Scope
JJF 1101-2019 (Calibration Specification for Environmental Testing Equipment for Temperature and Humidity Parameters) is available as an English-translated PDF.
JJF 1101-2019 is the Chinese standard "Calibration Specification for Environmental Testing Equipment for Temperature and Humidity Parameters".
Its scope clause reads: This Specification applies to the calibration of temperature and humidity parameters for environmental testing equipment - such as drying ovens, incubators, climatic aging chambers, mold test chambers, salt spray test chambers, corrosive gas test chambers, high/low-temperature test chambers, alternating damp-heat test chambers, and constant temperature and humidity chambers - operating within a temperature range of -80°C~ 300°C and a humidity range of 10% RH~100% RH.
Temperature and humidity parameters of other types of environmental testing equipment may also be calibrated with reference to this Specification. Its clauses include terms; overview; metrological characteristics; calibration conditions; calibration items and calibration methods.
It was issued by the State Administration for Market Regulation on 2019-09-27 and took effect on 2020-03-27.
Found what you need?
Document preview — JJF 1101-2019
National Standard of the People's Republic of China
Issued by: State Administration for Market Regulation
Contents
- Introduction...5
- 1 Scope...7
- 2 Normative references...7
- 3 Terms...7
- 4 Overview...9
- 5 Metrological characteristics...9
- 6 Calibration conditions...10
- 6.1 Environmental conditions...10
- 6.2 Load conditions...10
- 6.3 Measurement standards and other equipment...10
- 7 Calibration items and calibration methods...11
- 7.1 Calibration items...11
- 7.2 Calibration methods...12
- 7.3 Data processing...14
- 8 Expression of calibration results...17
- 9 Interval between recalibration...18
1 Scope
This Specification applies to the calibration of temperature and humidity parameters for environmental testing equipment - such as drying ovens, incubators, climatic aging chambers, mold test chambers, salt spray test chambers, corrosive gas test chambers, high/low-temperature test chambers, alternating damp-heat test chambers, and constant temperature and humidity chambers - operating within a temperature range of -80°C~
300°C and a humidity range of 10% RH~100% RH.
Temperature and humidity parameters of other types of environmental testing equipment may also be calibrated with reference to this Specification.
2 Normative references
This Rules refers to the following documents.
JJF 1001-2011, General Terms in Metrology and Their Definitions
GB/T 5170.1-2016, Inspection methods for environmental testing equipment for electric and electronic products -- Part 1.General
GB/T 5170.2-2017, Inspection methods for environmental testing equipment -- Part
2.Temperature testing equipment
GB/T 5170.5-2016, Inspection methods for environmental testing equipment for electric and electronic products -- Part 5.Damp heat testing equipment
For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies to this
Specification.
3 Terms
The terms and definitions defined in JJF 1001-2011 and GB/T 5170.1-2016 apply to this Specification.
3.1 environmental testing equipment
Equipment used to conduct environmental testing on products by simulating one or more environmental conditions.
3.2 working space
The portion of the space within environmental testing equipment where specified temperature and relative humidity conditions are maintained within defined tolerance limits.
3.3 steady state
The state in which the variations in temperature and relative humidity at any point within the working space of environmental testing equipment meet the equipment's own performance specifications.
3.4 temperature deviation
For environmental testing equipment in a steady state, temperature deviation refers to the upper and lower deviations of the highest and lowest temperatures actually measured at various points within the working space over a specified period relative to the set temperature. Temperature deviation comprises both upper temperature deviation and lower temperature deviation.
[GB/T 5170.1-2016, Definition 3.2.4]
3.5 relative humidity deviation
For environmental testing equipment in a steady state, the relative humidity deviation refers to the upper and lower deviations of the highest and lowest relative humidity values - measured at various points within the working space over a specified period - relative to the set relative humidity. Relative humidity deviation comprises both the upper relative humidity deviation and the lower relative humidity deviation.
[GB/T 5170.1-2016, Definition 3.2.5]
3.6 temperature fluctuation
The variation in temperature over time at any point within the working space of environmental testing equipment, under steady-state conditions and within specified time intervals.
3.7 relative humidity fluctuation
The variation of relative humidity over time at any point within the working space of environmental testing equipment, under steady-state conditions and within specified time intervals.
3.8 temperature uniformity
The maximum temperature difference between any two points in the working space of environmental testing equipment at any given instant under steady-state conditions.
[GB/T 5170.1-2016, Definition 3.2.8]
3.9 relative humidity uniformity
The maximum difference in relative humidity between any two points in the working space of environmental testing equipment at any given instant under steady-state conditions.
