YY 1468-2016Oxygen concentrator supply systems for use with medical gas pipeline systems (English PDF)
用于医用气体管道系统的氧气浓缩器供气系统
This is a limited preview
Buy now to download the full PDF (35 pages)
Need quotation for your management?
Get a formal quotation for YY 1468-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
January 26, 2016
Implementation date
January 1, 2018
Scope
YY 1468-2016 is the English-translated version of 用于医用气体管道系统的氧气浓缩器供气系统.
YY 1468-2016 is the Chinese pharmaceutical industry standard for the oxygen concentrator supply system feeding a medical gas pipeline system built to ISO 7396-1 or GB 50751-2012.
It specifies the design and installation requirements for such a supply system and for its compressor, covering the oxygen concentration delivered and its monitoring, the supply pressure and the reserve, the alarms and the safety devices, the filtration and the drying of the gas and the testing and commissioning of the installation.
It applies only to systems producing oxygen-enriched air of 93 % oxygen, and it does not cover oxygen concentrators for use at home. Issued on 26 January 2016, it applies from 1 January 2018.
Found what you need?
Document preview — YY 1468-2016
National Standard of the People's Republic of China
- ICS
- 11.040.10
- Classification
- C 46
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Preface
- Introduction
- 1 Scope
- 2 normative reference document
- 3 Terms and definitions
- 4 General requirements
- 4.1 Safety and continuity of gas supply
- 4.2 * alternative structure
- 4.3 Materials
- 4.4 System Design
- 4.5 Oxygen-enriched air (93% oxygen) index
- 5 supply source
- 5.1 General requirements
- 5.2 Main supply air supply
- 5.3 Auxiliary supply source
- 5.4 standby air supply
- 5.5 Gas supply with gas cylinders
- 5.6 Oxygen concentrator supply system location
- 6.1 Oxygen concentrator unit
- 6.2 Oxygen-enriched air (93% oxygen) tank
- 6.3 Oxygen Analyzer
- 6.4 Pressure relief valve
- 6.5 globe valve
- 6.6 sampling port
- 6.7 Pressure regulator
- 7 Monitoring and alarm system
- 7.1 General requirements
- 7.2 Monitoring and alarm signals
- 7.3 Operation alarm
- 7.4 Information signal
- 8 mark
- 9 Installation
- 9.1 General requirements
- 9.2 Electrical system
- 10 test, test run and proof
- 10.1 General requirements
- 10.2 Testing and Procedures
- 10.3 Performance Testing
- 10.4 commissioning and certification
- 11 Information provided by the manufacturer
- 11.1 Installation Guide
- 11.2 User's Manual
- 11.3 Preventive Maintenance Guide
- 11.4 Operation and management information
- 11.5 Completed drawings
- 11.6 Electrical Schematic Figure
- 11.7 Manufacturer Announces
- 12 Oxygen-enriched air (93% oxygen) used in the implementation of
- 12.1 Acceptance of oxygen-enriched air (93% oxygen)
- 12.2 Timing
- 12.3 Oxygen-enriched air (93% oxygen) and oxygen mixture
- 12.4 Calibration of Medical Devices
- 12.5 mark
- 12.6 Complies with ISO 7396-1 or GB 50751-2012
- Appendix A (informative) Oxygen concentrator gas supply system
- Appendix B (informative) General guidelines for site selection of gas supply systems
- Appendix C (informative) Guidelines for Emergency Procedures
- Appendix D (informative) Test and commissioning procedures
- Appendix E (informative) Typical form for proof of gas supply system for oxygen concentrators
- Appendix F (informative) Minimum requirements recommended for preventive maintenance
- Appendix G (informative) Installation proposal
- Appendix H (informative) Risk and risk management
- Appendix I (informative) Size and capacity recommendations
- Appendix J (informative) Basic principles Reference
Foreword
The technical content of this standard is mandatory.
This standard is drafted in accordance with the rules given in GB/T 1.1-2009.
This standard uses the re-drafting method to modify the use of ISO 10083.:2006 "for the medical gas pipeline system oxygen concentrator supply system System ".
The technical differences between this standard and ISO 10083.:2006 are as follows.
--- on the normative reference documents, the standard made a technical differences in the adjustment to meet China's technical conditions, adjust the situation The situation is reflected in Chapter 2, "Normative references", as follows.
- Increase the normative reference document GB 50751-2012 (see 4.4, 5.12, 7.1.1.1, 12.6);
- Increase the normative reference document WS1-XG-008-2012 (see 4.5.1, 10.3.4.2.2);
Replace ISO 21969 with YY/T 1440 (YY/T 1440-2016, ISO 21969.:2009, MOD) (see 5.5.4);
Replace GB 5145 (see 5.5.5) with GB 15383 (GB 15383-2011, ISO 5145.:2004, NEQ);
Replace ISO 10524-2 (see 6.7) with YY/T 1439.2 (YY/T 1439.2-2016, ISO 10524-2.:2005, MOD);
--- Removal of the relevant contents of the filling (ISO 10083.:2006, 4.6, 7.3.1i), 11.2, item 4, item 6 and appendix J.
- Revised Oxygenated Air (93%) in Table E.1 according to the relevant requirements of WS1-XG-008-2012. This standard makes the following editorial changes to ISO 10083.:2006.
- to add the relevant statement in the second paragraph of the introduction;
- delete or modify the contents of ISO 10083.:2006 relating to "country or region"; Fixed an editorial error in ISO 10083.:2006 (see Figure A.2);
--- reference to the international standards involved in the literature, there are corresponding to be used as national standards and industry standards, the standard to the corresponding national standard The quasi-industry standard is used as a reference and adjusts the order of the references.
