YY 0065-2016Ophthalmic instruments - Slit-lamp microscopes (English PDF)

眼科仪器 裂隙灯显微镜

Preview PDF

This is a limited preview

Buy now to download the full PDF (25 pages)

Need quotation for your management?

Get a formal quotation for YY 0065-2016 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Mandatory

Issue date

January 26, 2016

Implementation date

January 1, 2018

Scope

YY 0065-2016 is the English-translated version of 眼科仪器 裂隙灯显微镜.

YY 0065-2016 is the Chinese pharmaceutical industry standard for the slit-lamp microscope used in ophthalmic examination, applying together with ISO 15004-1 and ISO 15004-2:2007.

It sets the requirements and the test methods for the instrument, covering the magnification and the field of view of the microscope, the width, height and illuminance of the slit and their adjustment, the filters, the alignment of the illumination and the observation systems and the mechanical movements of the stand.

It applies to universal slit-lamp microscopes and not to accessories such as cameras and lasers, and where it differs from the ISO standards it takes precedence. Issued on 26 January 2016, it has applied since 1 January 2018.

Found what you need?

Document preview — YY 0065-2016

National Standard of the People's Republic of China

Replacing
YY 0065-2007

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Terms and definitions5
  • 4 Requirements6
  • 5 Test methods9
  • 6 Accompanying files15
  • 7 Marks15

Foreword

This standard was drafted in accordance with the rules given in GB/T 1.1-2009.

This standard replaces YY 0065-2007 "Ophthalmic instruments - slit-lamp microscopes"; compared with YY 0065-2007, the main technical changes of this standard are as follows.

-- Add the requirements for focus difference between left and right observation systems (see Table 1, serial number 7);

-- Add the requirements for optical radiation hazards (see 4.4);

-- Delete the requirements for minimum spot size (2007 version 4.1.3);

-- Delete the requirements for slit spot illuminance (2007 version 4.1.6);

-- Delete the requirements for diameter of field of view (2007 version 4.1.5);

-- Delete the requirements for mechanical movement (2007 version 4.2);

-- Delete the requirements for environmental test (2007 version 4.5).

This standard uses the redrafted law to amend and adopt ISO 10939:2007 "Ophthalmic Instruments - slit-lamp microscopes".

Compared with ISO 10939:2007, there are technical differences in this standard. The terms involved in these differences have been indicated by a vertical single line (|) in the marginal position on its outer side. The technical differences between this standard and ISO 10939:2007 and its causes are as follows.

-- With regard to normative references, this standard has made adjustments with technical differences to adapt to China's technical conditions. The adjustments are centrally reflected in Chapter 2 "normative references" and the specific adjustments are as follows.

- Use GB 9706.1 which is adopted to be equivalent to international standard to replace IEC 60601-1;

-- Add "this standard applies to universal slit-lamp microscopes" in the scope of Chapter 1.

-- Add Chapter 5 test methods. As the recommended test methods, it makes the requirements of this standard operable. There is no test method in ISO 10939:2007.

1 Scope

This standard, together with ISO 15004-1 and ISO 15004-2:2007, specifies the requirements and test methods for slit-lamp microscopes.

This standard applies to universal slit-lamp microscopes. This standard does not apply to slit-lamp microscope accessories, such as camera equipment and lasers. If there are differences, this standard takes precedence over ISO 15004-1 and ISO 15004-2:2007.

2 Normative references

The following documents are essential for the application of this document. For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.

GB 9706.1 Medical electrical equipment - Part 1. General requirements for safety (GB 9706.1-2007, IEC 60601-1:1988, IDT)

ISO 15004-1 Ophthalmic instruments - Fundamental requirements and test methods - Part 1. General requirements applicable to all ophthalmic instruments

ISO 15004-2:2007 Ophthalmic instruments - Fundamental requirements and test methods - Part 2. Light hazard protection

3 Terms and definitions

The following terms and definitions apply to this document.

3.1 Slit-lamp microscope

It is an instrument that consists of microscope and rotating illumination system which is capable of producing slit image.

e) At the highest visual angular magnification, the resolution of field-of-view center shall be at least 1800Nlp/mm (N is numerical aperture).

