
- 4 Requirements7
- 5 Test methods9
- 6 Inspection rule12
- 7 Marking, operating instructions12
- 8 Packaging and Storage12
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
This standard replaces YY/T 0079-2006 "Implants for surgery - Metallic clip".
The main differences between this standard and YY/T 0079-2006 are as follows.
-- Modify the name of the standard, which is changed from "Implants for surgery - Metallic clip" to "Metallic clip";
-- Revise structure type and size requirements;
-- Revise material trademark and hardness range of titanium metal clip;
-- Delete tantalum metal clip and related content;
-- Add clamping performance, corrosion resistance and package sealing requirements and test methods;
-- Add microbiological indicators requirements and test methods;
-- Improve biocompatibility requirements and test methods.
1 Scope
This standard specifies many requirements related to metallic clip, such as type and basic dimension, requirements, test methods, inspection rules, marking, operating instructions, packaging, transport and storage requirements.
This standard applies to silver clip that clamps the capillaries of brain and titanium clip that clamps the abdominal tubular tissue. The following is referred to as (metallic clip).
2 Normative references
The following documents are essential for the application of this document. For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB/T 4135 Silver ingots
GB/T 4340.1 Metallic materials - Vickers hardness test - Part 1. Test method
GB/T 6682 Water for analytical laboratory use - Specification and test methods
GB/T 13810-2007 Wrought titanium and titanium alloy for surgical implants
GB/T 14233.2-2005 Test methods for infusion, transfusion, injection equipment for medical use - Part 2. Biological test methods
GB/T 16886.3-2008 Biological evaluation of medical devices - Part 3. Tests for genotoxicity, carcinogenicity and reproductive toxicity
GB/T 16886.5-2003 Biological evaluation of medical devices - Part 5. Test for in vitro cytotoxicity
GB/T 16886.6-1997 Biological evaluation of medical devices - Part 6. Tests for local effects after implantation
GB/T 16886.7 Biological evaluation of medical devices - Part 7. Ethylene oxide sterilization residuals
4 Requirements
4.3 Appearance
4.3.1 Both sides of metallic clip shall be symmetric, and the appearance shall be smooth without any defects such as sharp edges, burrs, cracks, or spots.
4.3.2 The stripe and/or groove on metallic clip shall be clear and complete, and there shall be no overflow angle or sharp edges.
4.4 Surface roughness
The surface roughness parameter Ra of metallic clip shall not exceed 0.8 µm.
4.5 Hardness
The hardness of silver metallic clip is 60 HV0.1~100 HV0.1, and the hardness of titanium metallic clip is 100 HV0.2 ~ 200 HV0.2.
4.6 Toughness
Metallic clip shall have good toughness and must not be cracked or broken after being clamped.
4.7 Clamping performance
Metallic clip and clamp are well-matched; it shall have good clamping performance.
4.8 Corrosion resistance Metallic clip shall have good
corrosion resistance; it will not produce any corrosion trace after test.
4.9 Sterility
If metallic clip is delivered in sterilized state, it shall undergo an effective and validated sterilization process to make the product sterile.
4.10 Microbiological indicators
If metallic clip is delivered in non-sterilized state, its initial contaminating bacteria shall be no more than 100 CFU/piece.
4.11 Ethylene oxide residual amount
If metallic clip is sterilized with ethylene oxide gas, the residual amount shall be no more than 10µg/g.
5 Test methods
5.11 Ethylene oxide residual amount
Test according to the method specified in GB/T 16886.7, which shall comply with the provisions of 4.11.
Preparation of the test solution. metallic clip and test water are configured in a ratio of 1.0g. 5mL, sealed at a temperature of 60° C ±1° C, and balanced for 40 minutes.
5.12 Packaging seal test
Conduct a test according to the method specified in YY/T 0681.4, which shall comply with the provisions of 4.12.
5.13 Biocompatibility
5.13.1 Cytotoxicity test
Conduct a test according to the method specified in 8.2 (extract liquid test) of
GB/T 16886.5-2003, which shall meet the requirements of 4.13.1.
5.13.2 Sensitization, intradermal reaction test
Conduct a test according to the methods specified in 7.4 (maximum dose test) and Appendix B (intradermal reaction test) of GB/T 16886.10-2005, which shall meet the requirements of 4.13.2 and 4.13.3.
5.13.3 Evaluation of systemic and sub-chronic toxicity
Evaluate according to the method specified in chapter 5 and chapter 6 of GB/T 16886.11-2011, which shall meet the requirements of 4.13.4 and 4.13.5.
5.13.4 Genetic toxicity evaluation
Evaluate according to the method specified in 4.4 of GB/T 16886.3-2008, which shall meet the requirements of 4.13.6.
5.13.5 Implantation reaction Evaluation
Evaluate according to the method specified in 3.4 of GB/T 16886.6-1997, which shall meet the requirements of 4.13.7.
5.13.6 Bacterial endotoxin
Evaluate according to the method specified in GB/T 14233.2-2005, which shall meet the requirements of 4.13.8..
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 4135Silver ingots
- GB/T 4340.1Metallic materials - Vickers hardness test - Part 1: Test method
- GB/T 6682Water for analytical laboratory use - Specification and test methods
- GB/T 13810-2007Wrought titanium and titanium alloys for surgical implants
- GB/T 14233.2-2005Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods
- GB/T 16886.3-2008Biological evaluation of medical devices -- Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
Editions of YY/T 0079
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0079-2016 | Metallic clip | current edition | Current |
| YY/T 0079-2006 | Metallic clip | previous edition | Superseded |
This page sells the current edition, YY/T 0079-2016. Earlier editions are listed for reference only.
How to Download YY/T 0079-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 14233.2-2005 — Test methods for infusion, transfusion, injection equipment for medical use-Part 2: Biological test methods
GB/T 16886.7-2015 — Biological evaluation of medical devices—Part 7: Ethylene oxide sterilization residuals
GB/T 4135-2016 — Silver ingots
Secure payment via Stripe
Payments accepted
YY/T 0079-2016
$105.00