Introduction
This standard is used to specify the supply of oxygen-enriched air (93% oxygen) oxygen concentrator gas supply system to the medical gas piping distribution system (eg medical Molecular sieve oxygen system) safety and performance of the minimum requirements. Specifies the minimum oxygen concentration of the gas supply system for the gas concentrator gas supply system.
Oxygen concentrators can be used to deliver oxygen to the medical gas piping system as a substitute for medical oxygen (eg, "oxygen" in Chinese Pharmacopoeia)
Air (93% oxygen), oxygen-enriched air (93% oxygen) is known in the US Pharmacopoeia as "oxygen93percent", "oxygen 93%", in the European Pharmacopoeia Is referred to as "oxygen (93percent)", "oxygen (93%)", and the oxygen concentrator is available with an air supply source containing 100% Cylinders or cryogenic containers).
Depending on the characteristics of the oxygen concentrator and the supply flow, the oxygen concentrator can supply a change in the oxygen concentration within a specified range Product gas.
The decision to use oxygen-enriched air (93% oxygen) should be made by the health care organization at an early stage and not within the scope of this standard. Whether to use rich Oxygen air (93% oxygen) and oxygen mixture is also determined by the health care agency. The use of a gas supply system containing an oxygen concentrator may require a country Authority approval.
This standard should not be considered a recognition or recommendation for another oxygen concentration.
The gas supply system with an oxygen concentrator can be installed at the time of installation of the piping distribution system or as an alternative to the existing gas supply system plus. A gas supply system with an oxygen concentrator can be provided as a complete set of equipment and can be installed by a third party. In this case, connect the gas supply The system to the piping distribution system and before use, the manufacturer of the oxygen concentrator gas supply system must provide the installer with the appropriate installation and testing When the information. The purpose of this standard is to ensure that.
- Oxygen concentrator gas supply system is properly introduced into the health care facility;
- the quality of oxygen-enriched air (93% oxygen) delivered by the gas supply system;
--- Oxygen-enriched air (93% oxygen) continuous supply;
--- use the right material;
--- cleaning of components;
--- proper installation;
- Provide appropriate control, monitoring and alarm systems for the gas supply system;
--- test, test run and prove.
This standard is intended for use by the design, construction, inspection or operation of the health care organization. Those involved in the design, manufacture, calibration or testing It is also clear that the personnel of the equipment connected to the piping system supplied to the gas supply system of the oxygen concentrator should also be aware of the contents of this standard.
Appendix J contains a description of the requirements of certain requirements of this standard. Provides a further explanation of the principles of this standard that contain requirements and recommendations.
The chapter marked with * under the chapter number corresponds to the principle in Appendix J. The basic principles of such requirements are not only conducive to the correctness of this standard Use, and can simplify the subsequent changes.
1 Scope
1.1 This standard specifies the oxygen concentration of a medical gas piping distribution system for compliance with ISO 7396-1 or GB 50751-2012 Design and installation requirements for the compressor gas supply system.
1.2 This standard applies only to the production of oxygen-enriched air (93% oxygen) oxygen concentrator gas supply system (see 4.5.1).
1.3 This standard does not include oxygen concentrators for use at home.
Note. The requirements for the use of oxygen concentrators at home are specified by YY 0732.
2 Normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.
GB 15383 Cylinder valve outlet connection type and dimensions (GB 15383-2011, ISO 5145.:2004, NEQ) Technical specification for medical gas engineering GB 50751-2012
YY/T 0316 Medical Device Risk Management for Medical Devices (YY/T 0316-2008, ISO 14971.:2007, IDT)
YY/T 0882 Compatibility of anesthesia and breathing equipment with oxygen (YY/T 0882-2013, ISO 15001.:2003, IDT)
YY/T 1439.2 Medical gas pressure regulators - Part 2. Busbar pressure regulators and piping pressure regulators (YY/T 1439.2-2016, ISO 10524-2.:2005, MOD)
YY/T 1440 High pressure flexible connection for use with medical gas systems (YY/T 1440-2016, ISO 21969.:2009, MOD)
WS1-XG-008-2012 Oxygen-enriched air (93% oxygen)
ISO 7396-1 Medical gas piping systems - Part 1. Medical compressed gas and vacuum piping (Medicalgaspipeline systems-Part 1. Pipelinesforcompressedmedicalgasesandvacuum)
ISO 14644-1.:1999 Clean rooms and associated control environments - Part 1. Air cleanliness classifications (Cleanroomsandassoci-atedcontroledenvironments-Part 1. Classificationofaircleanliness EN 286-1 Simple, non-subject pressure vessel for storage of air or nitrogen Part 1. General pressure vessel (Simpleunfired pressurevesselsdesignedtocontainairornitrogen-Part 1. Pressurevesselsforgeneralpurposes
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 Trial run commission
Functional proof to verify that the agreed system characteristics are met and accepted by the user or his representative.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 35 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 15383Connection types and dimensions for gas cylinder valve outlets
- YY/T 0316Medical devices-Application of risk management to medical devices
- YY/T 0882Anaesthetic and respiratory equipment—Compatibility with oxygen
- YY/T 1439.2Pressure regulators for use with medical gases - Part 2: Manifold and line pressure regulators
- YY/T 1440High-pressure flexible connections for use with medical gas systems
WS1-XG-008-2012 · ISO 7396-1 · ISO 14644-1
Similar standards
How to Download YY 1468-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 15383-2011 — Connection types and dimensions for gas cylinder valve outlets
YY/T 0316-2003 — Medical devices - Application of risk management to medical devices
YY/T 0316-2008 — Medical devices - Application of risk management to medical devices
Secure payment via Stripe
Payments accepted
YY 1468-2016
$510.00