4 Requirements

4.3.2 High eye point eyepiece

If the manufacturer claims to be high eye point eyepiece, the distance between the exit pupil of observation system and the closest part of eyepiece shall be no less than 17mm.

4.4 Optical radiation hazard of slit-lamp microscopes

Slit-lamp microscopes shall meet the requirements of ISO 15004-2:2007 for light hazard protection.

Slit-lamp microscopes shall be classified as Class-1 instruments or Class-2 instruments according to ISO 15004-2:2007 Chapter 4. ISO 15004-2:2007 applied to slit-lamp microscopes is as follows.

a) For Class-1 slit-lamp microscopes.

1) 5.1, 5.2 and 5.4 of ISO 15004-2:2007 apply.

2) The test methods of 6.1, 6.2, 6.4 and 6.5 of ISO 15004-2:2007 apply.

3) If it is determined to be Class-1 instruments, there are no more requirements; if it is determined not to belong to Class-1 instruments, more requirements specified in b) shall apply.

b) For Class-2 slit-lamp microscopes.

1) 5.1, 5.3 and 5.5 of ISO 15004-2:2007 apply.

2) The test methods of 6.1, 6.2, 6.3, 6.4, 6.5 and 6.6 of ISO 15004-2:2007 apply.

3) ISO 15004-2:2007 Chapter 7 applies.

If the intended use of slit-lamp microscopes includes the use of an accessory 90D lens, measure the related exposure of cornea and lens. The 90D lens (such as Volk lens) shall be placed at 7mm position behind the focal plane of slit-lamp with the largest illuminated area. Exposure measurement is taken at the minimum dimension point of the illuminated area 7mm behind the 90D lens.

5 Test methods

5.6 Errors of the focal plane of lighting system with respect to mechanical axis of rotation Place the reticle with adjustable front-to-back distance on object plane; adjust the width of slit to be minimized; use the microscope with intermediate visual angular magnification to observe and adjust it to the zero-line coincidence of slit image and reticle. As shown in Figure 1, the lighting system and the observation system rotate within the left (right) <= 45. When the rotation angle is alpha, if the slit image deviates from the reticle zero-line, read out the lateral deviation value (deltaa)alpha from the reticle; then adjust the reticle so that the slit image coincides with the zero-line of the reticle. Record the axial distance deltaa before-after adjustment of the reticle. The values of (deltaa)alpha and deltaa measured shall be in accordance with the regulations of No.5 in Table 1.

5.1 Overview

microscope) to focus reticle, and the image is clear. Measure the maximum amount of movement between the two after changing the magnification. The result shall meet the requirements of No. 4 in Table 1. […]

5.7 Confocal errors of observation system and lighting system at slit surface

Use visual angular magnification to firstly adjust slit image to vertical position and record the clearest diopter-value of the slit image. When the rotating slit is tilted at any angle, use simple eye to observe; adjust the diopter loop of left (right) eyepiece to have clear slit; measure the displacement amount of left (right) eyepiece; use the same method to measure the diopter difference of left (right) eyepiece when it is at the highest and lowest visual angular magnification. The measured maximum displacement amount shall be in accordance with the provisions of No. 6 in Table 1.

5.8 Focus difference of left and right observation system

Adjust the eyepiece diopter of slit-lamp microscope to zero and measure it at the highest visual angular magnification. Regard cross reticle as test object, adjust the focus of the cross division at left (right) observation system of slit-lamp microscope, and then use the right (left) observation system to adjust the focus of the cross division and read out the displacement amount. The result shall meet the requirements of No.7 in Table 1.

5.9 Eyepiece

5.9.1 Zero diopter error of eyepiece

Use diopter-meter to test and the result shall meet the requirements of No.8 1) in Table 1.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY 0065

EditionTitleRevisionStatus
YY 0065-2016Ophthalmic instruments - Slit-lamp microscopescurrent editionCurrent
YY 0065-2007Ophthalmic instruments - Slit-lamp microscopesprevious editionSuperseded

This page sells the current edition, YY 0065-2016. Earlier editions are listed for reference only.

How to Download YY 0065-2016

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
25 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 0065-2016

$75.00