[GB/T 5170.1-2016, Definition 3.2.9]
4 Overview
Environmental testing equipment is a collective term for all testing apparatus designed to simulate one or more environmental conditions. Based on the categories of environmental factors simulated, such equipment is classified into types including climatic, mechanical, electromagnetic, and combined environmental testing equipment.
It is primarily used in sectors such as electrical and electronics, chemicals, and healthcare for applications including research testing, product qualification (type) testing, production inspection, product acceptance testing, safety and reliability testing, as well as failure analysis and verification.
5 Metrological characteristics
Common technical requirements for temperature deviation, temperature fluctuation, temperature uniformity, relative humidity deviation, relative humidity fluctuation, and relative humidity uniformity of environmental testing equipment are shown in Table 1.
Table 1 -- Technical requirements for temperature and humidity in environmental testing equipment
6 Calibration conditions
6.1 Environmental conditions
Temperature. 15°C~35°C;
Humidity. <= 85% RH;
Atmospheric pressure. 80 kPa~106 kPa.
The area surrounding the environmental testing equipment shall be free from strong vibrations and corrosive gases. The influence of other cold or heat sources shall be avoided. In actual operation, environmental conditions must also meet the requirements for the proper use of measurement standards.
6.2 Load conditions
Calibration is generally performed under no-load conditions; however, it can also be carried out under load conditions depending on user requirements, provided that the load conditions are specified.
6.3 Measurement standards and other equipment
6.3.1 Temperature measurement standards
For temperature measurement standards, multi-channel temperature display instruments or multi-channel temperature measurement devices shall generally be selected. Four-wire platinum resistance thermometers are the preferred choice for sensors. There shall be no fewer than nine sensors, and the setup must meet the requirements of the calibration work.
6.3.2 Humidity measurement standards
For humidity measurement standards, multi-channel temperature and humidity display instruments or multi-channel temperature and humidity measurement devices shall generally be selected. The number of sensors must be at least three and sufficient to meet calibration requirements.
6.3.3 Technical requirements
The number of standard temperature and humidity sensors used for measurement shall meet the requirements for calibration point distribution. Sensors of the same model and specifications shall be used for all channels. Technical specifications for the standard temperature and humidity sensors are provided in Table 2.
Table 2 -- Technical requirements for measurement standards
6.3.4 For calibration, the measurement standards listed in Table 2 may be used, or other measurement standards with compliant uncertainty may be selected.
7 Calibration items and calibration methods
7.1 Calibration items
See Table 3 for calibration items.
Table 3 -- Calibration items for temperature and humidity parameters of environmental testing equipment
7.2 Calibration methods
7.2.1 Selection of temperature and humidity calibration points
Temperature and humidity calibration points are generally selected based on user requirements - using commonly used values - or by choosing the lower limit, upper limit, and midpoint of the equipment's operating range.
7.2.2 Measurement point locations
The sensors are to be positioned at the measurement points used for equipment calibration and distributed across three distinct levels within the equipment's working space. the upper, middle, and lower levels. The middle level is the calibration plane that passes through the geometric center of the working space and lies parallel to the bottom surface. The distance from each measurement point to the inner wall of the equipment shall be one-tenth of the corresponding side length. Where air ducts are present, this distance may be increased but must not exceed 500 mm. If the equipment is equipped with a sample rack or trolley, the measurement points for the lower level may be positioned 10 mm above the rack or trolley.
The placement of sensor measurement points can also be arranged according to the user's actual operational requirements.
7.2.3 Number of measurement points
Temperature sensor measurement points are designated by numbers 1, 2, 3... Humidity sensor measurement points are designated by letters A, B, C...
7.2.3.1 For equipment with a volume of 2 m³ or less, there are 9 temperature measurement points and 3 humidity measurement points. Temperature point 5 and humidity point O are located at the geometric center of the middle layer of the equipment's working space, as shown in Figure 1.
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.
Referenced standards
Normative references
- JJF 1001-2011General Terms in Metrology and Their Definitions
- GB/T 5170.1-2016Inspection methods for environmental testing equipments for electric and electronic products—Part 1: General
- GB/T 5170.2-2017Inspection methods for environmental testing equipments—Part 2: Temperature testing equipments
- GB/T 5170.5-2016Inspection methods for environmental testing equipments for electric and electronic products—Part 5: Damp heat testing equipments
How to Download JJF 1101-2019
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 5170.1-2016 — Inspection methods for environmental testing equipments for electric and electronic products—Part 1: General
GB/T 5170.2-2017 — Inspection methods for environmental testing equipments—Part 2: Temperature testing equipments
GB/T 5170.5-2016 — Inspection methods for environmental testing equipments for electric and electronic products—Part 5: Damp heat testing equipments