YY 0581.1-2011Infusion access adapters - Part 1: Needle access adapters(Heparin plugs) (English PDF)
输液连接件 第1部分:穿刺式连接件(肝素帽)
This is a limited preview
Buy now to download the full PDF (23 pages)
Need quotation for your management?
Get a formal quotation for YY 0581.1-2011 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
December 31, 2011
Implementation date
June 1, 2013
Scope
YY 0581.1-2011 is the English-translated version of 输液连接件 第1部分:穿刺式连接件(肝素帽).
YY 0581.1-2011 is the Chinese pharmaceutical industry standard for the needle access adapter, the heparin plug through which a needle enters an infusion line.
It is the first part of the YY 0581 series and sets the requirements and the test methods for these connections, covering the dimensions and the conical fittings, the number of punctures the septum must take without leaking, the resealing and the freedom from fragmentation, the strength of the joints, the materials and their biological evaluation and sterility.
A device supplied integral with an indwelling intravascular device may follow this part as a reference. Issued on 31 December 2011, it has applied since 1 June 2013.
Found what you need?
Document preview — YY 0581.1-2011
National Standard of the People's Republic of China
- ICS
- 11.040.20
- Classification
- C 31
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- Introduction
- 1 Scope
- 2 Normative references
- 3 Mark
- 4 Materials
- 5 Physical Requirements
Foreword
YY 0581 "infusion connector" is divided into two parts.
--- Part 1. piercing connector (heparin cap);
--- Part 2. needle-free connectors. This is Part 1 YY 0581 of. This section drafted in accordance with GB/T 1.1-2009 given rules.
YY 0581 This part of Amendment YY 0581-2005. The main difference between this part and the YY 0581-2005 YY 0581 are.
--- Modify the standard name;
--- Modify the chemical test methods;
Increase the volume of the cavity --- labeling requirements on a single package.
Introduction
Common market infusion into piercing connectors (also known as pin connectors have) and non-penetrating formula (also known as needle-free connection) of two types. foreign On the market there is a blunt plastic pin connector type connections. Interface and after use to maintain its sealing its total requirement. Piercing Connector interface for patients with puncture risk off the crumbs into the infusion system, as well as medical staff have needle prick accident risk; non-punch Barbed connector member such risks piercing connection does not exist, but there is the risk of multiple uses microbial invasion infusion system.
Infusion connector YY 0581 relates generally intravascular long-term use of indwelling catheters (eg intravenous catheter, central venous catheters)
With the use, through which the liquid to the intravascular infusion and extraction of blood or liquid. Whether the use of needles or needle-free connection is connected, at the output At the end of the infusion liquid or liquid between different, it must be made in physiological saline via infusion catheter flushing pipe fittings, flush left in the catheter Drugs to prevent complications such as phlebitis. At the end of the infusion with saline or heparin saline through Pin connector, or needleless connectors on the catheter forward sealed tube to prevent backflow of blood transfusion due to pull out when the appliance caused.
Prolonged use in connection with intravascular indwelling catheter if improper connections between, there is the risk of accidental loss, which may cause Patients with massive bleeding into the patient and endanger their lives. Therefore, the lock connector and require clinical use is to maintain a reliable connection Necessary, recommendations on joint injection or related product packaging or instructions for use on the reliability of the connection holding given the necessary warnings.
1 Scope
YY 0581 This section provides puncture infusion connectors (also known as heparin cap, hereinafter referred to as. connections) requirements and test methods.
Connection with indwelling intravascular devices (such as intravenous catheter) into one integrated product supply, can refer to the implementation of this section.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
GB/T 1962.2 syringes, needles and other medical equipment - 6% (Luer) taper fittings - Part 2. Lock fittings
GB 8368 disposable infusion set gravity transfusion formula
GB 15810 single-use sterile syringes
GB 15811-time use of sterile needles Symbols - Part 1 YY/T 0466.1 medical equipment used with medical device labels, labeling and information. General requirements
3 Mark
Infusion connector marked by the text and description of the part number, numbered. Connection YY 0581.1
4 Materials
Material connecting member shall meet the requirements specified in Chapter 5, 6 and 7.
5 Physical Requirements
5.1 Particulate pollution
Liquid passage surface of the connecting member should be clean. When you press A.1 prescribed test shall not exceed the pollution index.
5.2 Connection strength
Press A.2 predetermined test connections and all connections between components should be able to withstand at least 15N static pull 15s.
5.3 Leak
Connections shall not permeable to air and liquid. When you press the test A.3 provisions, there should be no air or water leaks.
5.4 Connector
If there is a 6% (Luer) taper joint on the connecting member for use in connection with indwelling catheters in the other vessel, which should be a joint operator
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Normative references
Cited by
- GB/T 19335-2022Blood flow products for single use — General specification
- YY/T 1287.1-2016Intracranial external drainage system - Part 1: Collecting sets for external drainage by intracranial puncture
- YY/T 1287.2-2016Intracranial external drainage system - Part 2: Collecting sets for external drainage by lumbar puncture
Similar standards
How to Download YY 0581.1-2011
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.