YY 0055-2018Dentistry—Powered polymerization activators (English PDF)
牙科学 光固化机
This is a limited preview
Buy now to download the full PDF (25 pages)
Need quotation for your management?
Get a formal quotation for YY 0055-2018 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
September 28, 2018
Implementation date
April 1, 2020
Scope
YY 0055-2018 is the English-translated version of 牙科学 光固化机.
YY 0055-2018 is the Chinese pharmaceutical industry standard for the powered polymerization activator used in dentistry, the curing light that hardens polymer-based restorative materials in the mouth in the wavelength range from 385 nm to 515 nm.
It sets the requirements and the test methods, covering the emitted spectrum and the irradiance at the tip, the uniformity of the beam, the timer, the heat produced and the eye protection, for quartz tungsten halogen lamps and light emitting diode units powered internally by a rechargeable battery or from the mains.
It does not apply to laser or plasma arc equipment, nor to units used in the dental laboratory. Issued on 28 September 2018, it applies from 1 April 2020.
Found what you need?
Document preview — YY 0055-2018
National Standard of the People's Republic of China
- ICS
- 11.060.20
- Classification
- C 33
- Replacing
- YY 0055.1-2009
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions
- 4 Category
- 5 Requirements
- 5.1 Overview
- 5.2 Irradiance
- 5.3 Electrical Safety Requirements
- 6 sampling
- 7 Test method
- 7.1 Overview
- 7.2 Irradiance
- 8 Information provided by the manufacturer
- 8.1 Instruction Manual
- 8.2 Technical Specification
- 9 mark
- 10 Packing
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
This standard replaces YY 0055.1-2009 and YY 0055.2-2009, and YY 0055.1-2009 and YY 0055.2-2009 In addition to the editorial changes, the main technical changes are as follows.
--- Added the expression "This standard does not apply to laser or plasma arc equipment" (YY 0055.1-2009 and 1 of YY 0055.2-2009, 1) of this standard;
--- Increased irradiance test requirements for 90% and 110% rated voltage of type 1 LED curing light (YY 0055.2-5.2 of 2009, 5.2 of this standard;
--- Modified the working wavelength range from 385nm to 515nm (YY 0055.1-2009 and YY 0055.2-2009) 5.2.1, 5.2.1) of this standard;
--- Modified UV wavelength range from 200nm to 385nm (YY 0055.1-2009 and YY 0055.2-2009 5.2.2, this Standard 5.2.2);
--- Revised the irradiance requirement of the UV wavelength range of the LED lamp to be.200mW/cm2 (5.2.2 of YY 0055.2-2009, 5.2.2) of this standard;
---Modified electrical safety requirements. from the application of GB 9706.1 part of the provisions should be in accordance with GB 9706.1, YY 0505 and IEC 80601-2-60 requirements and the addition of "if the curing light can be connected or combined with other equipment such as dental treatment machines, then The curing light should meet the requirements of GB 9706.1, YY 0505 and IEC 80601-2-60 in the connected or combined state.
The light curing machine can be connected to other equipment such as dental treatment machines, and the test should be carried out under the condition of connection." (YY 0055.1-2009 and YY 0055.2-2009 5.3, 5.3 of this standard);
--- Increased sampling requirements. According to new products in the industry, new "light curing devices without light guides are used.
Expression of "test under conditions" (Chapter 6 of YY 0055.1-2009 and YY 0055.2-2009, No. 6 of this standard chapter);
--- Removed the terms of pre-processing, repair and improvement in the 7.1 general test clauses (YY 0055.1-2009 and YY 0055.2-7.1 of 2009, 7.1 of this standard;
--- Modified the test 190nm quartz filter to 200nm quartz filter, no longer use 400nm filter (YY 0055.1-7.2.1 of 2009 and YY 0055.2-2009, 7.2.1 of this standard;
--- Revised the wavelength range of the radiometer from 190nm to 1100nm to 200nm~1100nm (YY 0055.1-2009 and 7.2.1 of YY 0055.2-2009, 7.2.1 of this standard;
--- Revised the irradiance test procedure form to clarify the test time as "10s or the time specified by the manufacturer" (YY 0055.1-2009 And 7.2.2 of YY 0055.2-2009, 7.2.2 of this standard);
---Modified information and marking requirements provided by the manufacturer (Chapter 8 of YY 0055.1-2009 and YY 0055.2-2009, this Chapter 8 of the standard).
This standard uses the redrafting method to modify the use of ISO 10650:2015 "dental light curing machine". This standard and ISO 10650. The main differences in 2015 are as follows.
