YY 0847-2011Medical endoscopes - Endotherapy device - Stone retrieval baskets (English PDF)
医用内窥镜 内窥镜器械 取石网篮
This is a limited preview
Buy now to download the full PDF (16 pages)
Need quotation for your management?
Get a formal quotation for YY 0847-2011 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
December 31, 2011
Implementation date
June 1, 2013
Scope
YY 0847-2011 is the English-translated version of 医用内窥镜 内窥镜器械 取石网篮.
YY 0847-2011 is the Chinese pharmaceutical industry standard for the stone retrieval basket, the endotherapy device passed through an endoscope to capture and remove a stone.
It gives the scope, the terms and definitions, the requirements and the test methods, covering the dimensions of the basket and of the sheath, the opening and closing of the wires and the force needed for it, the tensile strength of the basket and of its joints, the corrosion resistance and the surface finish, the materials and their biological evaluation and, where the device is supplied sterile, its sterility.
It applies to baskets sold as a product in their own right. Issued on 31 December 2011, it has applied since 1 June 2013.
Found what you need?
Document preview — YY 0847-2011
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions5
- 4 Requirements5
- 5 Test methods9
- 6 Inspection rules12
Foreword
This Standard was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This Standard specifies the scope, terms and definitions, requirements and test methods of stone retrieval forceps.
This Standard is applicable to stone retrieval forceps that is used for endoscopic surgery as an independent product.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 1962.1-2001 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1. General requirement
GB/T 1962.2-2001 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2. Lock fittings
GB 4234-2003 Stainless steel for surgical implants
GB/T 14233.1-2008 Test methods for infusion transfusion injection equipment for medical use - Part 1. Chemical analysis methods
GB/T 14233.2-2005 Test methods for infusion, transfusion, injection equipment for medical use - Part 2. Biological test methods
GB/T 16886.1 Biological evaluation of medical devices - Part 1. Evaluation and testing within a risk management process
YY/T 0149-2006 Medical instruments of stainless steel - Test methods of corrosion resistance
YY 0294.1-2005 Surgical instruments - Metallic materials - Part 1. Stainless steel
YY/T 1052-2004 Marking requirements for surgical instruments
3 Terms and definitions
The following terms and definitions are applicable to this document.
3.1 Wire basket
A mesh basket that is used for taking stone, foreign matter or tissue sample.
3.2 Stone retrieval forceps
An endotherapy device that works with designated endoscope, which is mainly composed of handle, slender guiding-rope and guiding-duct, and wire basket on the head. It traps the stone, foreign matter or tissue sample via wire basket's circle; and firmly catches the stone, foreign matter or tissue sample through contraction of wire basket.
3.3 Insertion portion
A part of stone retrieval forceps which can be inserted into human nature orifice, surgical incision, endoscope's device channel or endotherapy device's channel.
3.4 Maximum insertion portion width
The maximum external width of insertion portion.
3.5 Working length
The maximum length of insertion portion of stone retrieval forceps.
3.6 Endotherapy device
A medical device, which is used in endoscopic surgery, goes into human nature orifice or surgical incision through the same or different channel with endoscope, so as to realize examination, diagnosis or treatment.
Note 1. Endotherapy device includes some devices that lead the endotherapy device to insert.
It also includes some devices that are used under observation of endoscope and go from human opening which is different from the endoscope, for safety purpose.
4 Requirements
4.1 Material used in the part of contacting with patients
4.1.1 Requirements for chemical composition
Material used in the part of contacting with patients shall be clearly indicated in any possible form by the manufacturer. Metal material shall be marked with trademark and/or code as well as the material's chemical composition requirements. Metal material's chemical composition shall be verified by experiment.
4.1.2 Biocompatibility
For material used in the part of contacting with patients, it shall conduct biological safety evaluation according to principles and requirements in GB/T 16886.1, so as to prove that it has good biocompatibility.
Biological evaluation shall consider the results of biological test. The selection of test items are conducted according to the guideline of GB/T 16886.1.
Relevant parts of GB/T 16886 are preferred for all the tests.
For material that has previously been proved as applicable, if it can prove that the following manufacturing process is insufficient to produce any biological safety hazards, then it may be exempted from repeated biological test.
Note 1. When device's material in design has demonstrable history of using in specific application OR any obtainable material's and/or device's information from other aspects, it shall be regarded that the material has been proved as applicable.
Note 2. If metal material uses medical metal material that is suitable in application scope of national or industrial standard, it may be exempted from repeated biological test.
4.1.3 Polymeric material's
dissolving precipitates in the part of contacting with patients
4.1.3.3 Heavy metal. leachables are less than 5.0 µg/mL
4.1.3.2 pH. pH difference is less than 2.0
4.1.3.1 Surface / appearance
Colorless and transparent, no foreign matter, inspected by visual inspection.
4.1.3.4 Potassium permanganate reducing substances. consumption
difference with standard test solution is less than 2.0 mL.
4.1.3.5 Evaporation residue. less than 2.0 mg
4.2 Appearance
4.2.1 Except for special purpose, there shall not have any possible safety-harm on the outer surface.
4.2.2 There is no floss, loosing, other obvious abnormalities or damage for wire
5 Test methods
5.3.4 Use a universal measuring tool to measure, according to requirements in 4.3.4.
5.4 Performance tests
5.4.2 Coil the insertion portion into a circle or arc of which the radius is 10 cm; push the handle; the stone retrieval forceps shall be open and close smoothly.
5.4.3 Coil the insertion portion into a circle or arc of which the radius is 10 cm; open the stone retrieval forceps; turn the handle; the stone retrieval forceps shall rotate smoothly.
5.4.1 Tensile strength test
Apply the nominal tension, provided by the manufacturer, to the whole stone retrieval forceps. No damage or fracture on each part of stone retrieval forceps.
5.5 Tolerance tests
5.5.1 High temperature and high pressure resistance test
After 20 tests of high temperature and high pressure resistance, it shall still comply with requirements in 4.2 and 4.4.
5.5.2 Corrosion resistance test
Conduct boiling water test in YY/T 0149-2006 for the portion that uses stainless steel, according to requirements in 4.5.2. For other materials, it shall conduct once according to the sterilization method, as designated in the instructions for use by the manufacturer. There shall have no corrosion.
5.5.3 Repeated operation resistance
For operation of actual-practice simulation, it shall comply with requirements in 4.5.3.
5.6 Sterility requirement test
5.6.1 Conduct in the method specified in GB/T 14233.2-2005, according to requirements in 4.6.1.
5.6.2 Conduct in gas chromatography specified in GB/T 14233.1-2008, according to requirements in 4.6.2.
5.7 Luer fitting test
Conduct according to relevant tests in GB/T 1962.1-2001, GB/T 1962.2-2001, and requirements in 4.7. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 1962.1-2001Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirement
- GB/T 1962.2-2001Conical fittings with a 6 %(Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
- GB/T 14233.1-2008Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods
- GB/T 14233.2-2005Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- YY/T 0149-2006Medical instruments of stainless steel - Test methods of corrosion resistance
GB 4234-2003 · YY 0294.1-2005
Similar standards
How to Download YY 0847-2011
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 14233.1-2008 — Test Methods for Infusion, Transfusion, Injection Equipment for Medical Use — Part 1: Chemical Analysis Methods
GB/T 14233.2-2005 — Test methods for infusion, transfusion, injection equipment for medical use-Part 2: Biological test methods
GB/T 16886.1-2022 — Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
Secure payment via Stripe
Payments accepted
YY 0847-2011
$125.00