--- Modified some formatting formats in accordance with GB/T 1.1-2009;
---Delete the ISO preamble;
--- For other international standards cited in this standard, if it has been converted into Chinese standards, this standard will replace the referenced international standard number with Corresponding national or industry standard number, and indicate the adoption relationship in Chapter 2 of this standard;
--5.2.1385nm ~ 515nm (blue light) wavelength range of irradiance. for the specific requirements of the relevant content, delete "this standard is not regulated The repeated expression of the specific irradiance requirement for the wavelength range from 385 nm to 515 nm (preface has been described);
---7.2.1 Device. Considering the actual operation needs of the inspection work, the filter is added according to the YY 0055.1-2009 industry standard index requirements. Thickness requirements of the light sheet;
---6 Sampling. According to the new product situation in the industry, a new "light curing machine without light guiding elements is added, and the test is carried out under normal use conditions. "Expression".
Introduction
This standard specifies the requirements for the optical radiation value of the light curing machine in the wavelength range of 200nm~385nm and the wavelength range above 515nm.
And test methods. For the 385 nm to 515 nm wavelength range, no minimum value is given. 385nm to 515nm wavelength range value Not less than the manufacturer's specified value.
This standard uses a wavelength range based on a cut-off filter. Therefore, the wavelength range of 200 nm to 385 nm includes not only the ultraviolet range, It also includes a near-blue light range with a wavelength around 380 nm. The wavelength range of 385 nm to 515 nm is considered to be the wavelength range of photocuring.
The wavelength range of 515nm~1100nm is the detection limit of the detectors involved in this standard. The test methods provided in this standard cannot be given absolutely.
The logarithmic value does not reflect the energy similar to blackbody radiation. The data obtained is not a true radiance value, but a test provided using this standard.
The value obtained by the method. However, the data obtained by these methods should be used in conjunction with this standard.
This standard cites the general standard for medical electrical equipment safety GB 9706.1, correspondingly, the applicable place marked GB 9706.1 corresponding chapter Article number. Dentistry light curing machine
1 Scope
This standard specifies that in the wavelength range of 385nm~515nm, it is used to irradiate and cure dental clinical polymer-based repair materials. The requirements and test methods for the light curing machine.
This standard applies to quartz tungsten halogen lamps and light-emitting diode (LED) lamps. The light curing unit can be powered by an internal power source (rechargeable battery) or Use a connected external power supply (network power supply). This standard does not apply to laser or plasma arc equipment.
This standard does not apply to light curing machines used in dental laboratories for indirect restorations, veneers, dentures and other oral products. this The relevant provisions of the standard are prioritized over the requirements of GB 9706.1.
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only dated versions apply to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB 9706.1 Medical electrical equipment - Part 1. General requirements for safety (GB 9706.1-2007, IEC 60601-1:1988, IDT) A semiconductor-based illumination source.
GB/T 9937 (all parts) Oral vocabulary [ISO 1942 (all parts)]
YY/T 0466.1 Medical devices for use in medical devices - Labels, labels and symbols for information - Part 1. General requirements (YY/T 0466.1-2016, ISO 15223-1:2012, IDT)
YY 0505 Medical Electrical Equipment - Part 1-2. Safety General Requirements Parallel Standard. Electromagnetic Compatibility Requirements and Test (YY 0505-2012, IEC 60601-1-2:2004, IDT)
YY/T 0628 dental equipment graphic symbol (YY/T 0628-2008, ISO 9687:1993, IDT)
IEC 80601-2-60 Medical electrical equipment - Part 2-60. Particular requirements for basic safety and basic performance of dental equipment (Medical electricalequipment-Part 2-60. Particularrequirementsforthebasicsafetyandessentialperformance Ofdentalequipment)
3 Terms and definitions
The following terms and definitions as defined in GB/T 9937 and GB 9706.1 apply to this document.
Note. This clause corresponds to Chapter 3 of IEC 60601-1:2005 A1.2012.
3.1 Light curing machine
It mainly emits light in the wavelength range of 385nm~515nm, which is used for dental base filling, repairing, and sealing material polymerization. device.
3.2 LED
A semiconductor-based illumination source.
3.3 Fully charged battery
The battery is initially 100% fully charged.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 9937Dentistry - Vocabulary
- YY/T 0466.1Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
- YY 0505Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
- YY/T 0628Dentistry—Graphical symbols for dental equipment
IEC 80601-2-60
Similar standards
Editions of YY 0055
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 0055-2018 | Dentistry—Powered polymerization activators | current edition | Current |
| YY 0055.1-2009 | Dentistry—Powered polymerization activators | previous edition | Superseded |
This page sells the current edition, YY 0055-2018. Earlier editions are listed for reference only.
How to Download YY 0055-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
GB/T 9937.2-2008 — Dental vocabulary -- Part 2: Dental materials
GB/T 9937-2020 — Dentistry - Vocabulary
Secure payment via Stripe
Payments accepted
YY 0055-2018
$285.00